ILAC and IAF to Global ACI: what changed for calibration and testing labs

ILAC and IAF handed their roles to Global ACI, fully operational since January 1, 2026. Which ILAC documents were replaced (P14, G8, P9, G17…), which have no Global ACI version yet (P10, G24), what to do with the ILAC MRA mark on certificates, and what to update in your quality system.

By David BoschUpdated Sources checked 9 primary sourcesHow we researchCite

On this page
  1. What Global ACI is
  2. The document map for labs
  3. What changed inside the documents labs cite most
  4. The accreditation mark on your certificates
  5. What to update in your quality system
  6. If you buy calibration
  7. FAQ
  8. Sources

ILAC and IAF have handed their roles to a single organization, the Global Accreditation Cooperation Incorporated (Global ACI), fully operational since January 1, 2026. For a calibration or testing lab, three things follow. Most ILAC documents you cite are being reissued as Global ACI documents with new codes: ILAC P14 is now Global ACI-TECH-1-009 and ILAC G8 is now TECH-1-002, both with the same content. A few have no Global ACI version yet: ILAC P10 (traceability) stays in force, and ILAC G24 (calibration intervals) is still the latest published guidance. And the ILAC MRA mark stays on your certificates for now: the new Global ACI MRA Mark had not been authorized as of November 5, 2026.

This page covers the documents that affect laboratories and inspection bodies, and was checked against the official Global ACI–IAF and ILAC Document Cross-Reference Table, version 14.0 (September 14, 2026). Global ACI updates that table often; when this page and the table disagree, the table wins.

What Global ACI is

Global ACI is the international association of accreditation bodies, the organizations that accredit laboratories, inspection bodies and certification bodies. It does not accredit anyone itself. It took over the roles of the International Laboratory Accreditation Cooperation (ILAC) and the International Accreditation Forum (IAF), and became fully operational on January 1, 2026 (Global ACI FAQ). The ILAC Mutual Recognition Arrangement (MRA) and the IAF Multilateral Recognition Arrangement (MLA) continue as the Global ACI MRA.

For a lab, the practical point is that recognition carries over. An accreditation body that signed the ILAC MRA for calibration and joined Global ACI is a signatory of the Global ACI MRA. In the United States, the accreditation bodies that are signatories for calibration are A2LA, ANAB, IAS, NAC, NVLAP and PJLA (Global ACI MRA Signatory List, October 19, 2026).

The old documents did not all lapse on January 1. For the purpose of peer evaluations, General Assembly resolution 2025-25 keeps the ILAC P-series policies and the IAF mandatory documents in force until Global ACI adopts an equivalent; resolution 2025-29 keeps specific documents on the arrangements and the marks in force, including ILAC R7 and P8 (cross-reference table, section 1). So an ILAC document can be replaced by a published Global ACI document, still in force (the P-series policies, and R7 and P8 for the old mark), or a guidance document with no Global ACI version yet.

Date What happened
October 23, 2025 Global ACI-MRA-006 issued (use of the MRA mark and accreditation claims)
January 1, 2026 Global ACI fully operational; MRA-006 implemented, with the old ILAC and IAF marks allowed during the transition
June 26, 2026 First technical documents issued, including TECH-1-002 (formerly G8) and TECH-1-009 (formerly P14)
September 14, 2026 Cross-reference table version 14.0, the one this page uses
October 19–26, 2026 Global ACI annual meetings in Panama City. As of November 5, no decision authorizing the Global ACI MRA Mark had been published

The document map for labs

These ILAC documents have a published Global ACI replacement. The dates are the issue dates in the cross-reference table.

  1. ILAC P14:09/2020

    Policy for measurement uncertainty in calibration: CMCs and the uncertainty on calibration certificates.

  2. ILAC G8:09/2019

    Guidelines on decision rules and statements of conformity.

  3. ILAC G17:01/2021

    Guidelines for measurement uncertainty in testing.

