FDA calibration requirements: QMSR (ISO 13485 7.6), 21 CFR 211.68, 211.160 and Part 11

What FDA regulations require for calibration, in plain words: what replaced 21 CFR 820.72 under the QMSR, ISO 13485 clause 7.6, drug CGMP (211.68, 211.160(b)(4)), GLP (58.63), Part 11 for electronic records, how often to calibrate, and what investigators cite.

Updated 17 primary sourcesHow we researchCite

On this page
  1. Which FDA calibration rule applies to you
  2. Medical devices: what happened to 21 CFR 820.72
  3. ISO 13485 clause 7.6, requirement by requirement
  4. Drugs: 21 CFR 211.68 and 211.160(b)(4)
  5. GLP labs: 21 CFR 58.63
  6. How often does the FDA require calibration?
  7. Electronic calibration records and Part 11
  8. What FDA investigators cite: Form 483 data
  9. Audit-ready checklist
  10. FAQ
  11. Sources

FDA has no single calibration rule. Each FDA regulation that touches measurement asks for the same core: a written program or procedure, calibration or checks at intervals you set and can justify, limits the equipment must meet, records, and action when equipment fails. Which text applies depends on what you make. For medical devices, it is now ISO 13485:2016 clause 7.6, incorporated into 21 CFR 820 by the Quality Management System Regulation (QMSR), in force since February 2, 2026; the old 21 CFR 820.72 no longer exists. For drugs, it is 21 CFR 211.68(a) and 211.160(b)(4). For nonclinical (GLP) labs, 21 CFR 58.63. If you keep any of those records electronically, Part 11 applies as well.

None of these rules sets a calibration interval, and none of their texts requires an ISO/IEC 17025 accredited lab. What FDA investigators look at is whether your program is written, followed, traceable and documented, and what you did when an instrument was found out of tolerance.

A note on sources: the FDA regulations are quoted from the current eCFR text (up to date as of October 13, 2026) and paraphrased. ISO 13485 is a paid standard: we do not reproduce it, we cite clause numbers and describe clause 7.6 from the audit guidance that regulators publish (the MDSAP Audit Approach, which FDA hosts) and from FDA’s own inspection program. FDA points to a read-only copy of ISO 13485:2016 on ANSI’s Incorporated by Reference portal, at no cost to the user (QMSR FAQ 12; 89 FR 7500, 7505).

Which FDA calibration rule applies to you

Medical devices Drugs (finished) APIs Nonclinical labs (GLP)
Rule QMSR: 21 CFR 820.10 → ISO 13485:2016 7.6 21 CFR 211.68(a) (production equipment) and 211.160(b)(4) (laboratory) ICH Q7 §5.3 (FDA guidance, not a regulation) 21 CFR 58.63
What to calibrate Monitoring and measuring equipment that shows product conformity Automatic, mechanical and electronic equipment, computers included; lab instruments, apparatus, gauges and recorders Equipment critical to the quality of intermediates or APIs Equipment that generates, measures or assesses data
Program Documented procedures Written program designed to assure proper performance; for the lab, specific directions, schedules, accuracy and precision limits, remedial action Written procedures and an established schedule Written SOPs with methods, materials, schedules and the person responsible
Interval Set by you (none in the rule) “Routinely” (211.68) and “at suitable intervals” (211.160) Established schedule Schedules in the SOP
Traceability To national or international standards, where they exist Not stated in these sections To certified standards, if they exist Not stated (OECD GLP: “where appropriate”)
Failure Assess and record validity of earlier results; act on equipment and product Remedial action; do not use equipment out of specification Do not use; investigate effect since the last successful calibration Remedial action in the SOP; records of non-routine repairs
Records Calibration and verification results (7.6, 4.2.5) Written records (211.68(a)); complete lab calibration records (211.194(d)) Records maintained; current status known and verifiable Dated records of every operation, and whether the SOP was followed
Electronic records Part 11 applies to all four when FDA-required records are kept electronically (21 CFR 11.1(b))
FDA calibration requirements by sector. Requirements paraphrased; the full texts are linked under Sources. Source: 21 CFR 820.7, 820.10 (eCFR, October 13, 2026); ISO 13485:2016 7.6 via MDSAP Audit Approach Tasks 13–15 and FDA CP 7382.850; 21 CFR 211.68, 211.160, 211.194; ICH Q7 §5.3; 21 CFR 58.63; 21 CFR 11.1.

