Calibration records: what to keep, how long, and a free template

What a calibration record must contain, how it differs from a certificate and a master list, how long to keep it under each standard, with an annotated example and a printable template.

Updated 23 primary sourcesHow we researchCite

On this page
  1. Calibration record, certificate, equipment record, master list: what’s the difference
  2. What a calibration record must contain
  3. Example: an equipment record with its calibration history
  4. Free calibration record template
  5. How long to keep calibration records
  6. Paper, spreadsheet or software?
  7. Common audit findings on calibration records
  8. Common myths
  9. Related guides
  10. FAQ
  11. Sources

A calibration record is your evidence that a specific instrument was calibrated or verified: which item, when, by whom, against which traceable standard, with what results (as found and as left), against which tolerance, what you decided, and when it is due next. Keep it for as long as the strictest rule that applies to you requires: ISO 9001 and ISO/IEC 17025 set no number of years, while medical device makers must keep records for at least the lifetime of the device and never less than two years from product release (ISO 13485 clause 4.2.5, enforced by FDA through the QMSR). This guide covers what to record, an annotated example, a printable template, retention by standard, and the findings auditors write most often.

Calibration record, certificate, equipment record, master list: what’s the difference

People use these words loosely, and no standard defines a “calibration log”. These are the working definitions we use in our guides. They follow how ISO/IEC 17025 separates the report a lab issues (clause 7.8) from the equipment records a user keeps (clause 6.4.13, as summarized by PJLA and A2LA G149).

Document Who creates it What it holds Scope
Calibration certificate (or calibration report) Whoever calibrates: an external lab or your in-house lab Results, uncertainty, traceability and any conformity statement, on one date One calibration
Calibration record You The result of each calibration or verification, the reference standard, your acceptance decision, any out-of-tolerance link, the next due date One calibration or check
Equipment record (gauge history card) You Identity, location, requirement, interval, and the history of every calibration, repair and status change One instrument, over its life
Master list (equipment register, calibration log or tracker) You One row per item: ID, location, interval, last and next due date, status All instruments

In a small program, the calibration records are simply the rows of the equipment record’s history, as in the example below. Software usually stores them that way too. What matters is that each piece exists and links to the others through the asset ID.

What a calibration record must contain

No single free source lists every field. The clearest public minimum came from FDA’s former device rule, which required calibration records to show the equipment identification, the calibration dates, the person who performed each calibration and the next calibration date, displayed on or near the equipment or readily available (former 21 CFR 820.72(b)(2), in force until February 1, 2026). ISO/IEC 17025 gives the most complete list for equipment records (clause 6.4.13), and its technical-record clause adds the date and identity of the people responsible for each activity, the results, and enough information to identify the factors that affect them (clause 7.5.1, as summarized by SADCAS and NATA). Put together:

Field Why it is there Basis
Asset ID, description, make, model, serial, firmware Proves which item the results belong to Former 820.72(b)(2); ISO/IEC 17025 6.4.13
Location and owner Recall, and tracing uses after a failure ISO/IEC 17025 6.4.13
Required tolerance (acceptance criteria) and its basis Without it, “pass” has no meaning ISO/IEC 17025 6.4.13; 21 CFR 211.160(b)(4) (accuracy and precision limits)
Procedure or method Shows the calibration was done the defined way 21 CFR 211.160(b)(4); former 820.72(b)
Calibration date and who performed it The minimum FDA used to require; accountability Former 820.72(b)(2); ISO/IEC 17025 7.5.1
Reference standard ID and its certificate Traceability for in-house work ISO 9001 7.1.5.2; former 820.72(b)(1)
As-found and as-left results per test point, with uncertainty Evidence of past measurements; drift data for intervals NASA-HDBK-8739.19-2 §10.2.7; ISO/IEC 17025 6.4.13
Adjustment made (yes/no, what) Separates the two sets of results ISO/IEC 17025 6.4.13
Decision, and the decision rule for pass/fail Who accepted the item and on what basis Global ACI-TECH-1-002, formerly ILAC G8 (decision rules)
Out-of-tolerance assessment reference Required when equipment is found not to conform ISO 13485 7.6 (MDSAP Tasks 13–14)
Next due date or interval What the recall schedule runs on Former 820.72(b)(2); ISO/IEC 17025 6.4.13
Reviewed and approved by, date Second look before release ISO/IEC 17025 7.5.1; 21 CFR 11.50 for FDA e-signatures
Environmental conditions, where they affect the result Lets you judge the result later Good practice (required on lab certificates by ISO/IEC 17025 7.8.4.1)