    Global ACI-TECH-1-003Replaced

    Version 1.0, June 26, 2026.

  4. ILAC P9:01/2024

    Policy for participation in proficiency testing activities.

    Global ACI-TECH-1-008Replaced

    Version 1.0, June 26, 2026. The title now also names interlaboratory comparisons other than proficiency testing.

  5. ILAC R7:05/2015 + ILAC P8:11/2023

    Rules for the ILAC MRA mark, and for accreditation symbols and claims of accreditation status.

    Global ACI-MRA-006Merged

    One document for the MRA mark and accreditation claims, implemented January 1, 2026. The old mark continues during the transition (see below).

  6. ILAC P15:05/2020

    Application of ISO/IEC 17020 for the accreditation of inspection bodies.

    Global ACI-TECH-2-001Replaced

    Version 1.0, August 23, 2026.

  7. ILAC G27:07/2019

    Guidance on measurements performed as part of an inspection process.

    Global ACI-TECH-2-002Replaced

    Version 1.0, June 26, 2026.

  8. ILAC G28:07/2018

    Formulation of scopes of accreditation for inspection bodies.

    Global ACI-TECH-2-003Replaced

    Version 1.0, August 23, 2026.

  9. ILAC G26:11/2018

    Implementation of a medical accreditation scheme.

    Global ACI-TECH-1-006Replaced

    Version 1.0, July 31, 2026.

  10. ILAC G19:06/2022

    Modules in a forensic science process.

    Global ACI-TECH-1-004Replaced

    Version 1.0, June 26, 2026.

  11. ILAC G7:04/2021

    Accreditation requirements for horseracing laboratories.

    Global ACI-TECH-1-001Replaced

    Version 1.0, June 26, 2026.

Replaced, same textReplacedMerged

ILAC documents for laboratories and inspection bodies and their Global ACI replacements. Our selection and summaries; titles shortened. Source: Global ACI–IAF and ILAC Document Cross-Reference Table v14.0 (September 14, 2026), section 2; TECH-1-009 §0.1.4; TECH-1-002 §1.1; MRA-006 cover page and Annex 1.

Replaced, same text means Global ACI states in the document itself that the ILAC text was reformatted without substantive change. Replaced means a new Global ACI version exists; we have not compared it line by line with the ILAC original, so check it before you rely on a specific clause.

No Global ACI version yet: keep citing these

For these documents, the Global ACI successor is listed as in process in table version 14.0. Only P10, a policy, is marked as still in force (resolution 2025-25 covers the P-series). The three guidance documents are not, but nothing has replaced them, so they remain the latest published guidance:

ILAC document Topic Future Global ACI document
ILAC P10:07/2020 Metrological traceability of measurement results TECH-1-011
ILAC G24:2022 (OIML D 10:2022) Determination of calibration intervals TECH-1-005
ILAC G18:01/2024 Describing scopes of accreditation TECH-1-010
ILAC G29:06/2020 Scopes for WADA anti-doping laboratories TECH-1-007

If your calibration interval procedure cites ILAC G24, there is nothing to change yet. The same goes for traceability statements that cite ILAC P10.

The table also maps governance, arrangement and form documents, and most of IAF’s documents for certification bodies (the IAF MD and ID series). Those are outside the scope of this page; see the official table.

What changed inside the documents labs cite most

ILAC P14 → Global ACI-TECH-1-009. The policy on calibration and measurement capabilities (CMCs) and on the uncertainty stated on calibration certificates is the same. Global ACI says the text was reformatted without any substantive change (TECH-1-009 §0.1.4), and the numbered sections that labs and assessors quote, such as the CMC rules and the rules for reporting uncertainty, keep their numbers. Our measurement uncertainty guide and certificate guide already cite the new reference.

