How to calibrate a thermometer: ice point, boiling point and comparison calibration

How to calibrate a thermometer: the ice-point check done right, the boiling-water check corrected for altitude, what to do if it reads wrong, and the comparison procedure labs use, with test points, immersion, as-found and as-left results, uncertainty and records.

Updated 19 primary sourcesHow we researchCite

On this page
  1. How to calibrate a thermometer with ice water (ice-point method)
  2. Boiling-water check: correct for altitude and pressure
  3. What to do if your thermometer reads wrong
  4. Thermometer calibration vs a check
  5. What you need to calibrate a thermometer
  6. Test points and number of readings
  7. Step-by-step thermometer calibration procedure
  8. Acceptance criteria and measurement uncertainty
  9. What to record on a thermometer calibration certificate
  10. How often to calibrate a thermometer
  11. How to calibrate an infrared thermometer
  12. Common mistakes when checking a thermometer
  13. Related guides
  14. FAQ
  15. Sources

To calibrate a thermometer at home or in a kitchen, use the ice-point method: pack a glass with crushed ice, add water to the top and stir, put the probe in the middle at least 2 inches deep without touching the sides or bottom, wait at least 30 seconds, and check that it reads 32 °F (0 °C) (USDA FSIS). A food thermometer passes if it reads within ±2 °F, or ±1 °C if it reads in Celsius (FDA Food Code, 4-203.11). If it is off and has an adjustment nut or calibration button, adjust it while it is still in the ice. That check tests one temperature. A full calibration compares the thermometer with a calibrated reference at several temperatures across the range you use, and records the error and its uncertainty.

This page covers both: the ice and boiling-water checks done properly, what to do when a thermometer reads wrong, and then the comparison procedure a lab or in-house calibration room follows, with a worked example, the records to keep and how often to recalibrate.

How to calibrate a thermometer with ice water (ice-point method)

The ice point is the most reliable check you can make without special equipment, because a properly made mixture of ice and water sits at 0 °C without needing another thermometer to tell you so. NIST recommends it as a secondary fixed point for calibrating thermometers, reproducible in a lab to about 0.002 °C when done carefully (NIST SP 1088, §7.2). The catch is “properly made”.

✓ Correct≥ 2 in(5 cm)no contact with sides or bottom✗ Wrongcubes floating, tip on the bottom

Stir, wait at least 30 seconds, then read. A correct ice bath reads 32 °F (0 °C); a food thermometer passes if it reads within ±2 °F (±1 °C). Schematic, not to scale.

The ice-point check done right and done wrong. Crushed ice packed to the top with water filling the gaps; probe in the middle, deep enough, touching nothing. Source: method: USDA FSIS, Kitchen Thermometers; acceptance: FDA Food Code 2026, 4-203.11. Diagram by Acribi.
  1. Fill a large glass with finely crushed ice. USDA specifies finely crushed ice, and NIST shaves it to particles of 2 to 5 mm for lab work (NIST SP 1088, §7.7): small pieces pack tightly, with no pockets of plain water.
  2. Add clean tap water to the top of the ice and stir well. The ice point is ice saturated with water: the ice should fill the glass, not float in a layer of water (NIST SP 1088, §7.7).
  3. Insert the stem or probe at least 2 inches, in the middle, touching neither the sides nor the bottom of the glass. A dial (bimetal) thermometer senses along the last 2 to 2½ inches of the stem, so the whole sensing area has to be in the slush (USDA FSIS).
  4. Wait at least 30 seconds (longer if the reading is still changing).
  5. Read it. It should show 32 °F (0 °C).
  6. Adjust if you can and need to. On most dial thermometers, hold the adjusting nut under the dial with a suitable tool and turn the head until the pointer reads 32 °F, without taking the stem out of the ice (USDA FSIS). Digital models with a calibration or reset function have their own steps: follow the manual.

What counts as a pass? The FDA Food Code, the model code FDA offers to state and local regulators for restaurants and retail food, requires food thermometers to be accurate to ±1 °C if they read in Celsius or in both scales, and ±2 °F if they read only in Fahrenheit, over their intended range of use (4-203.11). Thermometers for ambient air and water, such as the one in a refrigerator, may be off by up to ±1.5 °C or ±3 °F (4-203.12). The 2026 edition, published in September 2026, keeps the limits of the 2022 edition. Your state or local code decides which edition applies.

