Pressure gauge: types, accuracy grades and calibration standards
An overview of pressure gauge calibration: gauge types, ASME B40.100 grades and EN 837-1 classes, the free DKD and EURAMET guidelines, reference standards, intervals and what NIST-traceable means for a gauge. The step-by-step method is in our separate procedure.
- Calibration procedureHow pressure gauge calibration works: what you can check yourself, how to choose between a deadweight tester and a comparator, test points up and down, hysteresis, mounting position, head and gravity corrections, accuracy grades, uncertainty and records.Read the procedure
- Is your reference good enough?Enter the tolerance and your calibration uncertainty to get the test uncertainty ratio.Open the TUR calculator
Pressure gauge calibration compares a gauge’s readings with a more accurate, traceable pressure reference, a deadweight tester or a calibrated reference gauge on a comparator, at several points across its range, first with pressure rising and then falling, and records the deviation and its uncertainty at each point (DKD-R 6-1, §3 and §7; VIM 2.39). Resetting the pointer to zero is an adjustment, not a calibration (VIM 3.11), and a gauge that reads zero correctly can still be off at mid-scale.
This page is the overview. For the checks you can do yourself, deadweight tester vs comparator, test points, hysteresis and a worked example, see pressure gauge calibration step by step.
Types of pressure gauges
| Type | What to know for calibration |
|---|---|
| Bourdon tube dial gauge | The most common; rated by accuracy grade or class as a percentage of span. DKD’s recommended interval depends on the class (DKD-R 6-1, App. F) |
| Electronic (digital) gauges and transmitters | Calibrated the same way; need their warm-up time. Also used as references, but less stable over time than a deadweight tester (DKD-R 6-1, §3 and App. F) |
| Compound and vacuum gauges | Need test points below zero: at least two when calibrated on both sides of zero (DKD-R 6-1, Table 1, notes) |
| Absolute gauges (including barometers) | Do not read zero at atmospheric pressure, so a zero check does not apply; they need an absolute reference (DKD-R 6-1, §8.6.2) |
| Low-pressure differential gauges (Magnehelic) | Read inches of water; most ranges ±2 % of full scale (Dwyer, Series 2000 data sheet). Mounting position matters |
Accuracy grades and classes
In the US, dial gauges are graded to ASME B40.100; in Europe, to EN 837-1 classes. In both, the percentage is of the span, not of the reading. ASME B40.100 (2022) is a paid standard; each grade below is confirmed by at least two of these manufacturer and distributor sources (Ashcroft; WIKA data sheets; Boshart). Lower grades also exist; check your data sheet:
| ASME B40.100 grade | Permissible error, % of span |
|---|---|
| 4A | ±0.1 % |
| 3A | ±0.25 % |
| 2A | ±0.5 % |
| 1A | ±1 % |
| A | ±2 % in the first and last quarters, ±1 % in the middle half |
| B | ±3 % in the first and last quarters, ±2 % in the middle half |
EN 837-1 classes work the same way: a class 1.0 gauge is good to ±1 % of span, class 0.6 to ±0.6 % (WIKA data sheets; DKD-R 6-1, App. B, E and F).
Standards and guides that apply
| Document | Scope | Access |
|---|---|---|
| DKD-R 6-1 (2014, rev. 3) | Calibration of pressure gauges: references, sequences, points, uncertainty, certificate, recommended intervals | Free |
| EURAMET cg-17 (v4.1) | Calibration of electromechanical and mechanical manometers: Basic, Standard and Comprehensive procedures | Free |
| EURAMET cg-3 (v2.1) | Calibration of pressure balances (deadweight testers) | Free |
| USBR 1040-89 | A US government procedure for calibrating pressure gauges against a deadweight tester and a master gauge | Free |
| ASME B40.100 (2022) | US standard for dial pressure gauges, including accuracy grades | Paid |
| EN 837-1 | European standard for Bourdon tube gauges, including accuracy classes | Paid |
The reference: a deadweight tester or a reference gauge with a calibration certificate. About four times more accurate than the gauge is the long-standing rule of thumb (Mensor; Fluke); check your ratio with the TUR calculator. Never put a gauge marked for oxygen service on an oil-filled tester (Ashcroft; AIGA SP 09/20).
How often to calibrate a pressure gauge
DKD-R 6-1 leaves the interval to the user and recommends, under normal use, 1 year for Bourdon gauges of class 0.6 or better and electronic gauges of 0.5 % of span or better, 2 years for less accurate ones, and 5 years for deadweight testers (App. F). Twelve months is a common starting point in the US: the Bureau of Reclamation procedure and Ashcroft both use it (USBR 1040-89, §4.2; Ashcroft). Recalibrate after an overload, a repair or rough handling (DKD-R 6-1, App. F), then let your as-found history set the interval: see calibration intervals.
