Pipette: types, ISO 8655 parts, error limits and calibration intervals
An overview of pipette calibration: pipette types, the parts of ISO 8655:2022, calibration vs routine testing, error limits, free guides and intervals. The gravimetric method step by step, with a calculator, is in our separate procedure.
- Calibration procedureHow to calibrate a pipette by the gravimetric method of ISO 8655-6:2022: the quick check vs a full calibration, what changed in 2022, balance and room requirements, test volumes and tips, the Z factor, systematic and random error, a worked example and a free calculator.Read the procedure
- Is your reference good enough?Enter the tolerance and your calibration uncertainty to get the test uncertainty ratio.Open the TUR calculator
Pipette calibration measures how much liquid a pipette really delivers and how much that amount varies, and compares both with a limit. The reference method is gravimetric: distilled water dispensed onto an analytical balance, each weight converted to volume with the Z factor, then the systematic error (how far the mean is from the set volume) and the random error (the scatter of the deliveries) calculated (ISO 8655-6; VIM 2.17, 2.19). Under ISO 8655-6:2022 a variable-volume pipette is tested at three volumes, at least ten weighings each, with a tip change at each volume (ISO 8655-6, 8.1; Sartorius white paper). Turning the adjustment ring to correct the volume is an adjustment, a separate step (VIM 3.11).
This page is the overview. For the quick check, the equipment, the step-by-step gravimetric method, a worked example and a free calculator, see pipette calibration step by step.
Pipette types
ISO 8655 covers piston-operated volumetric apparatus: pipettes, burettes, dilutors, dispensers and syringes (European Instruments, 2023; SelectScience, 2022). For pipettes, what changes the calibration is:
| Pipette | What to know |
|---|---|
| Fixed-volume | Tested at its nominal volume only (ISO 8655-6, 8.1.1) |
| Variable-volume | Tested at the nominal volume, 50 % of it, and the lower limit or 10 % of nominal, whichever is greater (ISO 8655-6, 8.1.1) |
| Multichannel | Tested channel by channel; separate error limits in ISO 8655-2:2022 (Sartorius; ISO 8655-2, foreword) |
| Electronic (motorized) | Same gravimetric test, with the manufacturer’s speed settings (ISO 8655-6) |
Air-displacement pipettes are the subject of DKD-R 8-1, the free German guideline for “piston-operated pipettes with air cushions”. Since 2022, the pipette and its tip are treated as one system, so results are valid only for the tip type used (ISO 8655-2, foreword; SelectScience).
ISO 8655 and the guides that apply
| Document | Scope | Access |
|---|---|---|
| ISO 8655-1:2022 | Terminology, general requirements and user recommendations | Paid |
| ISO 8655-2:2022 | Pipettes: design, adjustment, maximum permissible errors at 100 %, 50 % and 10 % of nominal | Paid |
| ISO 8655-6:2022 | Gravimetric reference measurement procedure | Paid |
| ISO 8655-7:2022 | Alternative procedures, including routine tests | Paid |
| ISO 8655-8:2022 | Photometric reference procedure, up to 5,000 µL | Paid |
| ISO 8655-10:2024 | User guidance, competence, training and suitability | Paid |
| DKD-R 8-1 (rev. 1) | Calibration of air-displacement pipettes, updated to reference ISO 8655:2022 | Free |
| EURAMET cg-19 (v4.1) | Uncertainty of gravimetric volume calibration | Free |
| NIST IR 7383-2019 (GLP 10, SOP 14) | Water density and the gravimetric calibration procedure | Free |
The parts of ISO 8655 are as summarized by European Instruments (2023) and SelectScience (2022). ISO 8655 describes how to calibrate; ISO/IEC 17025 sets the requirements for the lab that does it (Sartorius; see ISO/IEC 17025 explained).
Error limits
The 2022 edition sets limits at 100 %, 50 % and 10 % of nominal, where the 2002 edition gave them for the nominal volume only and applied the same absolute error across the range (ISO 8655-2:2022, foreword; Eppendorf, 2023). Eppendorf compared both editions: for single-channel pipettes above 100 µL, the limits at nominal volume stayed at 0.8 % systematic and 0.3 % random error, while several small-volume limits changed, most of them tightened (Eppendorf, 2023). We don’t reproduce the 2022 tables, which are not in the free preview; take the values for your pipette from ISO 8655-2 or the manufacturer.
How often to calibrate a pipette
ISO 8655 gives no single interval. Sartorius, summarizing the 2022 edition, lists the factors to weigh: the risk of the application, how often and by how many people the pipette is used, the liquids, the permissible errors you need and the manufacturer’s information (Sartorius). Routine tests should run more often than full calibrations (Sartorius white paper). Older manufacturer documents written for the 2002 edition state a yearly calibration (Eppendorf SOP, 2019, §4), which we could not confirm from the 2022 text. Recalibrate after a drop or a repair, and when a routine check fails; see calibration intervals. If you send pipettes out, the procedure lists what to ask a service, and our certificate guide shows how to review what comes back.