Each framework states the core differently:

  • ISO 9001 requires documented evidence that monitoring and measuring resources are fit for purpose (clause 7.1.5.1) and, when no national or international standards exist, a record of the basis used for calibration or verification (clause 7.1.5.2) (ISO/TC 176 N1286). The 2026 edition does not change clause 7.1.5 in substance, as far as the official change summaries show (Global ACI TECH-3-TR, §3.1). Auditing guidance expects records of calibration, uncertainty and traceability to match how the equipment is used (ISO 9001 APG).
  • ISO 13485 (and so FDA’s QMSR since February 2, 2026) requires records of calibration and verification results, and a recorded assessment of the validity of earlier results when equipment doesn’t conform (clause 7.6; MDSAP Tasks 13–14, pp. 107–108).
  • US drug CGMP requires written records of the calibration, inspection or checking of automatic, mechanical and electronic equipment (21 CFR 211.68(a)), and complete records of the periodic calibration of laboratory instruments, apparatus, gauges and recording devices (21 CFR 211.194(d)).
  • IATF 16949 has its own clause on calibration and verification records (7.1.5.2.1). We don’t list its content here because we couldn’t verify it in a free primary source.

Example: an equipment record with its calibration history

The example below is the record a manufacturer might keep for the digital pressure gauge from our example calibration certificate. It continues the same story: the gauge came back from the lab out of tolerance at 300 psi, which opened the out-of-tolerance assessment OOT-2026-007. Everything in it is fictitious. Select a number to jump to its note.

Equipment recordSample Manufacturing Co. · Measuring equipment control
1Asset IDPG-014

Identification

Description
Digital pressure gauge
Make / model
Example Instruments DPG-300
Serial no.
21-0456
Firmware
2.1 (example)
Range
0 to 300 psi · 0.01 psi
In service since
2022-03-08
2

Location and owner

Location
Final test bench 3 (valve line)
Owner
Production (valve line)
3

Requirement

Tolerance
± 0.75 psi
Basis
Process requirement; same as the manufacturer spec (±0.25 % FS)
Risk class
A (accepts product)
4

Method and provider

External calibration, ISO/IEC 17025 accredited lab, 5 points ascending

5

Interval and status

Interval
6 months · Shortened from 12 months by OOT-2026-007 (2026-10-02)
Intermediate check
Monthly at 300 psi against PG-022
Next due
2027-03-16

Calibration history (newest last)

6Date · by · certificate7Standard / traceability8As found · As left · U (k=2)9Decision10Reviewed
2025-09-15
Calibration (external)
Example Calibration Laboratory, Inc.
Cert. ECL-2025-03920
Lab reference, traceable to SI (on certificate)In tolerance
No adjustment
U = 0.12 psi
AcceptedR. Patel · 2025-09-19
2026-09-16
Calibration (external)
Example Calibration Laboratory, Inc.
Cert. ECL-2026-04817
ECL-STD-031 (cert. RC-2026-1123)OUT at 300 psi (+0.90)
Zero and span adjusted; in tolerance (max 0.03)
U = 0.12 psi
Accepted after adjustment; OOT-2026-007 openedR. Patel · 2026-09-18

Certificates are attached to this record as PDF files, named by asset ID and certificate number. Errors in psi; error = reading − reference.

11

Status changes and repairs

  • 2026-09-18 Tagged out of service; OOT-2026-007 opened
  • 2026-09-22 Returned to service, new label applied
  • 2026-10-02 OOT-2026-007 closed; interval 12 → 6 months; monthly check added
12

Change log

2026-10-02 · R. Patel
Interval changed 12 → 6 months (was: 12 months). Reason: OOT-2026-007.

Record owner: R. Patel, metrology coordinator · Fictitious example by acribi.com. Company, people and results are invented.