The lab version of the ice point

For checking reference and liquid-in-glass thermometers, NIST describes a stricter recipe (NIST SP 1088, §7.7–7.8):

  • Ice made from distilled or deionized water (or the clear part of commercial ice), shaved to particles of 2 to 5 mm, packed in a vacuum (Dewar) flask and saturated with distilled water.
  • After 15 to 30 minutes, siphon off the meltwater and add ice. The bath is ready when the ice is saturated but not floating and there is no water on top. Make a fresh one every day.
  • Clean the thermometer and your hands (salt from skin contaminates the bath), pre-chill the thermometer, and leave at least 3 cm of slush below the bulb.
  • Read a mercury thermometer after about three minutes, an organic-liquid one after about 15 minutes, looking perpendicular to the stem to avoid parallax.

Ice made from tap water is often good enough for routine work: EURAMET notes that in many countries it gives an ice point between −0.025 °C and 0 °C, although deionized water is best (EURAMET cg-8, §4.1).

Boiling-water check: correct for altitude and pressure

The boiling-water method works the same way as the ice point: immerse the stem at least 2 inches in water at a full rolling boil, wait at least 30 seconds and compare with the boiling point (USDA FSIS). The problem is the boiling point itself. Water boils at 212 °F only in distilled water at a pressure of one standard atmosphere, 29.921 inHg; USDA notes that tap water under unknown conditions will probably boil at least 2 °F, and perhaps as much as 5 °F, lower (USDA FSIS). Strictly, pure water at standard pressure boils at 99.974 °C (211.95 °F) on today’s temperature scale (NIST Chemistry WebBook), which rounds to 212 °F.

Pressure falls with altitude, and the boiling point falls with it:

Boiling point of pure water by altitude (standard atmosphere)
AltitudePressureWater boils at
0 ft 0 m29.92 inHg 1,013 hPa212.0 °F 100.0 °C
1,000 ft 305 m28.86 inHg 977 hPa210.1 °F 99.0 °C
2,000 ft 610 m27.82 inHg 942 hPa208.3 °F 97.9 °C
3,000 ft 914 m26.82 inHg 908 hPa206.5 °F 96.9 °C
4,000 ft 1,219 m25.84 inHg 875 hPa204.7 °F 95.9 °C
5,000 ft 1,524 m24.89 inHg 843 hPa202.8 °F 94.9 °C
6,000 ft 1,829 m23.98 inHg 812 hPa201.0 °F 93.9 °C
7,000 ft 2,134 m23.09 inHg 782 hPa199.2 °F 92.9 °C
8,000 ft 2,438 m22.22 inHg 753 hPa197.3 °F 91.8 °C
10,000 ft 3,048 m20.57 inHg 697 hPa193.7 °F 89.8 °C
Boiling point of pure water by altitude in the standard atmosphere, rounded to 0.1° (at sea level, 99.97 °C or 211.95 °F). Real pressure changes with the weather, so treat these values as a guide. Source: computed by Acribi: saturation temperature from IAPWS-IF97 (IAPWS R7-97(2012), Eq. 31); pressure at altitude from the NOAA National Weather Service pressure-altitude formula; 1 inHg = 3.386 38 kPa (NIST SP 811). Checked against the NIST Chemistry WebBook (boiling point within 0.001 °C), the FAA standard-atmosphere table (pressures, FAA-H-8083-25B, Fig. 11-2) and USDA FSIS (208 °F at 2,000 ft; about 198 °F at 7,500 ft).

That is roughly 1.8 °F less for every 1,000 ft, in line with USDA’s rule of thumb of just under 1 °F for every 500 ft (USDA FSIS, High Altitude Cooking). Two more things move the result:

  • The weather. The table assumes the standard atmosphere, but the actual pressure changes from day to day. Near 100 °C, the boiling point changes by about 0.28 °C (0.5 °F) for every kPa (0.3 inHg) of pressure (NIST Chemistry WebBook). If you know the actual pressure, the table’s pressure column gives the boiling point more accurately than the altitude does.
  • Which pressure. The pressure in weather reports is usually corrected to sea level (the altimeter setting). Boiling depends on the actual pressure where you are, the station pressure, which is lower at altitude (NOAA NWS, Station Pressure).

Use the boiling-water check for thermometers you use at high temperatures (candy, frying), and treat it as a rough one. The ice point has none of these problems.

What to do if your thermometer reads wrong

First rule out the method: a bath with too much water, a probe touching the glass, or a reading taken too early. Then you have three options (USDA FSIS):

  • Adjust it, if it has an adjustment: a calibration nut on dial thermometers, a calibration or reset function on some digital ones, or, on some liquid-filled thermometers, moving the glass stem in its holder.
  • Allow for the error. USDA’s example: a thermometer that reads 214 °F in boiling water reads 2 °F high, so you subtract 2 °F from its readings, or cook ground beef to 162 °F on it instead of 160 °F. This only works if the error is the same at the temperatures you use, which a one-point check can’t prove.
  • Replace it.

