How to read and review a calibration certificate

A field-by-field guide to calibration certificates, with an annotated example, how to judge uncertainty and pass/fail statements, red flags, and a printable review checklist.

Updated Sources checked 26 primary sourcesHow we researchCite

On this page
  1. What a calibration certificate is (and what it isn’t)
  2. Anatomy of a calibration certificate: an annotated example
  3. How to read the results table
  4. Measurement uncertainty: what k=2 means and whether it’s good enough
  5. “Pass” isn’t always pass: decision rules and guard bands
  6. Accredited, traceable, or neither?
  7. When the as-found data is out of tolerance
  8. Red flags and common myths
  9. Calibration certificate review checklist
  10. What auditors expect, by standard
  11. Does a calibration certificate expire?
  12. FAQ
  13. Sources

A calibration certificate records what was measured, against which reference, with what uncertainty and with what result, on one date. Reviewing it means answering four questions: is this the right document for the right instrument, is the traceability and accreditation real, is the uncertainty small enough for your tolerance, and what happened before and after any adjustment? This guide walks through each part of a certificate, using an annotated example, and ends with a printable review checklist.

What a calibration certificate is (and what it isn’t)

Calibration compares an instrument’s indications with reference standards of known value and uncertainty, and documents that relationship (VIM 2.39). It is not the same as adjustment, which changes the instrument so it reads closer to the reference. A calibration service can include an adjustment, but the adjustment is a separate operation, and the instrument must then be calibrated again (VIM 2.39 Note 2; 3.11 Note 3). That is why you can get two sets of results, before and after (ISO/IEC 17025, clause 7.8.4.1).

Not every document that arrives with an instrument is a calibration certificate. The paperwork usually falls into one of these types:

Document What it contains When it is enough
Accredited calibration certificate (ISO/IEC 17025) Results with units, measurement uncertainty, traceability statement, accreditation symbol The preferred evidence in most certified quality systems (and required by IATF 16949), as long as the calibration is within the lab’s scope
Calibration certificate with data (“long form”, not accredited) Results and usually uncertainty, traceability claim Acceptable when you can show the provider is competent and traceable (extra evidence needed)
Certificate without data (“short form”, “traceable, no data”) A statement that the instrument met its specification Rarely enough for audits or out-of-tolerance investigations: there is nothing to review
Certificate of conformity, accuracy or inspection A statement that a product met a specification Not a calibration certificate. Some explicitly say so
In-house calibration record Your own results against your traceable standards Fine for many instruments if your reference standards are calibrated and your method is documented

Anatomy of a calibration certificate: an annotated example

ISO/IEC 17025 sets the minimum content of a calibration certificate. Clause 7.8.2.1 lists what every lab report needs (identification, customer, method, item, dates, results, authorization). Clause 7.8.4.1 adds the calibration-specific items: measurement uncertainty, the conditions that influence the results, a traceability statement, results before and after any adjustment or repair, and, when relevant, a statement of conformity (summarized in NIST’s 17025 crosswalk and SOP 1).

The example below is a fictitious certificate for a digital pressure gauge. It follows that structure, with numbered notes. Select a number to jump to its note.

2Example Calibration Laboratory, Inc.100 Sample Street, Anytown, ST 00000, USAAccredited to ISO/IEC 17025:2017 by an ILAC MRA signatory · Accreditation cert. no. 0000.01 (example)
1Certificate of CalibrationNo. ECL-2026-04817Page 1 of 1
3

Customer

Sample Manufacturing Co., 25 Example Avenue, Springfield, ST 00000

5

Dates

Received
2026-09-14
Calibrated
2026-09-16
Issued
2026-09-17
Due date
Not stated (interval set by customer)
4

Item calibrated

Description
Digital pressure gauge
Manufacturer / model
Example Instruments DPG-300
Serial no. / asset ID
21-0456 / PG-014
Range / resolution
0 to 300 psi / 0.01 psi
Condition on receipt
Received in working order, no visible damage
6

Method and conditions

Comparison against a reference pressure standard, procedure ECL-P-112 rev. 4. Pneumatic, ascending points, gauge mounted vertically.