  1. Unique ID and identity

    One ID per item ties the instrument, its status label, its certificates and every record together. ISO 9001 asks for equipment to be identified so its calibration status can be determined (clause 7.1.5.2). ISO/IEC 17025 equipment records start with the identity, including software and firmware version, plus manufacturer, model and serial number (clause 6.4.13).

  2. Location and owner

    Where the item is and who uses it. You need it to recall the item when it is due and, if it fails, to trace what it measured. ISO/IEC 17025 lists current location among the equipment-record fields (clause 6.4.13).

  3. Your requirement

    The tolerance the item must meet over the range you use, and where it comes from. ISO/IEC 17025 puts acceptance criteria in the equipment record (clause 6.4.13). Without this field nobody can say whether a result passes. The risk class tells people how much control the item gets (see our program guide).

  4. Method and provider

    Which procedure is used and who calibrates. The drug CGMP rule requires specific directions in the written calibration program (21 CFR 211.160(b)(4)). For in-house work, quote the procedure number and revision.

  5. Interval, intermediate checks and next due date

    ISO/IEC 17025 asks for the due date or the interval (clause 6.4.13), and the former FDA device rule asked for the date of the next calibration (former 21 CFR 820.72(b)(2)). Record why the interval is what it is, so a change is defensible (see calibration intervals).

  6. When, by whom, which certificate

    Date of each calibration and who performed it: the minimum list in the former 21 CFR 820.72(b)(2), and in ISO/IEC 17025 technical records the date and identity of the people responsible (clause 7.5.1). The certificate number links the row to the lab's document.

  7. Reference standard and traceability

    For in-house calibrations, the ID of the reference standard you used and its own certificate. When measurement traceability is required, ISO 9001 asks for calibration or verification against standards traceable to national or international standards, or a record of the basis used when none exist (clause 7.1.5.2). For external calibrations, the certificate carries this, and the record points to it.

  8. As-found and as-left results, with uncertainty

    NASA asks for the status of every parameter, in or out of tolerance, with as-found and as-left data, and recommends keeping the measured values with their uncertainty so drift can be analyzed later (NASA-HDBK-8739.19-2, §10.2.7). "Pass" alone throws away the data you need to set intervals.

  9. Your decision

    Accepted, accepted with limits, or removed from service, and a link to the out-of-tolerance assessment if the as-found result failed. ISO 13485 requires you to assess and record the validity of earlier results when equipment is found not to conform (clause 7.6; MDSAP Tasks 13–14). See our out-of-tolerance guide.

  10. Review and sign-off

    Who reviewed the results and when. ISO/IEC 17025 expects technical records to show the date and the identity of the people responsible for each activity (clause 7.5.1, as summarized by SADCAS). Recording who reviewed the results is how you show that step. In an FDA-regulated electronic system, the signature must show the printed name, date and time, and its meaning, such as review or approval (21 CFR 11.50).

  11. Status changes, maintenance and repairs

    Out-of-service tags, returns to service, damage, malfunctions, modifications and repairs. ISO/IEC 17025 lists maintenance and damage or repair details among the equipment-record fields (clause 6.4.13).

  12. Changes that leave a trail

    When an entry changes, keep the original visible, with the date, what changed and who changed it (ISO/IEC 17025 clause 7.5.2, as summarized by SADCAS). On paper that is a single line through the old value; in software it is the audit trail.

Example equipment record for a fictitious company, with two calibration records in its history. Names, numbers and results are invented. Source: fields based on ISO/IEC 17025:2017 clauses 6.4.13 and 7.5 (as summarized by PJLA, A2LA, SADCAS) and the former 21 CFR 820.72(b)(2); example values by Acribi.

Three details make this record hold up in an audit. The 2026 row keeps the as-found failure next to the as-left result, so the adjustment does not hide the problem. The decision points to the out-of-tolerance report, so an auditor can follow the trail. And the interval change has a reason and a date in the change log, so nobody has to guess why a 12-month gauge is now on 6 months.

Free calibration record template

Use this form for one calibration or verification, whether in-house or external. For external calibrations, fill in the header, copy the key results from the certificate, and record your decision. Print it, or copy the fields into your own spreadsheet or system. It is free, with no sign-up.