Combination products (21 CFR Part 4) and contract manufacturers can fall under more than one column. If your organization also follows ISO 9001 clause 7.1.5 or runs an internal lab under ISO/IEC 17025, one calibration program can serve all of them; the calibration program guide shows how to set one up.

Medical devices: what happened to 21 CFR 820.72

On February 2, 2024, FDA published the final rule that turned 21 CFR 820 into the Quality Management System Regulation (QMSR); it took effect on February 2, 2026 (89 FR 7496). The new Part 820 incorporates by reference ISO 13485:2016 and clause 3 of ISO 9000:2015 (21 CFR 820.7), and requires manufacturers to document a quality management system that complies with the applicable requirements of ISO 13485 and of Part 820 (820.10(a)).

What that means for calibration:

  • 820.72 is gone. The current Part 820 has only six sections: 820.1, 820.3, 820.7, 820.10, 820.35 and 820.45. None of them mentions calibration or measuring equipment. The requirement now comes from ISO 13485 clause 7.6, Control of monitoring and measuring equipment (eCFR, October 13, 2026).
  • FDA inspects clause 7.6. Since February 2, 2026, investigators follow Compliance Program 7382.850. Its list of inspection elements maps “Control of Monitoring and Measuring Equipment” to clause 7.6, and “Software Changes” to clauses 4.1.6, 7.5.6 and 7.6 (CP 7382.850, Attachment A, pp. 71 and 76). FDA stopped using QSIT and the old device program 7382.845 the same day (FDA QMSR page; QMSR FAQ 9).
  • The final rule does not discuss calibration. It judges ISO 13485 to be, taken as a whole, “substantially similar” to the old Quality System regulation (89 FR 7496), and mentions clause 7.6 only in a list of clauses where the standard builds in risk management (89 FR 7505). The paragraph-by-paragraph mapping below is ours, built from the old text and the MDSAP audit guidance, whose 2024 revision paired 820.72 with clause 7.6 (MDSAP Audit Approach AU P0002.009, Tasks 13–14).
  1. 820.72(a)

    Equipment suitable for its intended purpose and capable of producing valid results.

    ISO 13485 7.6Carried over

    Determine the monitoring and measurement you need and the equipment for it; equipment capable of valid results.

  2. 820.72(a)

    Procedures so equipment is routinely calibrated, inspected, checked and maintained; activities documented.

  3. 820.72(a)

    Handling, preservation and storage that keep accuracy and fitness for use.

    ISO 13485 7.6Carried over

    Protect equipment from damage and deterioration in handling, maintenance and storage; account for environmental conditions.

  4. 820.72(b)

    Calibration procedures with specific directions and limits for accuracy and precision.

    ISO 13485 7.6Reworded

    No longer spelled out as “limits for accuracy and precision”. MDSAP auditors expect them defined during planning when they affect the validity of results.

  5. 820.72(b)

    Limits not met: remedial action, and evaluate whether device quality was adversely affected.

  6. 820.72(b)(1)

    Standards traceable to national or international standards; else an independent reproducible standard; else an in-house standard.

    ISO 13485 7.6Reworded

    Calibration traceable to national or international standards where they exist; external calibration under supplier controls.

  7. 820.72(b)(2)

    Record ID, calibration dates, who calibrated, next due date; on or near the equipment, or readily available.

  8. 820.70(i)

    Validation of software used in production or the quality system (a different section of the old rule).

Carried overRewordedMoved hereNo longer prescribed

Former 21 CFR 820.72 and where each obligation sits under the QMSR. Our mapping; ISO 13485 wording paraphrased from the MDSAP Audit Approach. Source: 21 CFR 820.72 as of January 15, 2026 (eCFR historical version); MDSAP AU P0002.009, Tasks 13–15 (pp. 107–109); FDA CP 7382.850, Attachment A.