Temperature 20.4 °C ± 1.0 °C · Relative humidity 43 % ± 10 %

7

Reference standards and traceability

Reference pressure monitor, 0–500 psi · ID ECL-STD-031 · cert. RC-2026-1123 · due 2027-03-02

Results are metrologically traceable to the International System of Units (SI) through an unbroken chain of calibrations, each contributing to the stated uncertainty, realized by national metrology institutes (e.g., NIST).

8

Results: As found

Nominal
(psi)
Reference value
(psi)
Gauge reading
(psi)
Error
(psi)
Tolerance
(± psi)
U, k=2
(psi)
Result
00.000.05+0.050.750.12Pass
7575.0275.20+0.180.750.12Pass
150150.01150.38+0.370.750.12Pass
225224.98225.55+0.570.750.12Pass
300300.03300.93+0.900.750.12Fail*
9

Results: As left (after adjustment)

Zero and span adjusted per manufacturer instructions after as-found test.

Nominal
(psi)
Reference value
(psi)
Gauge reading
(psi)
Error
(psi)
Tolerance
(± psi)
U, k=2
(psi)
Result
00.000.01+0.010.750.12Pass
7575.0175.04+0.030.750.12Pass
150150.02150.00−0.020.750.12Pass
225225.00224.97−0.030.750.12Pass
300300.02300.05+0.030.750.12Pass
10

Measurement uncertainty

The expanded uncertainty U is stated as the standard uncertainty multiplied by a coverage factor k = 2, corresponding to a coverage probability of approximately 95 %.

11

Statement of conformity

Manufacturer specification: ±0.25 % of full scale (±0.75 psi).

As found: OUT OF TOLERANCE at 300 psi. As left: IN TOLERANCE at all test points.

Decision rule: Simple acceptance with TUR ≥ 4:1 (TUR at every test point: 6.2:1). Agreed with customer at order review.

* Out of tolerance.

12


Authorized by: J. Rivera, Technical Manager (approved signatory)

Results relate only to the item calibrated. This certificate may not be reproduced except in full without written approval of the laboratory.

  1. Title, unique number, page count

    Every page carries the certificate number and "page X of Y", so a missing page is obvious. Quote this number in your equipment record.

  2. Who calibrated it, and under which accreditation

    Lab name and address, plus the accreditation body symbol and accreditation number if the work is accredited. The symbol only means something if this calibration is inside the lab's scope (see below).

  3. Customer

    Your company. Information you supplied (for example, the tolerance) is your responsibility, not the lab's, and should be identified as such.

  4. Item identification

    Model, serial number and your asset ID must match the instrument on your shelf. If they do not match, the certificate is not evidence for the instrument on your shelf.

  5. Three dates

    Received, calibrated and issued. A missing issue date is one of the most common certificate findings A2LA reports. On an accredited certificate, a due date normally appears only if you asked for one.

  6. Method and conditions

    Which procedure was used and under which conditions (temperature, humidity). Check that the method fits how you use the instrument, and that the test points cover your working range.

  7. Reference standards and traceability

    Which standards were used, and a statement that the results are traceable to the SI. "NIST traceable" on its own, or a NIST test report number, is not proof of traceability.

  8. As-found results

    How the instrument behaved when it arrived. This is the data that tells you whether your past measurements were good. Here, 300 psi is out of tolerance.

  9. As-left results

    Results after adjustment or repair. When nothing was adjusted, there is only one set of results.

  10. Measurement uncertainty

    The expanded uncertainty U and its coverage factor k (here k = 2, about 95 %). Without U and k you cannot judge the pass/fail statements or your own uncertainty budget.

  11. Statement of conformity and decision rule

    Which specification was used, which results pass or fail, and how uncertainty was taken into account. Check that the specification is your tolerance, not just the manufacturer's.

  12. Authorization and scope of the results

    The person who approved the certificate, the statement that results relate only to the item calibrated, and that the certificate may only be reproduced in full.

Example certificate for a fictitious lab and instrument. Names, numbers and the accreditation symbol are placeholders. Source: structure based on ISO/IEC 17025:2017 clause 7.8 as summarized by NIST OWM; example values by Acribi.