Calibration record

Equipment
Asset ID
Description
Make / model
Serial no.
Firmware / software version
Location / owner
Range / resolution
Risk class
Requirement and method
Required tolerance and where it comes from
Procedure / method and revision
Calibration date
Calibrated byIn-house (name):External lab (name, accreditation no.):
Certificate or record no.
Reference standard ID(s)
Standard cert. no. / due date
Environmental conditions
Results
Test pointReference valueAs-found readingAs-found errorAs-left readingAs-left errorTolerance ±Pass / fail
Units
Expanded uncertainty U (k = __)
Adjusted or repairedNoYes (what):
Decision rule for pass / fail
Decision and close-out
As foundIn toleranceOut of tolerance: OOT report no.:
DecisionAccepted, back in serviceAccepted with limits (labelled)Removed from service
Next due date
Interval (change and reason)
Status label applied
Certificate attached to record
Performed by / date
Reviewed and approved by / date

Master list columns. For the list of all your equipment, one row per item is enough: asset ID, description, serial number, location, owner, risk class, tolerance, interval, last calibration date and certificate, next due date and status. Our program guide shows a filled-in row for this same gauge. A due-date tracker without results is a useful master list, but it is not a calibration record.

How long to keep calibration records

Most sources answer “it depends”. Here is what it depends on, by framework. For paid standards we describe requirements in our own words, based on the free sources cited.

Framework What the record must show How long to keep it
ISO 9001 (2015 and 2026) Evidence that equipment is fit for purpose (7.1.5.1); basis for calibration when no traceable standard exists (7.1.5.2) No fixed period. You define and control retention (clause 7.5.3)
ISO/IEC 17025 (labs) Equipment records (6.4.13); technical records with dates, people and results (7.5.1); traceable amendments (7.5.2) No fixed period. Retain records for a time consistent with your contractual obligations (8.4.2). Your accreditation body may set a minimum: NATA in Australia, for example, requires at least 4 years, and for equipment records at least the longest recalibration interval, whichever is longer
ISO 13485 and FDA QMSR (since Feb. 2, 2026) Calibration and verification results; recorded assessment of earlier results when equipment doesn’t conform (7.6) At least the lifetime of the device as you define it, and never less than 2 years from product release (ISO 13485 4.2.5, required through 21 CFR 820.10, together with 820.35). National rules elsewhere can require longer
21 CFR 211 (drug CGMP) Written records of the calibration program (211.68(a)); complete records of laboratory instrument calibration (211.194(d)) No period specific to calibration records. Records associated with a batch: at least 1 year after the batch expiration date (211.180(a))
21 CFR Part 11 (electronic FDA records) Applies only when an FDA-required record is kept electronically Records must stay protected and retrievable for the whole retention period set by the underlying rule, and audit trails are kept as long as the record (11.10(c), (e))
IATF 16949 Has a dedicated clause on calibration and verification records (7.1.5.2.1); we could not verify its content in a free primary source We found no published period for calibration records. Customer-specific requirements may set one: GM, for example, points to its own records specification under 7.5.3.2.1

Sources: ISO 9001 via ISO/TC 176 N1286; ISO/IEC 17025 via PJLA (8.4.2) and NATA (§8.1.1); ISO 13485 via MDSAP Task 8 (p. 26) and FDA’s QMSR final rule; 21 CFR 211.180; 21 CFR 11.10; GM IATF 16949 customer-specific requirements.

What to take from the table:

  • If you make medical devices, the device lifetime sets the floor, and a calibration record can matter for as long as any device measured with that instrument is in use. MDSAP auditors check that records are kept for at least the device lifetime and not less than two years from product release (MDSAP Task 8, p. 26). FDA inspects this under ISO 13485 clause 4.2.5 together with 21 CFR 820.35 (FDA CP 7382.850, Attachment A).
  • If you make drugs, Part 211 sets no calibration-specific period. Our recommendation: when a calibration supports the release of a batch, keep the record at least as long as that batch record. That is our reading, not a stated rule.
  • If you follow ISO 9001 or ISO/IEC 17025 only, write the period into your procedure. Base it on the longest of product life, warranty, contract and legal liability. Even a large metrology program sets its own schedule: NASA’s metrology standard simply refers to the agency’s records retention schedules (NASA-STD-8739.12A, §4.6.1).
  • Keep the history while the item is in service. Interval reviews use past calibration results (ILAC G24), so discarding old records also discards the evidence for your intervals. That is our recommendation, not a requirement.