Two details deserve attention. The old rule allowed an independent reproducible standard or an in-house standard when national or international standards were not practical or available (former 820.72(b)(1)); clause 7.6, as MDSAP describes it, asks for traceability to national or international standards where they exist (see the note on unconfirmed wording below for what happens when none does). And the old rule’s list of record fields (equipment ID, dates, who calibrated, next due date, on or near the equipment) is no longer written into the regulation. It is still a sensible minimum, and our calibration records guide uses it.

ISO 13485 clause 7.6, requirement by requirement

Clause 7.6 kept the number and title that ISO 9001 used before 2015, which is why it reads like the older ISO 9001 clause; the ISO 9001 equivalent is now clause 7.1.5. What the clause requires, as described in the MDSAP Audit Approach (rev. .009, Tasks 13–15, pp. 107–109), and what an investigator or auditor can be shown:

ISO 13485 7.6 · applies to every manufacturer under the QMSR

7.6

Know what you measure, and with what

Decide which monitoring and measurement show that product conforms, and which equipment does it.

  • Equipment list linked to the product characteristics it checks
  • Accuracy and precision needed, set during planning
7.6

Documented procedures

Procedures for controlling the equipment, including environmental, handling and storage needs.

  • Program procedure plus instrument-specific calibration procedures
  • Storage and handling rules where accuracy depends on them
7.6

Calibrate or verify, traceably

Calibration traceable to national or international standards where they exist, on a schedule you define and justify.

  • Certificates showing the traceability chain
  • A schedule with the reason for each interval
  • Supplier approval for the outside lab
7.6

Identify, adjust, maintain, protect

Equipment identified, adjusted when needed, maintained and protected from damage or deterioration.

  • Unique ID and status on the instrument or in the system
  • Maintenance history
7.6

Out of tolerance? Assess and record

Assess and record whether earlier results are still valid, then act on the equipment and any affected product.

  • Impact assessment covering in-process, finished and released devices
  • Actions taken, with the decision recorded
7.6

Records of results

Keep calibration and verification results as controlled records.

  • As-found and as-left results, not just pass or fail
  • Records kept at least for the device lifetime and not less than two years (4.2.5)
7.6

Validate measurement software

Software used for monitoring and measurement validated for its intended use, with software changes controlled.

  • Requirements, protocol with acceptance criteria, results
  • Change control for updates
ISO 13485:2016 clause 7.6: requirement and the evidence that shows it. Requirements paraphrased, not quoted from the standard. Source: MDSAP AU P0002.009, Tasks 13–15 (pp. 107–109); FDA CP 7382.850, Attachment A (p. 76); MDSAP Task 8 for record retention (clause 4.2.5).

Points the audit guidance makes that are easy to miss:

  • Out-of-tolerance results. You assess and record the validity of earlier measurement results and act on the equipment and any affected product; the auditor checks that the assessment covered in-process, finished or already released devices, based on risk (MDSAP Task 13 and Task 14). The old rule also required a documented evaluation of adverse effect on device quality, so the change is mostly in framing: earlier results and released product are named explicitly. Our out-of-tolerance guide walks through the assessment.
  • Outside labs. When a supplier calibrates for you, supplier controls apply so the work is done competently (MDSAP Task 14). Accreditation is not demanded, but it is the simplest evidence; see ISO/IEC 17025.
  • Software. Software used in the quality system, including off-the-shelf quality system software such as a calibration database, must be validated for its intended use against a protocol with predetermined acceptance criteria; you do not have to review the vendor’s code (MDSAP Task 15).
  • Wording we could not confirm in a primary source: that equipment is calibrated or verified at specified intervals or before use, and that you record the basis used when no national or international standard exists. Secondary descriptions of clause 7.6 by calibration labs and ISO trainers say both, and both mirror the pre-2015 ISO 9001 clause, but neither appears in the regulator guidance we used. Treat them as likely (medium confidence) and check the read-only copy on ANSI’s portal.