Two details in this example are easy to miss on a real certificate. The as-found reading at 300 psi was outside the ±0.75 psi tolerance, and the lab then adjusted the gauge. The as-left results look perfect, but the as-found row is the one that tells you whether the parts you measured with this gauge since its last calibration were measured correctly.

How to read the results table

Results tables vary from lab to lab, and the column names are a common source of confusion. Map each column to one of these:

  • Nominal (test point, set point): the target value the lab aimed for.
  • Reference value: what the reference standard measured. This is the “true” value for the comparison.
  • Instrument reading (unit under test, UUT): what your instrument indicated.
  • Error (deviation): reading minus reference.
  • Tolerance and result: the limits used for pass/fail, and the verdict.

If a certificate shows only nominal values and readings, ask the lab for the reference values. Without them you can’t recompute the error.

The sign matters. Error = indication − reference value, and a correction is the same size with the opposite sign: it is what you add to a reading to compensate for a known systematic error (VIM 2.16, 2.53; GUM 3.2.3 and B.2.23 Note 1). If a 15 mm gauge block measures 14.998 mm on your micrometer, the error is −0.002 mm and the correction is +0.002 mm.

Error versus correctionReference value 15 mm, instrument reading 14.998 mm. Error equals reading minus reference, −0.002 mm. The correction has the opposite sign, +0.002 mm.Reference 15.000Reading 14.998error = reading − reference = −0.002 mmcorrection = +0.002 mm
Error and correction have the same magnitude and opposite signs.

As found and as left. ISO/IEC 17025 requires results before and after adjustment or repair “if available” (clause 7.8.4.1). If the lab did not adjust anything, one set of results is normal. If it adjusted the instrument and only reports as-left data, you have lost the information you need for an out-of-tolerance review. Ask for as-found data when you order the calibration.

Measurement uncertainty: what k=2 means and whether it’s good enough

Every calibration result has an uncertainty. On a certificate it appears as the expanded uncertainty U, which the lab gets by multiplying the standard uncertainty by a coverage factor k. Accreditation policy asks for a coverage probability of approximately 95 % and for the certificate to state k and the coverage probability (Global ACI-TECH-1-009, formerly ILAC P14, §4.3 and §5.2). Under the usual conditions, that means k = 2 (EA-4/02, §5).

Three checks take a minute:

  1. U and k are both stated. An uncertainty without its coverage factor and coverage probability is incomplete.
  2. U has at most two significant digits, and the result is rounded to the same decimal place (TECH-1-009, §5.3; ANAB AR 2251, §4.7–4.8). “0.123456 psi” is a red flag.
  3. U is not smaller than the lab’s CMC. The calibration and measurement capability (CMC) in the lab’s scope is the best uncertainty it can achieve. The uncertainty for your instrument is normally larger, because it includes your instrument’s own contributions (TECH-1-009, §5.4–5.5).

Then the question that matters: is U small enough for your tolerance? A common way to express this is the test uncertainty ratio (TUR): the tolerance divided by the 95 % expanded uncertainty (Global ACI-TECH-1-002, formerly ILAC G8, §2.13); for a two-sided tolerance, the tolerance span divided by 2U (A2LA G136). In the example, the tolerance is ±0.75 psi and U is 0.12 psi, so TUR = 1.50 / 0.24 ≈ 6.2:1.

Measurement uncertainty compared with the toleranceLab A: U = 0.12 psi: error 0.03 psi, U 0.12 psi, TUR 6.2 to 1. Lab B: U = 0.50 psi: error 0.45 psi, U 0.5 psi, TUR 1.5 to 1.lower limitupper limitLab A: U = 0.12 psierror +0.03 ± 0.12 psi · TUR 6.2:1Lab B: U = 0.50 psierror +0.45 ± 0.5 psi · TUR 1.5:1−0.750+0.75 psi
The same tolerance with a small and a large expanded uncertainty, drawn to scale. Shaded area: result ± U (about 95 %). Source: values from the example certificate; second case is hypothetical.

With Lab B, a reading that “passes” by a comfortable margin could still be outside the tolerance. Any time the uncertainty band crosses a tolerance limit, the pass/fail call carries real risk.