Paper, spreadsheet or software?

No standard requires a particular medium. The same rules apply to all of them: records must be legible, protected against damage and loss, backed up, retrievable and kept for the defined period (ISO/IEC 17025 clauses 8.4.1–8.4.2, as summarized by PJLA). Corrections must keep the original visible, with who changed what and when (clause 7.5.2, as summarized by SADCAS). How each option handles that:

Paper Controlled spreadsheet Calibration software
Fits Few items, one site Small to medium inventory, one owner Many items, users or sites; regulated records
Due-date recall Manual review of the list Formulas and filters Automatic notices
Traceable corrections Single line through the old entry, initials, date Needs a change log or version history; overwriting a cell leaves no trace by default Audit trail, if the system keeps one
Protection and backup Copies; fire and loss risk Protected cells, controlled access, scheduled backups Access control and backups (yours or the vendor’s)
Linking certificates Filed by asset ID Links to PDF files by asset ID Usually attached to the item
FDA Part 11 Does not apply to paper records Applies if it holds FDA-required records Applies if it holds FDA-required records

Whatever you choose, three regulated situations add requirements:

  • FDA-required records kept electronically. Part 11 applies to records in electronic form created, modified, maintained or archived under FDA record requirements (21 CFR 11.1(b)). For closed systems it asks for validation, accurate copies, protection throughout the retention period, limited access, and secure, time-stamped audit trails that don’t hide earlier entries (21 CFR 11.10). Electronic signatures must show the signer’s name, date and time and meaning, and be linked to their records (21 CFR 11.50, 11.70).
  • FDA’s enforcement discretion. In its 2003 guidance, FDA said it intends to exercise enforcement discretion over the Part 11 requirements for validation, audit trails, record copying and record retention, and over systems in operation before August 20, 1997 that meet certain conditions. It still enforces controls such as limited system access, operational and authority checks, training, accountability and electronic signatures. The record requirements of the underlying rules still apply in full, and FDA recommends basing your decisions on a documented risk assessment (FDA Part 11 guidance, §III). In drug plants, 21 CFR 211.68(b) separately requires controls so that only authorized people change records, plus checks and secure backups.
  • Medical device quality-system software. Software used in the quality system, including off-the-shelf calibration software, must be validated for its intended use (ISO 13485 4.1.6 and 7.6; MDSAP Task 15, p. 109).

Part 11 does not apply to records kept only for ISO 9001, ISO/IEC 17025 or IATF 16949. Our comparison of calibration management software covers what to check before buying.

Common audit findings on calibration records

Many calibration findings are about the record, not the instrument. These are the record problems that appear in FDA observation data and that follow from the requirements above:

  • Calibration done but not documented, or written records not kept (FDA observation codes under 820.72(a) and 211.68(a)).
  • Incomplete laboratory calibration records, for example without results, standard or acceptance limits (21 CFR 211.194(d)).
  • No reference standard or acceptance criterion in in-house records, so traceability and the pass decision can’t be shown (ISO 9001 7.1.5.2; ISO/IEC 17025 6.4.13).
  • Certificate filed but never reviewed, so an out-of-tolerance as-found result went unnoticed. ILAC recommends checking results against predetermined limits before equipment goes back into use (ILAC G24, §4.10).
  • Out-of-tolerance result with no documented assessment of earlier results (ISO 13485 7.6; former 820.72(b)).
  • Records that can’t be found, are illegible or were not kept long enough (former 21 CFR 820.180, including its retention paragraph (b)).
  • Uncontrolled computer systems or spreadsheets: anyone can change records, or backups are incomplete (21 CFR 211.68(b)).