If you want a structured, auditable system around these requirements, ISO 10012 describes one built around metrological confirmation.

Drugs: 21 CFR 211.68 and 211.160(b)(4)

The drug CGMP rule has two calibration requirements, one for production equipment and one for the laboratory. Neither changed with the QMSR.

21 CFR 211.68(a), automatic, mechanical and electronic equipment. Equipment of these kinds, computers included, may be used in making, processing, packing and holding drugs, but it must be routinely calibrated, inspected or checked under a written program designed to assure proper performance, and written records of those calibration checks and inspections must be kept. Paragraph (b) adds controls for computer systems: only authorized people may change master production and control records or other records, inputs and outputs are checked for accuracy, and backup data must be exact, complete and protected from alteration, erasure or loss (211.68(b)).

21 CFR 211.160(b)(4), laboratory controls. Instruments, apparatus, gauges and recording devices must be calibrated at suitable intervals under an established written program that contains:

  1. specific directions;
  2. schedules;
  3. limits for accuracy and precision;
  4. provisions for remedial action when the limits are not met.

Equipment that does not meet its established specifications must not be used (211.160(b)(4)). The lab must keep complete records of that periodic calibration (211.194(d)). The rule sets no separate retention period for calibration records; batch-related records are kept at least one year after the batch’s expiration date (211.180(a)), so a calibration record that supports a batch decision is worth keeping at least that long. That last point is our recommendation, not a stated rule.

What this looks like in an FDA case: in 2018, FDA sent a warning letter to an over-the-counter drug manufacturer whose densitometer, used to calculate fill volumes, was not calibrated and had no calibration procedure (FDA Warning Letter 320-18-71, observation 3, citing 211.160(a) and (b)(4)).

Active pharmaceutical ingredients. FDA’s guidance ICH Q7 recommends calibrating equipment critical to quality under written procedures and an established schedule, with standards traceable to certified standards if they exist, records, a current calibration status that can be verified, no use of instruments that fail, and an investigation of the possible effect on product made since the last successful calibration (ICH Q7, 5.30–5.35). It is guidance, not a regulation, but it is how FDA describes CGMP for APIs.

GLP labs: 21 CFR 58.63

Nonclinical laboratories that run studies under FDA’s Good Laboratory Practice rule have a short, specific requirement (21 CFR 58.63):

  • (a) Equipment is adequately inspected, cleaned and maintained, and equipment used to generate, measure or assess data is adequately tested, calibrated and/or standardized.
  • (b) Written SOPs set out, in enough detail, the methods, materials and schedules for routine inspection, cleaning, maintenance, testing, calibration and standardization, specify remedial action for failures where appropriate, and name the person responsible for each operation.
  • (c) Written records of every inspection, maintenance, testing, calibration and standardization, dated, stating whether the work was routine and followed the SOP. Non-routine repairs are recorded with the nature of the defect, how and when it was found, and the remedial action.

Part 58 does not mention traceability. The OECD GLP principles do, “where appropriate”, as FDA’s own comparison of FDA, EPA and OECD GLP rules shows (FDA BIMO, Equipment). The EPA’s equivalent, 40 CFR 792.63, uses almost the same words as 58.63.

How often does the FDA require calibration?

FDA regulations do not set calibration intervals. The texts ask for calibration “routinely” under a written program (211.68(a)), “at suitable intervals” with schedules (211.160(b)(4)), on schedules written into SOPs (58.63(b)), and, for devices, under procedures you define in your ISO 13485 system. You choose the interval and must be able to justify it.

Two official sources give a little more guidance:

  • For balances with built-in auto-calibration, FDA says the internal feature cannot replace an external performance check (211.68). External checks can be less frequent than for a balance without that feature, and their frequency depends on how often the balance is used and how critical and tight the process step is. If the auto-calibrator check reveals a problem, every batch made between the two successive verifications is affected. FDA adds that the auto-calibrator itself should be verified periodically, “a common frequency is once a year”, with NIST-traceable standards (FDA Q&A on Laboratory Controls, question 1). That is described as common practice, not a requirement.
  • For devices, MDSAP auditors expect the initial interval to come from the manufacturer’s recommendation and to be adjusted with experience and a documented rationale (MDSAP Task 13).