Not every thermometer can be adjusted, and each type has its own weak points:

Type How it senses Can you adjust it? Watch out for
Bimetal dial (oven-safe or instant-read) A coil of two metals; averages the temperature over the last 2–2½ in of the stem Some models (USDA FSIS) The whole sensing area must be immersed; can’t measure thin foods unless inserted sideways
Digital, thermistor probe Semiconductor resistor in the tip Not all models (USDA FSIS) Strongly nonlinear, with a narrower range than platinum (DKD-R 5-1, §3.2)
Digital, thermocouple probe Junction of two wires; fastest response Can be calibrated (USDA FSIS) The signal is generated along the wire, so depth matters; type K drifts after use at 500 °C or above (EURAMET cg-8, §9.1, §10.1)
Platinum resistance (RTD, Pt100) Platinum resistor; the usual reference thermometer Instrument dependent Hysteresis and self-heating (DKD-R 5-1, §8.8, §8.10)
Liquid-in-glass Liquid expanding in a capillary Some can be reset in their holder (USDA FSIS) Separated columns; partial vs total immersion; bulb changes after heating (NIST SP 1088, §2–3)
Oven-cord and fork thermometers Thermistor or thermocouple Cannot be calibrated (USDA FSIS) Check them and replace if off

For a digital thermometer with interchangeable probes, remember that the probe and the display are two parts. A calibrator that simulates the sensor’s electrical signal only tests the display, not the probe (EURAMET cg-11, §1.2). Check the probe and display together, in a bath or block.

Before blaming the thermometer, also check how you use it: the center of a food is usually cooler than the surface, so the tip belongs in the center of the thickest part (USDA FSIS), and a Food Code establishment also needs a thin-probe thermometer for thin foods such as burger patties and fish fillets (4-302.12).

Thermometer calibration vs a check

Ice-point or boiling check Calibration
Reference Melting ice (or boiling water, corrected) A calibrated reference thermometer, traceable to national standards, in a bath or dry block
Points One At least two, covering the range of use (DKD-R 5-1, §5)
Readings One, after it settles Several, after both thermometers have stabilized
Uncertainty Not evaluated Evaluated and stated for each point
Result Pass or adjust As-found and as-left errors, pass or fail against the tolerance, certificate
Good for Kitchens, daily checks, catching a damaged thermometer between calibrations Quality systems, labs, HACCP verification, reference thermometers, audits

Calibration and adjustment are different operations: calibration compares the thermometer with the reference and records the error; adjustment changes the thermometer, and after an adjustment it usually has to be calibrated again (VIM 2.39, 3.11). Our guide to what calibration is explains the difference.

“NIST traceable” on the box is a claim that the calibration links to NIST standards through an unbroken chain of calibrations with stated uncertainties. NIST does not certify other organizations’ traceability claims: the supplier has to support the claim, and the user has to judge it (NIST traceability policy). What you want to see is a certificate with results at stated temperatures and their uncertainty.

What you need to calibrate a thermometer

  • Reference thermometer

    A platinum resistance thermometer (PRT, or SPRT for the lowest uncertainty) with a calibrated readout, or a calibrated reference thermometer

    Calibrated by an accredited lab. A lab should have two reference standards per range, or one plus at least two fixed points such as the ice point, to catch drift.

    DKD-R 5-1, §4 and §9

  • Accuracy ratio

    Reference system uncertainty a small fraction of the tolerance being checked

    A common target is 4:1: for a ±1 °C food thermometer, about ±0.25 °C or better. Check yours with the TUR calculator.

  • Temperature source

    A stirred liquid bath or a dry block calibrator, characterized for uniformity and stability

    Both thermometers must sit in the useful volume, where the temperature is uniform and stable. In a dry block, the hole or insert should be at most 0.5 mm wider than the probe (up to 660 °C).

    DKD-R 5-1, §7; EURAMET cg-13, §2.2

  • Ice point

    Crushed ice and water in an insulated flask

    For the as-found check of the thermometer and as an intermediate check of the reference.

    NIST SP 1088, §7.7; DKD-R 5-1, §9

  • Environment

    Ambient conditions within the specifications of the readout and electrical equipment

    Monitor and record them if they affect the result.