FDA publishes the observations its investigators write on Form 483. These codes relate directly to calibration records:

FDA code Regulation Observation (paraphrased) FY2024 FY2025
3235 820.72(a) Calibration or maintenance activities not documented 6 8
3239 820.72(b) Out-of-tolerance evaluation not documented 3 —
3149 820.180 Records not available, illegible or poorly stored 2 5
3328 820.180(b) Records not kept for the required period 1 1
1261 211.68(a) Calibration records not kept 2 5
2034 211.194(d) Laboratory instrument calibration records incomplete 4 2
1263 211.68(b) Inadequate controls over computer systems (changes by authorized people only) 60 87
4305 211.68(b) Backup not accurate, complete or secure 6 8

FDA warns that these data sets are not a complete list of observations: they only include 483s generated with its electronic tools. Fiscal year 2025 ended on September 30, 2025, before the QMSR took effect, so device observations still use the former section numbers (FDA Inspection Observations). Code 1263 covers computer systems in general, not only calibration.

Common myths

Myth What the sources say
“The certificate is the calibration record.” The certificate is the lab’s report. The record adds your tolerance, decision, out-of-tolerance link and due date (ISO/IEC 17025 6.4.13).
“21 CFR 820.72 requires calibration records on the equipment.” 820.72 no longer exists since February 2, 2026. Even then, records could be on or near the equipment or readily available (former 820.72(b)(2)).
“ISO 9001 requires records to be kept for X years.” ISO 9001 sets no period; you define it (clause 7.5.3; N1286).
“Records must be on paper and signed in ink.” No standard we checked requires paper or ink. Corrections must leave the original visible (ISO/IEC 17025 7.5.2), and some accreditation bodies require records to be indelible as far as practicable (NATA).
“All calibration software must be Part 11 compliant.” Only for FDA-required records kept electronically (21 CFR 11.1(b)), and FDA applies enforcement discretion to some controls (2003 guidance).
“An in-house record doesn’t need to name the standard used.” Without the standard, you can’t show traceability (ISO 9001 7.1.5.2).

FAQ

What is the difference between a calibration record and a calibration certificate?

The certificate is issued by whoever performed the calibration and reports what was measured, against which standards, with what uncertainty and result. The calibration record is your own evidence: it links that certificate (or your in-house results) to the instrument, your tolerance, your decision, any out-of-tolerance assessment and the next due date. A certificate filed in a folder is not a complete record.

Is a calibration report the same as a calibration record?

Usually not. ISO/IEC 17025 calls the document a laboratory issues a report, and a calibration certificate is the report of a calibration. The record is what you keep about the instrument and what you did with that report. In-house calibrations often produce a single document that serves as both.

How long should calibration records be kept?

As long as the strictest rule that applies to you. ISO 9001 and ISO/IEC 17025 set no number of years: you define the period, taking contracts and laws into account. Medical device manufacturers must keep records for at least the lifetime of the device as they define it, and never less than two years from product release (ISO 13485 clause 4.2.5, enforced by FDA through the QMSR). The US drug CGMP rule sets no period specific to calibration records.

Does ISO 9001 say how long to keep calibration records?

No. ISO 9001 requires you to retain evidence that monitoring and measuring resources are fit for purpose (clause 7.1.5.1) and to control the retention of documented information (clause 7.5.3), but it does not set a period. You decide it, and an auditor will expect you to follow what you decided.

Can calibration records be kept electronically?

Yes. No standard requires paper. Whatever the medium, records must be legible, protected, backed up and retrievable, and corrections must keep the original entry visible (ISO/IEC 17025 clauses 7.5.2 and 8.4.2). If FDA regulations require the record and you keep it electronically, 21 CFR Part 11 also applies.

Does my calibration software have to be Part 11 compliant?

Only if it holds records that FDA regulations require you to keep, such as calibration records under 21 CFR 211 or the QMSR. Part 11 does not apply to ISO 9001, ISO/IEC 17025 or IATF 16949 records on their own. Even within its scope, FDA has said since 2003 that it intends to exercise enforcement discretion over some Part 11 controls, such as audit trails and validation, while the underlying record requirements still apply.

Do I need to write my own certificate for equipment calibrated by an external lab?

No. You don't need a second certificate. You need a record that links the lab's certificate to the instrument and shows your review: the results against your tolerance, the decision, any out-of-tolerance assessment and the next due date.

One record per instrument, or per location?

Per instrument, identified by a unique ID. The location is a field in the record. If you swap a thermometer at a fixed point, the new one gets its own record, and the location history shows which instrument measured there and when, which is what you need if one of them fails calibration.