“Pass” isn’t always pass: decision rules and guard bands

When a certificate says pass, in tolerance or meets specification, that is a statement of conformity. ISO/IEC 17025 requires the lab to say which results it applies to, which specification was used, and which decision rule was applied: the rule that sets how measurement uncertainty is taken into account (clauses 7.8.6 and 3.7, as described in TECH-1-002 and A2LA G136). The decision rule should be agreed with you, ideally before the work is done (clause 7.1.3).

The rule changes the verdict for results near a limit:

Simple acceptance
Guard band (w = U)
A
Pass
Pass
B
Passat risk
Fail
C
Fail
Fail
D
Fail
Fail

measured result ± expanded uncertainty U guard band (AL to TL) beyond the tolerance limit

Four results near an upper tolerance limit (TL). Under simple acceptance, case B passes even though its uncertainty band crosses the limit. With a guard band equal to U, only case A passes. Source: based on Global ACI-TECH-1-002 (formerly ILAC G8:09/2019), sections 5.1–5.2.
  • Simple acceptance (pass if the result is within tolerance): if a result sits right at the limit, the chance that the instrument is actually out of tolerance can be as high as 50 % (TECH-1-002, §5.1).
  • Guarded acceptance: the lab only passes results inside an acceptance limit set inside the tolerance. With a guard band equal to U, the specific risk of a false accept stays below about 2.5 % for a single limit with normally distributed results (TECH-1-002, §6.2, Table 1). The price is more false rejects (TECH-1-002, §6.4).
  • Non-binary rules add conditional pass and conditional fail for results inside the guard band (TECH-1-002, §5.2.3). Some labs call this zone indeterminate.

The same result can pass under one rule and fail under another. Whoever receives the statement carries the risk of the rule chosen (TECH-1-002, §1.2). Two things to check:

  • The certificate names the decision rule. “Simple acceptance; uncertainty not taken into account” is not a valid decision rule according to A2LA, because a decision rule by definition takes uncertainty into account (A2LA G136).
  • The specification used is the one you care about. A lab usually checks the manufacturer’s specification. If your process tolerance is tighter, the lab’s “pass” does not answer your question.

Accredited, traceable, or neither?

Accreditation is how you get confidence in a lab without auditing it yourself. In the US, calibration labs are accredited to ISO/IEC 17025 by six bodies that are signatories for calibration: A2LA, ANAB, IAS, NAC, NVLAP (run by NIST) and PJLA. All of them are signatories of the Global ACI Mutual Recognition Arrangement (Global ACI MRA), which took over from the ILAC MRA on January 1, 2026; you can check them in Global ACI’s signatory list. You will still see the ILAC MRA mark on many certificates during the transition (Global ACI-MRA-006, Annex 1). Three checks:

  1. The certificate carries the accreditation symbol (or a clear reference to accredited status) and the accreditation number. Only certificates that do so get the full benefit of the MRA for traceability (ILAC P10, Note 3; ANAB AR 2251, §2.1.2). Accredited labs also issue non-accredited certificates, and results outside the accreditation must be clearly identified (Global ACI-MRA-006, formerly ILAC P8, §9.8).
  2. The calibration is in the lab’s scope. Look the lab up in the accreditation body’s online directory and open its scope. The parameter and range of your calibration must be listed, and the uncertainty on your certificate cannot be better than the CMC listed there (A2LA R105; TECH-1-009, §5.5).
  3. The traceability statement points to the SI, for example “traceable to the SI through NIST” or another national metrology institute (ANAB AR 2251, §2.3.4). NIST states that it does not certify the traceability of results it did not produce, and that a NIST test report number is not proof of traceability (NIST policy and FAQ).

If you use a non-accredited provider (an instrument manufacturer, for example), you will need other evidence of competence, uncertainty and traceability. ILAC P10 lists what that evidence should cover.

When the as-found data is out of tolerance

An out-of-tolerance (OOT) as-found result means the instrument may have been giving wrong readings since its last good calibration. The certificate does not end the problem. It starts your investigation:

  • Find where the instrument was used. This is where good records pay off: which products, lots, tests or other calibrations depended on it since the last in-tolerance result.
  • Assess the impact. Compare the size of the error with the tolerances of what was measured. A 0.15 psi error matters for a ±0.2 psi test and not for a ±5 psi one.
  • Act on the equipment and on affected product, and record what you decided and why.