    DKD-R 5-1, §4

  • Paperwork

    The thermometer specification (range, tolerance), your procedure and a record form

    Know the tolerance and the range of use before you start. See how to write a calibration procedure.

The temperature source matters as much as the reference. The calibration of a dry block mainly relates to the temperature of the block: the thermometer you put in it can be at a different temperature, especially with a different insert or probe than in the block’s calibration (EURAMET cg-13, Annex C). That is why a comparison calibration reads an external reference thermometer in the same block, as in DKD’s worked example, rather than the block’s own display (DKD-R 5-1, Appendix A.3).

Reference readout50.012 °CUnder test50.6 °Cstirred bath or dry blockusefulvolumereference PRTthermometer under testimmersion≥ 10 × Øto top ofsensorsame depth

Error = indication − reference temperature, read after both thermometers have stabilized. In a dry block, the hole (or insert) should be no more than 0.5 mm wider than the probe up to 660 °C. Schematic, not to scale.

Comparison calibration: the reference and the thermometer under test side by side, sensing elements at the same depth in the zone of uniform temperature. Source: DKD-R 5-1, §7 and §8.3; EURAMET cg-13, §2.2. Diagram by Acribi.

Test points and number of readings

DKD’s guideline for resistance thermometers asks for at least two temperatures, with the points and range agreed with the customer, and states that results are valid only between the lowest and highest calibrated temperature (DKD-R 5-1, §1 and §5). The Food Code’s accuracy requirement applies across the whole range of use, from refrigeration to cooking temperatures if the thermometer is used for both (Food Code, Annex 3, 4-203.11). Put together:

  • Cover the lowest and highest temperatures you measure, plus at least one point in between.
  • Include 0 °C when it is in the range: it is the easiest point to realize and the one you will use for checks between calibrations. EURAMET asks for it with thermocouples (EURAMET cg-8, §12.6).
  • Calibrate thermocouples from the highest temperature down (EURAMET cg-8, §10.1). For industrial platinum thermometers, rising and falling temperatures can give different results (hysteresis), so a lab tests both directions at mid-range when it needs that figure (DKD-R 5-1, §8.10).

The plan for the example thermometer below, a probe used from freezer to cooking temperatures:

×3×3ice point×3mid×3high
  • −20 °C: 3 readings
  • 0 °C, ice point: 3 readings
  • 50 °C, mid: 3 readings
  • 100 °C, high: 3 readings
  • 12 readings in total. Wait for stability at each point; record every reading
Test-point plan for a −50 to 150 °C digital thermometer used between −20 and 100 °C: four points, three readings each. Source: at least two points within the range of use: DKD-R 5-1, §1 and §5; 0 °C included: EURAMET cg-8, §12.6. Number of readings chosen for this example. Diagram by Acribi.

How many readings? No free standard fixes a number for general thermometers. The worked example in DKD-R 5-1 averages 20 readings taken every 30 seconds over 10 minutes after an hour of stabilization (Appendix A.3). Whatever you choose, write it into your procedure and use the spread of the readings in the uncertainty.

Step-by-step thermometer calibration procedure

The sequence follows DKD-R 5-1 and EURAMET cg-13, plus the as-found and as-left logic of any calibration. Adjust it to your thermometer type and write your choices into your calibration procedure.

  1. Identify and inspect

    Record ID, model, range, resolution and tolerance. Check the probe, cable and display; for glass thermometers, look for separated columns, bubbles and cracks with a magnifier.

    Damaged: tag out, record as not calibratable

  2. Ice-point check

    Read the thermometer in an ice point. For reference and resistance thermometers, a change since the last calibration is the first sign of damage from shock or transport (DKD-R 5-1, §5–6). Wait 72 hours after use at high temperature for glass thermometers (NIST SP 1088, §7.4).

  3. Mount

    Place the thermometer and the reference side by side in the bath or block, sensing elements at the same depth in the useful volume. As a first estimate, immerse at least 10 times the probe diameter, measured from the top of the sensing element: that keeps the error below about 1 % of the difference between the bath and room temperature, and each further 10 diameters cuts it by a factor of 10 (DKD-R 5-1, §8.3).

  4. Stabilize

    Set the first temperature and wait until both readings are stable. Take readings of both thermometers at the same time, or average them over the same period (DKD-R 5-1, §7).

  5. As-found readings

    At each test point, record the reference temperature and the reading of the thermometer under test.