The balance answer shows why the interval matters: the longer it is, the more product you have to review when a calibration fails. Our calibration intervals guide covers how to set and adjust intervals with the recognized methods, and the label generator prints due-date labels for the instruments.

Electronic calibration records and Part 11

Part 11 applies when records that FDA regulations require are kept in electronic form (21 CFR 11.1(b)). Calibration records under 211.68(a), 211.194(d), 58.63(c) or the QMSR qualify, so a calibration database or a spreadsheet that holds them falls within Part 11. Paper records sent by email do not (11.1(b)).

For a closed system, Part 11 lists controls that include validation, accurate and complete copies, protection of records throughout the retention period, access limited to authorized people, and secure, computer-generated, time-stamped audit trails that do not obscure earlier entries and are kept as long as the records (11.10(a)–(e)), plus authority checks and training (11.10(g), (i)). Electronic signatures must show the signer’s name, the date and time, and the meaning of the signature, and be linked to their records (11.50, 11.70).

In its 2003 guidance, FDA said it intends to exercise enforcement discretion over the Part 11 requirements for validation, audit trails, record copying and record retention, while the underlying record requirements still apply in full and controls such as limited access, authority checks and electronic signatures are still enforced. It recommends basing those decisions on a documented risk assessment (FDA Part 11 guidance, §III). For device inspections, the compliance program tells investigators to follow agency procedures for electronic records and signatures (CP 7382.850, p. 43).

Two more layers apply by sector:

  • Drugs: 211.68(b) requires its own computer controls (authorized changes only, checked inputs and outputs, secure backups), whatever Part 11 enforcement policy says.
  • Devices: calibration software used in the quality system must be validated for its intended use (ISO 13485 7.6 and 4.1.6; MDSAP Task 15). FDA’s Computer Software Assurance guidance (February 2026) describes a risk-based way to provide that assurance for production and quality system software.

No software is “Part 11 compliant” on its own; compliance depends on how you configure, validate and use it. Our comparison of calibration management software shows what to ask vendors, and the calibration records guide covers paper, spreadsheet and software options. Part 11 does not apply to records kept only for ISO 9001 or ISO/IEC 17025.

What FDA investigators cite: Form 483 data

FDA publishes, by fiscal year, how often each regulation was cited on the Form 483s its investigators issue. These are the calibration-related citations in the two latest data sets:

Regulation What the citations say (paraphrased) FY2024 FY2025
21 CFR 820.72 (devices, former rule) Six codes (five cited in FY2025): procedures missing or inadequate, activities not documented, procedures without limits, no or inadequate remedial action, out-of-tolerance evaluation not documented, equipment not suitable 55 41
of which code 3233 Procedures for routine calibration, inspection, checks or maintenance missing or inadequate 37 30
21 CFR 211.68(a) (drug equipment) Routine calibration, inspection or checking not performed (code 1274); calibration records not kept (1261) 32 43
21 CFR 211.160(b)(4) (drug laboratory) No written calibration procedures, calibration not at suitable intervals or without a written program, instruments used out of specification (three codes cited in FY2024, five in FY2025) 20 15
21 CFR 58.63(a) (GLP) Equipment not adequately tested, calibrated or standardized (code 3954) 1 0
Calibration-related citations on FDA Form 483, by regulation. Frequency is the number of times the section was cited, not the number of firms. Source: FDA Inspection Observations, FY2024 and FY2025 data sets (Devices, Drugs and Bioresearch Monitoring tabs). Our count of the calibration codes listed.