FAQ
What does NIST traceable mean on a digital pressure gauge?
That the gauge's calibration results link to the SI, usually through NIST, by a documented chain of calibrations with stated uncertainties. NIST does not certify traceability claims made by others, so the words alone prove nothing. Ask for the calibration certificate: readings at several pressures across the range, rising and falling, the deviation at each point, the uncertainty, the reference used and its traceability.
Is ±1 % of span the same as ±1 % of reading?
No. Dial gauge grades and classes are a percentage of the span, so the allowed error is the same number of psi everywhere on the dial. On a 0–160 psi gauge, ±1 % of span is ±1.6 psi, which at 20 psi is 8 % of the reading. If you work mostly at the low end of a gauge, a smaller range or a better grade gives you a much smaller error there.
Do gauges marked for oxygen service need special handling during calibration?
Yes. Gauges for oxygen service are cleaned of hydrocarbons and marked USE NO OIL, because oil residue can ignite in oxygen. Never connect one to an oil-filled tester or comparator: calibrate it on separate, clean equipment and keep the media apart.
Related guides
- How to read and review a calibration certificateA field-by-field guide to calibration certificates, with an annotated example, how to judge uncertainty and pass/fail statements, red flags, and a printable review checklist.
- Calibration intervals: how often to calibrateNo standard says 12 months. How to set an initial calibration interval, adjust it with the five ILAC G24 methods and a reliability target (EOPR), and defend it to an auditor.
- Out of tolerance: what to do when a calibration failsA step-by-step out-of-tolerance impact assessment: confirm the result, set the look-back window, correct past readings, decide on product, with an example report and template.
Sources
- PTB / DKD. DKD-R 6-1, Calibration of Pressure Gauges, Edition 03/2014, Revision 3. 2014 — §3 (standards, 1/3 recommendation), §5 (capability), §6 (ambient), §7 and Table 1 (method, sequences A–C), §8.3 and §8.6 (model, hysteresis), §9 and §9.3 (certificate, minimum uncertainty), App. B (Bourdon example), App. F (intervals). Current edition (DOI 10.7795/550.20210422).
- EURAMET. EURAMET Calibration Guide No. 17, Guidelines on the Calibration of Electromechanical and Mechanical Manometers, Version 4.1. 2022 — §5.1 (installation), §5.2 (Basic, Standard, Comprehensive procedures), §5.3 (set-ups), §5.4 (sequences, preloading, initial check, adjustment), §6.2.1 and §6.3 (uncertainty, hysteresis), Example 1a
- EURAMET. EURAMET Calibration Guide No. 3, Guidelines on the Calibration of Pressure Balances, Version 2.1. 2025 — §3 (principle, eq. 3.1, head correction 3.6 and 3.7), §4.2 (installation), §6.2 (results at standard gravity), App. C4.5 (local gravity)
- Ashcroft. What are ASME accuracy grades for mechanical pressure gauges?. 2026 — Grades 4A, 3A, 2A, 1A, A, B (ASME B40.100, current edition 2022, is a paid standard, not read)
- WIKA. Pressure gauge data sheets (models 232.53/233.53, 332.30, 342.11, 111.12) — ASME B40.100 grades and EN 837-1 classes as declared by the manufacturer
- Boshart Industries. What is 3-2-3 gauge accuracy?
- Ashcroft. How to check the calibration of a pressure gauge — Reference at least four times more accurate, attributed to ASME; 0–100 psi example; annual checks as a rule of thumb
- Mensor. The Fundamentals of Pressure Calibration — 4:1 TUR as a long-standing rule of thumb (no standard cited)
- Fluke Calibration. 2 guidelines to make sure the pressure calibration equipment you're buying fits your needs — A good test accuracy ratio is typically 4:1; checked 2026-10-16
- Dwyer Instruments. Series 2000 Magnehelic Differential Pressure Gages (data sheet) — Accuracy by range, calibrated vertical, position of use
- Ashcroft. Cleaning for oxygen service
- Asia Industrial Gases Association. SP 09/20, Oxygen cylinder pressure check (safe practice). 2020
- U.S. Bureau of Reclamation. USBR 1040-89, Procedure for Calibrating Pressure Gauges. 1989 — §3.1, §4.2, §9.2, Fig. 2 (example form)
- NIST. NIST Policy on Metrological Traceability. 2024
- JCGM / BIPM. International Vocabulary of Metrology (VIM), JCGM 200:2012. 2012 — 2.39 calibration, 3.11 adjustment