  6. Check immersion

    Withdraw the thermometer slightly, about 10 % of the depth or 20 mm in a dry block. DKD's criterion: the change should not exceed 10 % of the uncertainty you are aiming for; if it does, heat conduction along the stem is affecting the reading and its uncertainty contribution must be increased (DKD-R 5-1, §8.3; EURAMET cg-13, Annex C).

  7. Evaluate

    Error = reading − reference temperature, at every point. Compare it with the tolerance.

    Outside tolerance: record as found out of tolerance, open an out-of-tolerance report

  8. Adjust only if needed

    Use the manufacturer's adjustment method, and only past the threshold your procedure sets.

  9. As-left readings

    After any adjustment, repeat all points as a separate set. Report both: DKD requires results before and after adjustment, especially for direct-reading thermometers (DKD-R 5-1, §10).

  10. Uncertainty, record and label

    Evaluate the uncertainty at each point, issue the certificate or record, and label the thermometer.

A contact thermometer only measures its own temperature. The calibration tells you how its reading relates to the temperature of its sensor; making sure the sensor is at the temperature you want to measure, in use, is up to the user (DKD-R 5-1, §2).

Acceptance criteria and measurement uncertainty

Worked example: passes at the ice point, fails at 100 °C

The example thermometer, TH-117, is a handheld digital thermometer with a Pt100 probe, −50 to 150 °C with 0.1 °C resolution, used in a food plant. Its tolerance is the Food Code limit, ±1 °C over the range it is used in. It was compared with a reference PRT in a stirred bath at four points, three readings each. The error of each reading is the indication minus the reference temperature. The values are invented for this example.

+1 °C−1 °CAs found, −20 °C: reading -20.4 °C, error −0.4 °CAs found, −20 °C: reading -20.3 °C, error −0.3 °CAs found, −20 °C: reading -20.4 °C, error −0.4 °CAs found, 0 °C: reading 0.1 °C, error +0.1 °CAs found, 0 °C: reading 0.1 °C, error +0.1 °CAs found, 0 °C: reading 0.2 °C, error +0.2 °CAs found, 50 °C: reading 50.6 °C, error +0.6 °CAs found, 50 °C: reading 50.7 °C, error +0.7 °CAs found, 50 °C: reading 50.6 °C, error +0.6 °CAs found, 100 °C: reading 101.2 °C, error +1.2 °CAs found, 100 °C: reading 101.3 °C, error +1.3 °CAs found, 100 °C: reading 101.2 °C, error +1.2 °CAs left, −20 °C: reading -20.1 °C, error −0.1 °CAs left, −20 °C: reading -20.2 °C, error −0.2 °CAs left, −20 °C: reading -20.1 °C, error −0.1 °CAs left, 0 °C: reading 0 °C, error 0.0 °CAs left, 0 °C: reading 0.1 °C, error +0.1 °CAs left, 0 °C: reading 0 °C, error 0.0 °CAs left, 50 °C: reading 50.1 °C, error +0.1 °CAs left, 50 °C: reading 50.2 °C, error +0.2 °CAs left, 50 °C: reading 50.1 °C, error +0.1 °CAs left, 100 °C: reading 100.2 °C, error +0.2 °CAs left, 100 °C: reading 100.3 °C, error +0.3 °CAs left, 100 °C: reading 100.2 °C, error +0.2 °C

As foundAs leftoutside ±1 °Cmean

Reference temperatureAs found: mean error (worst)As left: mean error (worst)
−20 °C−0.4 °C (−0.4)pass−0.1 °C (−0.2)pass
0 °C+0.1 °C (+0.2)pass+0.0 °C (+0.1)pass
50 °C+0.6 °C (+0.7)pass+0.1 °C (+0.2)pass
100 °C+1.2 °C (+1.3)3 of 3 outside+0.2 °C (+0.3)pass
Example calibration of a fictitious food-plant thermometer: error of every reading at each reference temperature, as found and as left, against the ±1 °C tolerance. Source: tolerance: FDA Food Code 2026, 4-203.11(A); example data by Acribi, computed at build time.

How to read it:

  • As found, 0 °C: +0.1 to +0.2 °C. An ice-point check would have passed this thermometer.
  • As found, 50 and 100 °C: the error grows with temperature, to +0.6 °C at 50 °C and +1.2 to +1.3 °C at 100 °C, outside ±1 °C. This is a span (slope) error, the kind a one-point check can’t see. The out-of-tolerance report has to ask what was measured above about 80 °C with this thermometer since its last good calibration.
  • As left: after a two-point adjustment in the instrument’s calibration menu, every reading is within ±0.3 °C.