Quality standards expect this. ISO 9001 (clause 7.1.5) expects the same kind of review when measuring equipment that provides traceability is found unfit for purpose. For medical devices, ISO 13485 (clause 7.6) explicitly requires the assessment and a record of it. FDA inspects medical device calibration under that clause since the Quality Management System Regulation took effect on February 2, 2026 (FDA Compliance Program 7382.850; MDSAP Audit Approach, Tasks 13–14). It replaced the former 21 CFR 820.72.

For a full impact-assessment method, see our guide to out-of-tolerance calibration results.

Red flags and common myths

Red flags on the certificate itself (several are among the findings A2LA assessors report most often in accredited certificates):

  • No measurement uncertainty, or U without k and the coverage probability.
  • U with more than two significant digits, or smaller than the lab’s CMC.
  • No results at all (“calibrated, meets specifications”).
  • Only as-left data after an adjustment.
  • No issue date, or no unique certificate number on every page.
  • A traceability statement that names only NIST or quotes a NIST test number.
  • An accreditation symbol on a calibration that is not in the lab’s scope.
  • Test points that don’t cover the range you use (one point for a thermometer you use from 2 to 120 °C).
  • Model or serial number that doesn’t match your instrument.

Myths you’ll see repeated online:

Myth What the sources say
“A calibration certificate is valid for one year.” A certificate has no expiry date. ISO/IEC 17025 labs may not recommend an interval unless you agreed to it or a regulation requires it (clause 7.8.4.3). The interval is the user’s decision (ILAC G24).
“ISO 17025 requires k = 2.” Accreditation policy asks for about 95 % coverage and a stated k. k = 2 is the usual result, not a rule (TECH-1-009, §4.3 and §5.2; EA-4/02, §5).
“The uncertainty on my certificate equals the lab’s CMC.” The CMC is the best case. Your instrument’s uncertainty is normally larger (TECH-1-009, §5.4).
“The accreditation symbol means the accreditation body checked these results.” It means the lab is accredited for that work. The accreditation body does not vouch for individual results (A2LA R105, §1.7.2–1.7.3; NIST HB 150, Annex A).
“4:1 is required by ISO 17025.” It is an industry convention (A2LA G136).
“FDA requires an ISO 17025 accredited lab.” The text of 21 CFR 211.68, 211.160(b)(4) and the QMSR does not say so. Your quality system and your customers may.

Calibration certificate review checklist

Use this sequence each time a certificate arrives, before the instrument goes back into use. ILAC recommends checking calibration results against predetermined limits before you use the equipment again (ILAC G24, §4.10). NIST publishes a lab-side review checklist based on ISO/IEC 17025 clause 7.8 (SOP 1, Appendix B).

  1. Right document, right instrument

    Calibration certificate (not a conformity statement), unique number on every page, your model, serial number and asset ID.

    Mismatch: return to the lab

  2. Accreditation and scope

    Accreditation symbol and number present; calibration and range inside the lab's published scope.

  3. Traceability

    Reference standards identified; traceability statement to the SI.

  4. Results versus your tolerance

    Test points cover your working range; errors compared with your process tolerance, not only the manufacturer's.

  5. Uncertainty and ratio

    U and k stated, two significant digits at most, not below the CMC; TUR acceptable for your use.

  6. Statement of conformity

    Specification and decision rule named; results near the limits flagged.

  7. As-found results

    Any out-of-tolerance as-found result opens an impact assessment.

    Out of tolerance: start the OOT process

  8. Accept and record

    Record the acceptance (who, when), update the equipment record and label, and review the interval.

Calibration certificate review checklist

Document and identity
Accreditation and traceability
Results and uncertainty
Conformity and as-found
Close-out

What auditors expect, by standard

Requirements differ by sector. This is what each framework looks for when it comes to calibration certificates from external providers. For paid standards we describe requirements in our own words, based on the official guidance cited.