How to read the table:

  • The data sets cover only 483s generated with FDA’s electronic inspection tools; some 483s are prepared by hand and are not included (FDA Inspection Observations page).
  • Fiscal year 2025 ended on September 30, 2025, before the QMSR took effect, so device citations still use 820.72. FY2025 is the latest data set FDA has published; the first QMSR-era figures will come with FY2026.
  • For scale, FDA’s electronic system issued 791 device 483s and 713 drug 483s in FY2025 (Summary tab). We do not convert the counts into percentages, because one 483 can cite several sections.
  • The most frequent problem is not a bad instrument but a missing or inadequate procedure, or calibration that simply was not done. Records and programs, not metrology, drive most citations.

Audit-ready checklist

Use this before an FDA inspection or an internal audit. It applies to all four rules; drop what does not apply to your sector.

FDA calibration readiness checklist

Program and procedures
Equipment and intervals
Traceability and suppliers
Out of tolerance
Records and software

For the program behind the checklist, see the calibration program guide, how to read a calibration certificate and what to do when a calibration fails.

FAQ

What are the FDA calibration requirements?

There is no single FDA calibration rule. Each regulation asks for the same core: a written program or procedure, calibration or checks at intervals you justify, limits for accuracy (and, in most rules, precision), records, and action when equipment fails. For medical devices that now comes from ISO 13485:2016 clause 7.6, incorporated into 21 CFR 820 by the QMSR since February 2, 2026. For drugs it is 21 CFR 211.68(a) and 211.160(b)(4); for nonclinical (GLP) labs, 21 CFR 58.63. If the records are electronic, Part 11 applies too.

Does 21 CFR 820.72 still apply?

No. Since February 2, 2026, 21 CFR 820 is the Quality Management System Regulation (QMSR). The current Part 820 has only sections 820.1, 820.3, 820.7, 820.10, 820.35 and 820.45, and none of them mentions calibration. Device manufacturers must document a quality system that meets ISO 13485:2016, and FDA now inspects measuring equipment against clause 7.6 of that standard (Compliance Program 7382.850). Procedures that still cite 820.72 should be updated.

What replaced 21 CFR 820.72?

ISO 13485:2016 clause 7.6, Control of monitoring and measuring equipment, which the QMSR incorporates by reference (21 CFR 820.7 and 820.10). Most of the old 820.72 carries over: suitable equipment, documented procedures, protection in handling and storage, traceable calibration, and action when equipment is out of tolerance. What changed: the out-of-tolerance check is framed as the validity of earlier measurement results and action on any affected product (the old rule asked for an evaluation of adverse effect on device quality, also documented), validation of measurement software now sits in the same clause (it was in 820.70(i)), and the regulation no longer prescribes a fixed list of record fields.

How often does the FDA require calibration?

FDA regulations do not set calibration intervals. 21 CFR 211.68(a) asks for routine calibration under a written program, 211.160(b)(4) for calibration at suitable intervals with schedules, and 58.63(b) for schedules in your procedures. You set and justify the interval. MDSAP audit guidance notes that the initial interval is usually based on the manufacturer's recommendation and can be adjusted with experience, with a documented rationale.

Does the FDA require an ISO 17025 accredited calibration lab?

The regulation texts do not require accreditation. They require calibration you can trust: traceable standards (for devices, through ISO 13485 7.6) and, for devices, supplier controls when an outside lab does the work. An ISO/IEC 17025 accredited lab with the calibration in its scope is the easiest way to show competence and traceability, but it is a choice, not an FDA rule.

Is an ISO 13485 certificate enough for the FDA?

No. FDA does not require ISO 13485 certification, does not issue certificates, and a certificate does not exempt a manufacturer from FDA inspection (QMSR FAQ 13; final rule, responses to comments 79 to 81). FDA inspects your compliance with the QMSR, including clause 7.6, itself.

What does 21 CFR 211.68 require for calibration?

211.68(a) says that automatic, mechanical or electronic equipment used in drug manufacturing, computers included, must be routinely calibrated, inspected or checked under a written program designed to assure proper performance, and that written records of those checks are kept. 211.68(b) adds controls over computer systems: only authorized people change records, inputs and outputs are checked, and backups are exact, complete and secure.