Close to the limit, whether a result counts as a pass depends on the uncertainty and on your decision rule (Global ACI-TECH-1-002, formerly ILAC G8). Our guard banding guide explains the options.

What goes into the uncertainty

The uncertainty of the calibration combines the reference and the source with contributions from the thermometer itself, which can be many times larger than the reference’s (DKD-R 5-1, §8). DKD’s worked example for a digital thermometer with a Pt100 probe, calibrated against a working-standard PRT in a dry block at about 350 °C, gives a real budget (DKD-R 5-1, Appendix A.3):

Contribution What it captures Standard uncertainty
Readings Spread of 20 readings (standard deviation of the mean) 0.070 K
Reference temperature The working standard and its readout, block uniformity 0.032 K
Resolution 0.1 °C display, rectangular 0.029 K
Self-heating Measuring current heating the Pt100 (30 mK default when not measured) 0.017 K
Heat conduction Change when the probe is withdrawn by twice its diameter 0.029 K
Hysteresis Difference between rising and falling calibrations at mid-range 0.087 K
Combined, expanded (k = 2) U = 0.25 K

The thermometer read 0.23 K high, with an expanded uncertainty of 0.25 K. Hysteresis, a property of the probe, was the largest single contribution; DKD notes it often is (DKD-R 5-1, §1). Build or check your own budget with the uncertainty calculator.

What to record on a thermometer calibration certificate

Whether you issue a certificate or fill in an internal record, keep:

  • Identification: thermometer and probe IDs or serial numbers, manufacturer, model, range, resolution.
  • Method: your procedure number and revision, and the guideline it follows (for example, DKD-R 5-1).
  • Equipment: the reference thermometer and readout with their traceability, the bath or block, and the insert used.
  • Conditions: immersion depth (DKD asks for it on the certificate so the calibration can be repeated, §8.3), ambient conditions, stabilization time.
  • Results per point: reference temperature, readings or their mean, error, as found and as left.
  • Uncertainty: the expanded uncertainty at each point, and for resistance thermometers the self-heating and hysteresis behavior and the measuring current (DKD-R 5-1, §10).
  • Decision: pass or fail against the stated tolerance, the decision rule used, and any adjustment made.
  • Who and when: technician, date, next due date.

Our guides to calibration certificates and calibration records cover each field in detail. Print a matching status sticker with the free calibration label generator.

How often to calibrate a thermometer

There is no universal interval, but the sources agree on the logic:

  • Food thermometers: calibrate according to the manufacturer’s specifications as often as necessary to keep them accurate (Food Code, 4-502.11(B)). In practice, that means a regular ice-point check and a check after any drop or shock.
  • Liquid-in-glass thermometers: check at the ice point at least once a year even when used near room temperature. New ones, or ones used at high temperatures, may start with checks as often as monthly, then every two months, then once or twice a year if the ice point stays stable (NIST SP 1088, §4 and §7.6). Mercury thermometers used at 200 °C or below can be recalibrated at the ice point alone; above 200 °C they need full recalibration, and NIST does not recommend single-point recalibration for organic-liquid thermometers (NIST SP 1088, §1).
  • Reference thermometers: DKD recommends recalibration at most every 2 to 3 years for SPRTs and every 1 to 2 years for industrial PRTs, depending on their history, with ice-point checks in between (DKD-R 5-1, §9, Table 1).
  • Thermocouples: there is no formal frequency; EURAMET recommends a checking and recalibration program built on your own experience, with in-place checks against a reference for installed thermocouples (EURAMET cg-8, §13).

Then let your own data set the interval: extend it if thermometers keep coming back in tolerance, shorten it if they don’t. Our calibration intervals guide explains the methods.

When a thermometer fails calibration

An as-found result outside tolerance means every measurement made with the thermometer since its last good calibration may be wrong, at the temperatures where it failed. In the example, the thermometer read more than 1 °C high near 100 °C: food it showed at a safe cooking temperature could have been more than 1 °C cooler. Our guide to out-of-tolerance results walks through the impact assessment.

How to calibrate an infrared thermometer

Infrared thermometers measure the thermal radiation from a surface, so they are calibrated against a radiation source of known temperature, a blackbody or flat-plate calibrator, not in a bath. ASTM E2847 is the US test method for handheld, direct-reading, wideband infrared thermometers that measure below 1000 °C, including the uncertainty calculation and the report (ASTM E2847-21(2026), scope). In practice (Fluke Calibration):

  • Emissivity: set the thermometer’s emissivity to that of the source, or correct for the difference.
  • Distance and alignment: the thermometer’s field of view must fall inside the source; aim it square to the surface, within 5°.
  • Points: one point if the thermometer is used over a narrow range, three (minimum, middle and maximum) for a wide range.
  • Reflected heat: keep other heat sources and people away from the target, especially at low temperatures.