FAQ
What is the difference between a pipette calibration and a routine test?
A calibration under ISO 8655-6 tests a variable-volume pipette at three volumes with at least ten measurements each, under controlled conditions, and states the measurement uncertainty. A routine test under ISO 8655-7, as summarized by Sartorius, can use fewer volumes and at least four measurements per volume, and the uncertainty is optional. Routine tests run between calibrations, at shorter intervals, to catch a pipette that has drifted.
Does ISO 8655 cover multichannel pipettes?
Yes. The 2022 edition of ISO 8655-2 has a separate table of permissible errors for multichannel pipettes, and each channel is tested like a single-channel pipette: a 12-channel pipette at three volumes means 360 weighings. On a single-channel balance all channels aspirate and only the one being measured dispenses into the vessel; a multichannel balance measures all channels in parallel.
Against which limits is a pipette judged?
The calibration measures the systematic and random error; pass or fail needs a limit. That can be the maximum permissible errors of ISO 8655-2, the manufacturer's specification for the model, or tighter limits set by your lab for a critical method. A good certificate says which one it used. The results are valid only for the tip type used in the calibration.
Related guides
- How to read and review a calibration certificateA field-by-field guide to calibration certificates, with an annotated example, how to judge uncertainty and pass/fail statements, red flags, and a printable review checklist.
- Calibration intervals: how often to calibrateNo standard says 12 months. How to set an initial calibration interval, adjust it with the five ILAC G24 methods and a reliability target (EOPR), and defend it to an auditor.
- Out of tolerance: what to do when a calibration failsA step-by-step out-of-tolerance impact assessment: confirm the result, set the look-back window, correct past readings, decide on product, with an example report and template.
Sources
- ISO. ISO 8655-6:2022, Piston-operated volumetric apparatus, Part 6: Gravimetric reference measurement procedure for the determination of volume (official preview). 2022 — Paid standard. Read on the free official preview (foreword, clauses 1 to 8.3.2, Tables 1 and 2); clauses cited, not reproduced
- ISO. ISO 8655-2:2022, Piston-operated volumetric apparatus, Part 2: Pipettes (official preview). 2022 — Paid standard. Read on the free official preview (foreword, clauses 1 to 9.3); the tables of maximum permissible errors are not in the preview
- ISO. ISO 8655-8:2022, Piston-operated volumetric apparatus, Part 8: Photometric reference measurement procedure for the determination of volume. 2022 — Paid standard; scope only, from the catalog entry of its European adoption
- ISO. ISO 8655-10:2024, Piston-operated volumetric apparatus, Part 10: User guidance, and requirements for competence, training, and POVA suitability (catalog entry). 2024 — Paid standard; publication date and title only
- EURAMET. EURAMET Calibration Guide No. 19 (cg-19), Guidelines on the Determination of Uncertainty in Gravimetric Volume Calibration, version 4.1. 2025 — Current version (12/2025). §3 equations (1) to (4), §6.3.7.2, §6.3.8. Cited, not reproduced: extracts need EURAMET's permission
- PTB / Deutscher Kalibrierdienst. DKD-R 8-1, Calibration of piston-operated pipettes with air cushions, edition 12/2011, revision 1 — Revision 1 updates the references to ISO 8655:2022; technical content as in 2011. §5.3, §6, §7, §8, Appendices A to C. CC BY-NC-ND 3.0
- NIST Office of Weights and Measures. NIST IR 7383-2019, Selected Procedures for Volumetric Calibrations (GLP 10 and SOP 14). 2019 — GLP 10: water purity, water density equations and Table 2; SOP 14: gravimetric calibration, §1 to §3
- JCGM / BIPM. International Vocabulary of Metrology (VIM), JCGM 200:2012. 2012 — 2.13 accuracy, 2.14 trueness, 2.15 precision, 2.17 systematic error, 2.19 random error, 3.11 adjustment
- Sartorius. Pipette Calibration and Compliance: A Practical Guide to ISO 8655 (includes the white paper Gravimetric Calibration and Testing of POVA according to Part 6 and 7 of the ISO 8655:2022). 2023 — Secondary source on Parts 1, 6 and 7: test types, tip change, reporting, interval factors
- SelectScience. Pipette calibration ISO 8655:2022: everything you need to know. 2022 — Secondary source on the 2022 changes (interview with Sartorius metrology staff)
- European Instruments. ISO 8655:2022: everything you need to know. 2023 — Secondary source on the 2022 changes
- Eppendorf. Change of measurement values according to ISO 8655:2022. 2023 — Comparison of the maximum permissible errors at nominal volume, 2002 and 2022 editions
- Eppendorf. Standard Operating Procedure for manual dispensing systems. 2019 — §8.2, table of Z factors (used to validate our Z calculation)