ILAC G8 → Global ACI-TECH-1-002. Same text, reformatted without substantive change (TECH-1-002 §1.1), but every section number goes up by one: the preamble and purpose, unnumbered in G8, are now section 1, Introduction. If your decision rule procedure points to a section, renumber it:

Topic ILAC G8:09/2019 Global ACI-TECH-1-002
Definitions (guard band, simple acceptance, specific and global risk) 1.1–1.16 2.1–2.16
ISO/IEC 17025 clauses on decision rules 2.1–2.7 3.1–3.7
Guard bands; binary and non-binary decision rules 4.1–4.2 5.1–5.2
Guard band based directly on the measurement uncertainty (Table 1) 5.2 6.2
Specific vs global risk; false accept vs false reject 5.3–5.4 6.3–6.4
Choosing a decision rule (flow chart) 6 7

The guard banding guide and the TUR calculator use the TECH-1-002 numbering.

ILAC P9 → Global ACI-TECH-1-008. The proficiency testing policy has a broader title: it now names interlaboratory comparisons other than proficiency testing (cross-reference table). If your PT plan cites P9, update the reference and read the new version when you next review the plan.

The accreditation mark on your certificates

ILAC R7 (the ILAC MRA mark) and ILAC P8 (accreditation symbols and claims) are merged into Global ACI-MRA-006, implemented on January 1, 2026. An accredited lab will use the new Global ACI MRA Mark only combined with its accreditation body’s own symbol, after written approval, and in economies where the mark is registered or its registration is pending (MRA-006 §6–7).

Until then, Annex 1 of MRA-006 sets a temporary process:

  • Accreditation bodies that were ILAC MRA signatories on December 31, 2025, and not suspended, can keep using the ILAC MRA mark, including for scope extensions granted after January 1, 2026 (Annex 1, 1.3.2).
  • Labs that were authorized to use the combined ILAC MRA mark before January 1, 2026 can keep using it during the transition, under their accreditation body’s rules (Annex 1, note 2).
  • The mark is used under the same rules as before, ILAC R7 and ILAC P8 (Annex 1, 1.3.3, which cites the 2019 edition of P8). Existing licenses and agreements stay valid until the Global ACI MRA Mark is legally registered and available (1.3.4).
  • The transition ends country by country (economy by economy, in Global ACI’s terms), when the new mark is legally registered and available there; Global ACI will give advance notice (Annex 1, 1.6–1.7). From then on, a lab must not put the combined ILAC MRA mark on any certificate issued, reissued or renewed, and switches to the new mark as its accreditation body instructs (note 2).

As of November 5, 2026, the new mark was not authorized. The Global ACI FAQ says members are not currently allowed to use it and points to a decision at the Arrangements Committee meeting in October 2026; we found no resolution or news item announcing one after those meetings. A2LA tells its accredited bodies to keep using their ILAC/IAF mark until the registration process is complete. UKAS describes the introduction of the new mark as paused, and expects a transition of three to five years once it starts, a period Global ACI has not confirmed (UKAS technical bulletin, May 15, 2026).

What to update in your quality system

None of the documents we checked sets a deadline for updating references in a lab’s own documents. The practical approach is to update each one when you next revise it, and to start with what assessors read first: the uncertainty and decision rule procedures, certificate templates and supplier criteria.

Global ACI transition checklist for labs

Procedures and quality manual (update when you next revise them)
Certificates, reports and labels
Purchasing and supplier approval
Keep watching

If you buy calibration

Nothing changes in what to ask for: an accredited calibration, from a lab whose scope covers the instrument and range, with a certificate that states the measurement uncertainty. Two wording details:

  • “ILAC accredited” is common shorthand, but not accurate. Labs are accredited by an accreditation body, and that body is a signatory to the Global ACI MRA (formerly the ILAC MRA). Check the body on the Global ACI MRA Signatory List, then check the lab’s scope on the accreditation body’s website.
  • A certificate issued in 2026 with the accreditation body’s symbol and the ILAC MRA mark is normal. That is the mark the transition rules allow today.