Framework What auditors look for
ISO 9001 (clause 7.1.5) Equipment calibrated or verified against traceable standards, identified and protected. The ISO/IAF auditing guidance treats calibration by a lab accredited under the ILAC MRA, now the Global ACI MRA (or a national metrology institute under the CIPM MRA), with the accreditation symbol on the certificate, as demonstrating traceability.
ISO 13485 (clause 7.6) and FDA QMSR Calibration traceable to national or international standards, controls over external calibration providers, and a recorded assessment of previous results when equipment is found out of tolerance. FDA inspects this clause under the QMSR (CP 7382.850).
IATF 16949 (clause 7.1.5.3.2) External labs accredited to ISO/IEC 17025 by an ILAC MRA signatory (now the Global ACI MRA), with the calibration in scope and the accreditation body’s mark on the certificate. A non-accredited lab (for example, the equipment manufacturer) needs documented evidence that it meets the internal-lab requirements (IATF SI 10, revised 2021; IATF FAQ 14).
FDA drug CGMP (21 CFR 211.68, 211.160(b)(4)) A written calibration program with schedules, accuracy limits and remedial action; instruments that don’t meet specifications must not be used.
ISO/IEC 17025 (labs as customers) Traceability through accredited providers or NMIs (ILAC P10); calibration results recorded in the equipment records and checked against acceptance criteria.

Does a calibration certificate expire?

No. A certificate reports what the instrument did on the calibration date. When to calibrate again is your decision, based on how the instrument drifts, how much risk an undetected error carries and how heavily it is used (ILAC G24). Labs accredited to ISO/IEC 17025 may not put a recommended interval on the certificate or label unless you agreed to it or a regulation requires it (clause 7.8.4.3). If your certificates show a due date, it is usually because your purchase order asked for one.

See our guide to setting calibration intervals and the TUR calculator.

FAQ

Who can issue a calibration certificate?

Any calibration provider can issue one, including the instrument manufacturer and your own in-house lab. What changes is how much evidence it carries. A certificate from a lab accredited to ISO/IEC 17025, with the accreditation symbol and the calibration inside the lab's scope, is the form that ILAC, A2LA, ANAB and IATF treat as demonstrating traceability. Other providers can be acceptable, but you then need extra evidence about the provider's competence, uncertainty and traceability.

Can we calibrate in-house and write our own certificate?

Yes. Many companies calibrate their own gauges. The record should still show what was measured, against which traceable reference standards, with what uncertainty, by whom and with which result, and the reference standards themselves must be calibrated by a competent provider. Automotive suppliers have extra requirements for internal laboratories under IATF 16949.

My new instrument arrived without a calibration certificate. Is that normal?

Often, yes. Some manufacturers offer a calibration certificate as a separate option when you order. If you need traceable results from day one, order the instrument with a certificate that includes data, or have it calibrated before first use.

What is a long-form calibration certificate?

It is a manufacturer or lab term for a certificate that includes the measurement data (test points, readings, errors and usually uncertainty), as opposed to a short statement that the instrument met its specifications. For review, audits and out-of-tolerance investigations you need the data.

Is a certificate of conformity the same as a calibration certificate?

No. A certificate of conformity (or of accuracy, or inspection) states that a product met a specification. A calibration certificate reports measured results against traceable reference standards, with their uncertainty. Some 'certificates of accuracy' even state that they are not calibration certificates.

Is 'NIST traceable' enough for ISO 9001?

The words alone are not evidence. ISO 9001 auditing guidance looks for calibration by a lab accredited under the ILAC MRA, now the Global ACI MRA (or a national metrology institute under the CIPM MRA), shown by the accreditation symbol on the certificate. Traceability is a property of a result, to the SI, through a documented chain of calibrations with stated uncertainties.

Where do I find the calibration certificate for my NI, Fluke or Keysight instrument?

Start with the manufacturer's support or calibration services site, or contact its service center with the model and serial number. Once you have it, review it like any other certificate: a manufacturer's certificate is not automatically accredited.

How long is a calibration certificate valid?

A certificate does not expire. It reports results on the calibration date. When to recalibrate is your decision, based on risk, drift history and use. ISO/IEC 17025 labs may not recommend an interval on the certificate or label unless you asked for it or a regulation requires it.