Fluke lists emissivity as a major source of uncertainty in the calibration; in use, too, the reading depends on the emissivity setting matching the surface you measure.

Common mistakes when checking a thermometer

Mistake Why it matters Do this instead
Ice without water, or cubes floating in water Neither is ice saturated with water, so neither is at the ice point Crushed ice packed to the top, water filling the gaps, stirred
Probe touching the glass or not deep enough Reads the glass or the air, not the slush At least 2 in deep, sensing area fully immersed, touching nothing
Reading too soon The thermometer is still settling Wait at least 30 s, or until the reading stops moving
Assuming water boils at 212 °F It depends on pressure: about 203 °F at 5,000 ft Use the ice point, or correct the boiling point for altitude and pressure
Adjusting at one point and calling it calibrated A span error stays hidden and the as-found result is lost Record as found first; test near the temperatures you use
Shallow immersion in a bath or block Heat flows along the stem and the reading lags At least 10 × the probe diameter; check by withdrawing slightly
Loose fit in a dry block Air gaps add error and uncertainty Insert hole at most 0.5 mm wider than the probe (up to 660 °C)
Trusting the dry block display The block’s sensor is not where your probe is Use an external reference thermometer in the block
Calibrating the display with a simulator only The probe is not tested Check probe and display together

FAQ

What should a thermometer read in ice water?

32 °F (0 °C). That is the answer to the food-safety exam question about the ice-point method. Use a glass packed with crushed ice, topped up with water and stirred, with the sensing area at least 2 inches deep and touching neither the sides nor the bottom, and wait at least 30 seconds. Under the FDA Food Code, a food thermometer must be accurate to ±2 °F (±1 °C if it reads in Celsius) over the range you use it.

What is one effective way to calibrate a thermometer?

The ice-point method: put the thermometer in a properly made mixture of crushed ice and water and check that it reads 32 °F (0 °C). If it doesn't and it has an adjustment nut or a calibration button, adjust it while it is still in the ice. The boiling-water method is the other common check, but water only boils at 212 °F at standard sea-level pressure.

How do you calibrate a digital thermometer?

Check it in an ice-water slurry first. If it is off and your model has a calibration or reset function, follow the manufacturer's steps with the probe in the ice; many models have no adjustment, and then you either note the error and allow for it or replace the thermometer. For a traceable calibration, have the thermometer and its probe compared with a calibrated reference at several temperatures across the range you use.

Can you calibrate a thermometer with boiling water?

You can check it, but correct for pressure. Pure water boils at 212 °F only at standard sea-level pressure (29.92 inHg). At 5,000 ft it boils at about 203 °F, and the weather moves the boiling point too. USDA notes that tap water under unknown conditions will probably boil at least 2 °F, and perhaps 5 °F, lower than 212 °F. The ice point does not have this problem, which is why it is the preferred check.

How accurate does a food thermometer have to be?

Under the FDA Food Code (2026 edition, section 4-203.11), food temperature measuring devices must be accurate to ±1 °C if scaled in Celsius or in both scales, and ±2 °F if scaled only in Fahrenheit, across the range they are used in. Thermometers for ambient air and water (for example, refrigerator air) are allowed ±1.5 °C or ±3 °F (4-203.12). The Food Code is a model: your state or local code decides which edition applies.

How often should you calibrate a thermometer?

There is no single interval. The Food Code says food thermometers must be calibrated per the manufacturer's specifications as often as needed to keep them accurate. NIST recommends checking liquid-in-glass thermometers at the ice point at least once a year, more often when they are new or used hot. Check any thermometer after it is dropped, and let your calibration history lengthen or shorten the interval.

My digital thermometer has no calibration button. What now?

Check it anyway. USDA advises that a thermometer that cannot be adjusted should still be checked, and any error taken into account or the thermometer replaced. If it reads 34 °F in a good ice bath, it reads 2 °F high near freezing, but that tells you nothing certain about 165 °F; for that you need a check near the temperatures you use.

My two thermometers disagree. Which one is right?

Possibly neither. Put both in the same ice bath: one that is off at 32 °F is out. If both pass there and still disagree when hot, compare them side by side in a stirred hot-water bath with a calibrated reference thermometer, at the same depth. Also check placement: a probe that is not in the center of the thickest part of the food reads a different temperature, which is not a calibration problem.