More on what a certificate should show: how to read and review a calibration certificate. More on the standard itself: ISO/IEC 17025 explained.

FAQ

Does ILAC still exist?

ILAC and IAF transferred their roles to Global Accreditation Cooperation Incorporated (Global ACI), which became fully operational on January 1, 2026. The ILAC website is still online and still lists the old documents, but new and revised documents are now published by Global ACI. Some ILAC documents remain in force until Global ACI adopts an equivalent.

Is ILAC P14 still valid?

No. ILAC P14:09/2020 was replaced by Global ACI-TECH-1-009, Policy for Measurement Uncertainty in Calibration, issued and implemented on June 26, 2026. Global ACI says the text was reformatted without any substantive change, and the section numbers you are likely to cite are the same, so in most procedures only the document reference needs updating.

What replaced ILAC G8?

Global ACI-TECH-1-002, Guidelines on Decision Rules and Statements of Conformity, issued on June 26, 2026. The content is unchanged, but every section number went up by one: for example, G8 section 5.2 (guard bands based on the measurement uncertainty) is now TECH-1-002 section 6.2.

Are ILAC P10 and ILAC G24 still valid?

Neither has a Global ACI version yet: TECH-1-011 (traceability) and TECH-1-005 (calibration intervals) were still in process in the cross-reference table of September 14, 2026. ILAC P10:07/2020 is a policy and stays in force under resolution 2025-25, which keeps the ILAC P-series in force until Global ACI adopts an equivalent. ILAC G24:2022 is guidance; the table does not list it as in force, but it is still the latest published guidance on calibration intervals, so keep citing it until TECH-1-005 appears.

Can I still use the ILAC MRA mark on calibration certificates?

Yes, if you were already allowed to, but only until the Global ACI MRA Mark is available in your country. Global ACI-MRA-006 (Annex 1, note 2) lets labs keep the combined ILAC MRA mark during the transition, under their accreditation body's rules. Once the new mark is legally registered and available in their economy, no certificate issued, reissued or renewed may carry the combined ILAC mark. As of November 5, 2026, Global ACI had not authorized the new mark. Do not use it until your accreditation body tells you to.

Do I need to rewrite my quality manual?

Not overnight. None of the documents we checked sets a deadline for updating references in a lab's own procedures. A practical approach is to update each reference when you next revise the document, starting with the ones auditors read most: the measurement uncertainty and decision rule procedures, and certificate templates.

Sources

  1. Global Accreditation Cooperation Incorporated (Global ACI). Global ACI–IAF and ILAC Document Cross-Reference Table, v14.0. 2026 — Last update September 14, 2026; sections 1 (resolutions 2025-25 and 2025-29) and 2 (table)
  2. Global ACI. Global ACI-TECH-1-009 (M) Policy for Measurement Uncertainty in Calibration, v1.0. 2026 — Issued and implemented June 26, 2026; §0.1.4
  3. Global ACI. Global ACI-TECH-1-002 (G) Guidelines on Decision Rules and Statements of Conformity, v1.0. 2026 — Issued June 26, 2026; §1.1
  4. Global ACI. Global ACI-MRA-006 Use of the MRA Mark by Accreditation Bodies and Conformity Assessment Bodies, and Claims of Accreditation Status by Conformity Assessment Bodies, v1.0. 2025 — Issued October 23, 2025, implemented January 1, 2026; §6–7 and Annex 1
  5. Global ACI. Global ACI MRA Signatory List. 2026 — Version of October 19, 2026
  6. Global ACI. Frequently asked questions. 2026 — Checked November 5, 2026
  7. Global ACI. General Assembly resolutions. 2026 — Checked November 5, 2026, after the October 2026 meetings in Panama City
  8. A2LA. Global ACI FAQs: updates, marks and more for A2LA CABs. 2026 — May 26, 2026
  9. UKAS. Transition to the Global ACI MRA Mark (technical bulletin). 2026 — May 15, 2026