Sources

  1. NIST Office of Weights and Measures. ISO/IEC 17025 Crosswalk – Reporting the Results (clause 7.8). 2018 — Free summary of ISO/IEC 17025:2017 clause 7.8 (certificate content, 7.8.4.1, 7.8.4.3, 7.8.6)
  2. NIST (NISTIR 6969). SOP 1, Calibration Certificate Preparation, Appendices A–C (example certificate and review checklist). 2019
  3. JCGM / BIPM. International Vocabulary of Metrology (VIM), JCGM 200:2012. 2012 — 2.39 calibration, 2.41 metrological traceability, 3.11 adjustment
  4. JCGM / BIPM. Guide to the Expression of Uncertainty in Measurement (GUM), JCGM 100:2008. 2008 — 3.2.3 and B.2.23 (correction)
  5. Global Accreditation Cooperation Inc. (Global ACI). Global ACI-TECH-1-009 (M), Policy for Measurement Uncertainty in Calibration (formerly ILAC P14:09/2020). 2026 — v1.0, June 26, 2026; same numbering as ILAC P14 (§4.3, §5.2–5.5)
  6. European Accreditation. EA-4/02 M:2022, Evaluation of the Uncertainty of Measurement in Calibration. 2022
  7. Global Accreditation Cooperation Inc. (Global ACI). Global ACI-TECH-1-002 (G), Guidelines on Decision Rules and Statements of Conformity (formerly ILAC G8:09/2019). 2026 — v1.0, June 26, 2026; same content as ILAC G8, sections renumbered +1 (G8 §5.2 is now §6.2)
  8. A2LA. G136, Guidance on Decision Rules in Calibration. 2022
  9. ILAC (still current under Global ACI). ILAC P10:07/2020, Policy on Metrological Traceability of Measurement Results. 2020
  10. Global Accreditation Cooperation Inc. (Global ACI). Global ACI MRA Signatory List. 2026 — List dated 2026-09-25: six US signatories for calibration (A2LA, ANAB, IAS, NAC, NVLAP, PJLA)
  11. Global Accreditation Cooperation Inc. (Global ACI). Global ACI-MRA-006, Use of the MRA Mark and Claims of Accreditation Status (formerly ILAC P8 and R7). 2025 — v1.0, implemented January 1, 2026; §9.8 results outside the scope; Annex 1 temporary use of the ILAC MRA mark
  12. Global Accreditation Cooperation Inc. (Global ACI). Global ACI, IAF and ILAC Document Cross-Reference Table, v14.0. 2026 — P14 → TECH-1-009, G8 → TECH-1-002, P8/R7 → MRA-006; P10 and G24 still current
  13. A2LA. R105, Requirements When Making Reference to A2LA Accredited Status. 2025
  14. A2LA. Common Calibration Certificate Findings. 2024
  15. ANAB. AR 2251, Accreditation Requirements: ISO/IEC 17025 Calibration Laboratories. 2026 — §2.1.2 accredited certificates for traceability; §4.7–4.8 rounding
  16. NIST. NIST Policy on Metrological Traceability. 2024
  17. NIST. Metrological Traceability: Frequently Asked Questions and NIST Policy
  18. NIST NVLAP. NIST HB 150, NVLAP Procedures and General Requirements (2020, update 1). 2020
  19. ILAC / OIML. ILAC G24:2022 / OIML D 10:2022, Guidelines for the determination of recalibration intervals. 2022
  20. ISO/TC 176 and IAF. ISO 9001 Auditing Practices Group, Guidance on Measurement Traceability. 2016
  21. MDSAP (hosted by FDA). MDSAP AU P0002.009, Audit Approach (Tasks 13–15, monitoring and measuring equipment). 2024 — Task 13 p. 107, Task 14 p. 108, Task 15 p. 109
  22. FDA. Compliance Program 7382.850, Inspection of Medical Device Manufacturers (Attachment A). 2026
  23. eCFR. 21 CFR Part 211 (§211.68, §211.160(b)(4)). 2026
  24. eCFR. 21 CFR Part 820, Quality Management System Regulation. 2026
  25. IATF. IATF 16949:2016 Sanctioned Interpretations (SI 10, clause 7.1.5.3.2). 2025
  26. IATF. IATF 16949:2016 Frequently Asked Questions (FAQ 14). 2026