What does NIST traceable mean on a thermometer?

It is a claim that the thermometer's calibration links to NIST standards through an unbroken chain of calibrations, each with a stated uncertainty. NIST does not certify or vouch for other organizations' traceability claims; the supplier has to support the claim and the user has to judge it. Look for a calibration certificate with results at stated temperatures, the uncertainty and the lab's accreditation, not just the words on the box.

How do you calibrate an infrared thermometer?

Against a calibrated radiation source (a blackbody or flat-plate calibrator) at known temperatures, with the thermometer's emissivity set to match the source, at the distance that keeps its field of view inside the target, and with reflected heat controlled. ASTM E2847 is the US test method for handheld wideband infrared thermometers. A contact-thermometer method like the ice bath does not apply directly.

Can a thermocouple be calibrated to make it more accurate?

Calibration gives you its error at the tested temperatures, and you can correct for it, but only within limits. A thermocouple's signal is generated along the wire where the temperature changes, not at the tip, so an inhomogeneous or aged thermocouple can read differently at a different immersion depth than in the calibration. Type K thermocouples also change when cycled to 500 °C or above. EURAMET recommends checking used thermocouples in place, with a reference alongside.

Sources

  1. U.S. Food and Drug Administration. FDA Food Code 2026 (4-203.11, 4-203.12, 4-302.12, 4-502.11 and Annex 3). 2026 — Published September 17, 2026; pp. 91, 100, 106
  2. USDA Food Safety and Inspection Service. Kitchen Thermometers (section: Calibrating a Thermometer) — Read via the Internet Archive copy of 2025-03-11 (the live page blocks automated access)
  3. USDA Food Safety and Inspection Service. High Altitude Cooking and Food Safety — Read via the Internet Archive copy of 2026-10-04
  4. NIST (Cross, Miller, Ripple, Strouse). NIST SP 1088, Maintenance, Validation, and Recalibration of Liquid-in-Glass Thermometers. 2009 — §1, §2, §4, §7.2, §7.4, §7.6–7.8
  5. PTB / DKD. DKD-R 5-1, Calibration of resistance thermometers. 2023 — §1, §2, §4, §5, §7, §8.3, §8.8, §8.10, §9, §10, Appendix A.3
  6. EURAMET. EURAMET cg-13, Guidelines on the Calibration of Temperature Block Calibrators (v4.0). 2017 — §2.2, §3.5, Annex C
  7. EURAMET. EURAMET cg-8, Guidelines on the Calibration of Thermocouples (v3.1). 2020 — §4.1, §9, §10.1, §12.6, §13
  8. EURAMET. EURAMET cg-11, Guidelines on the Calibration of Temperature Indicators and Simulators by Electrical Simulation and Measurement (v2.0). 2011 — §1.2
  9. IAPWS. IAPWS R7-97(2012), Revised Release on the IAPWS Industrial Formulation 1997 for the Thermodynamic Properties of Water and Steam. 2012 — §8.2, Eq. (31), Tables 34 and 36 (saturation temperature)
  10. NIST. NIST Chemistry WebBook, SRD 69: Thermophysical Properties of Fluid Systems (water, saturation properties) — Normal boiling point 99.974 °C at 101.325 kPa; queried 2026-10-16
  11. NOAA National Weather Service. Pressure Altitude (wxcalc)
  12. NOAA National Weather Service. Station Pressure (wxcalc)
  13. Federal Aviation Administration. Pilot's Handbook of Aeronautical Knowledge (FAA-H-8083-25B), Chapter 11. 2016 — Figure 11-2, Properties of standard atmosphere (pressure in inHg by altitude)
  14. NIST. NIST SP 811, Guide for the Use of the International System of Units (SI). 2008 — App. B.8: 1 inHg (32 °F) = 3.386 38 kPa
  15. NIST. NIST Policy on Metrological Traceability. 2024
  16. ASTM International. ASTM E2847-21(2026), Standard Test Method for Calibration and Accuracy Verification of Wideband Infrared Thermometers. 2026 — Paid standard; scope only
  17. Fluke Calibration. Infrared Thermometer Calibration – A Complete Guide
  18. JCGM / BIPM. International Vocabulary of Metrology (VIM), JCGM 200:2012. 2012 — 2.39 calibration, 3.11 adjustment
  19. Global Accreditation Cooperation (Global ACI). Global ACI-TECH-1-002 (G), Guidelines on Decision Rules and Statements of Conformity (formerly ILAC G8:09/2019). 2026