How do I record an instrument with a dedicated probe?

If the readout and probe are always calibrated and used together, record them as one item and list both serial numbers. If they are calibrated separately or swapped between readouts, give each its own ID and record. Either choice works if it is written down and applied consistently.

Is a spreadsheet good enough for calibration records?

It can be, for a small inventory, if it is controlled: one owner, protected cells, a change log, regular backups and certificates filed by asset ID. Spreadsheets become hard to control with many users, sites or items, and they don't keep an audit trail of changes on their own.

Sources

  1. eCFR. 21 CFR Part 820, Quality Management System Regulation (§820.10, §820.35). 2026
  2. eCFR (historical version). 21 CFR Part 820 as in force on January 15, 2026 (former §820.72 and §820.180). 2026 — Superseded on February 2, 2026; cited as history
  3. Federal Register. Medical Devices; Quality System Regulation Amendments (final rule), 89 FR 7496. 2024 — Discussion of §820.35, pp. 7513–7514
  4. MDSAP (hosted by FDA). MDSAP AU P0002.009, Audit Approach (Task 8, document and record controls; Tasks 13–15, monitoring and measuring equipment). 2024 — p. 26 and pp. 107–109
  5. FDA. Compliance Program 7382.850, Inspection of Medical Device Manufacturers (Attachment A). 2026
  6. eCFR. 21 CFR Part 211 (§211.68, §211.160(b)(4), §211.180, §211.194(d)). 2026
  7. eCFR. 21 CFR Part 11, Electronic Records; Electronic Signatures (§11.1, §11.10, §11.50, §11.70). 2026
  8. FDA. Part 11, Electronic Records; Electronic Signatures — Scope and Application (guidance for industry). 2003
  9. FDA. Inspection Observations (FY2024 and FY2025 data sets). 2025
  10. Perry Johnson Laboratory Accreditation (PJLA). Requirements in Section 6.4 Equipment (webinar slides on ISO/IEC 17025:2017, clause 6.4.13). 2023
  11. A2LA. G149, Guidance for Application of P102 In-House Calibrations. 2026 — Summarizes ISO/IEC 17025 clause 6.4.13
  12. Perry Johnson Laboratory Accreditation (PJLA). A look at the ISO/IEC 17025:2017 Requirements Concerning Document Control and Control of Records (webinar slides, clauses 8.3–8.4). 2024
  13. SADCAS. F 60(c) (fr), Évaluation verticale à l'ISO/CEI 17025:2017, version 3 (items 7.5.1 and 7.5.2). 2019
  14. NATA. General Accreditation Criteria: ISO/IEC 17025 Standard Application Document (§7.5.1, §8.1.1). 2024
  15. ANAB. AR 2251, Accreditation Requirements: ISO/IEC 17025 Calibration Laboratories (§1.1.4). 2026
  16. ISO/TC 176/SC 2 (copy hosted by IAQG). ISO/TC 176/SC 2/N1286, Guidance on the requirements for Documented Information of ISO 9001:2015. 2016
  17. ISO/TC 176 and IAF. ISO 9001 Auditing Practices Group, Guidance on Monitoring and measuring resources. 2016
  18. Global Accreditation Cooperation (Global ACI). Global ACI-TECH-3-TR 2029-09-30 (M), Transition Requirements for ISO 9001:2026, v1.0. 2026 — §3.1
  19. General Motors (hosted by IATF). IATF 16949 Customer Specific Requirements, General Motors (clauses 7.1.5.2.1 and 7.5.3.2.1). 2025
  20. NASA. NASA-HDBK-8739.19-2, Measuring and Test Equipment Specifications (§10.2.7 Calibration Records). 2010
  21. NASA. NASA-STD-8739.12A, Metrology and Calibration (§4.6.1). 2024
  22. ILAC / OIML. ILAC G24:2022 / OIML D 10:2022, Guidelines for the determination of recalibration intervals of measuring equipment. 2022
  23. Global Accreditation Cooperation Inc. (Global ACI). Global ACI-TECH-1-002 (G), Guidelines on Decision Rules and Statements of Conformity (formerly ILAC G8:09/2019). 2026 — Same content as ILAC G8; sections renumbered +1