What does 21 CFR 211.160(b)(4) require?

Laboratory instruments, apparatus, gauges and recording devices must be calibrated at suitable intervals under an established written program that contains specific directions, schedules, limits for accuracy and precision, and provisions for remedial action when limits are not met. Equipment that does not meet its specifications must not be used. 211.194(d) requires complete records of that calibration.

Does Part 11 apply to calibration records and calibration software?

Yes, when the records are ones FDA regulations require you to keep, such as calibration records under 21 CFR 211, 58 or the QMSR, and you keep them electronically (21 CFR 11.1(b)). Then the system needs Part 11 controls such as validation, limited access and secure audit trails, within the enforcement discretion FDA described in its 2003 guidance. Part 11 does not apply to records you keep only for ISO 9001 or ISO/IEC 17025.

Is the 4:1 rule an FDA requirement?

No. None of the FDA regulations on this page sets a test accuracy or uncertainty ratio. A 4:1 test uncertainty ratio is a common industry target for making sure the reference is good enough for the tolerance; see our TUR calculator for what it means and how to work it out.

Sources

  1. FDA / eCFR. 21 CFR Part 820, Quality Management System Regulation (current text: 820.1, 820.3, 820.7, 820.10, 820.35, 820.45). 2026 — Title 21 up to date as of October 13, 2026
  2. FDA / eCFR (historical version). 21 CFR 820.72 Inspection, measuring, and test equipment (former Quality System regulation). 2026 — Version of January 15, 2026; removed on February 2, 2026
  3. Federal Register. Medical Devices; Quality System Regulation Amendments (final rule), 89 FR 7496. 2024 — Published February 2, 2024; effective February 2, 2026. Pages 7496, 7500, 7505, 7518–7519 cited
  4. FDA. Quality Management System Regulation (QMSR). 2026 — Content current as of February 2, 2026
  5. FDA. Quality Management System Regulation: Frequently Asked Questions. 2026 — Content current as of October 2, 2026; questions 5, 7, 8, 9, 12 and 13 cited
  6. FDA. Compliance Program 7382.850, Inspection of Medical Device Manufacturers (Attachment A). 2026 — Implementation date February 2, 2026; pp. 43, 71 and 76
  7. MDSAP (hosted by FDA). MDSAP AU P0002.009, MDSAP Audit Approach (Task 8, p. 26; Tasks 13–15, pp. 107–109). 2024 — Revision of August 6, 2024, still the file at this FDA URL. The current revision is AU P0002.011 (August 3, 2026, on mdsap.global); Tasks 13–15 not yet checked against it
  8. FDA / eCFR. 21 CFR Part 211, Current Good Manufacturing Practice for Finished Pharmaceuticals (211.68, 211.160, 211.180, 211.194). 2026
  9. FDA. Questions and Answers on Current Good Manufacturing Practice Requirements: Laboratory Controls (question 1). 2022 — Content current as of November 16, 2022
  10. FDA / ICH. Q7 Good Manufacturing Practice Guidance for Active Pharmaceutical Ingredients, Rev. 1 (§5.3 Calibration). 2016
  11. FDA. Warning Letter 320-18-71, Les Emballages Facoteck Inc. (observation 3). 2018
  12. FDA / eCFR. 21 CFR Part 58, Good Laboratory Practice for Nonclinical Laboratory Studies (58.61, 58.63). 2026
  13. FDA. FDA Bioresearch Monitoring Information: Equipment (FDA, EPA and OECD GLP compared). 2015
  14. FDA / eCFR. 21 CFR Part 11, Electronic Records; Electronic Signatures (11.1, 11.10, 11.50, 11.70). 2026
  15. FDA. Part 11, Electronic Records; Electronic Signatures: Scope and Application (guidance). 2003
  16. FDA. Computer Software Assurance for Production and Quality Management System Software (guidance). 2026
  17. FDA. Inspection Observations: data sets FY2024 and FY2025. 2025 — Devices, Drugs and Bioresearch Monitoring tabs; page content current as of December 16, 2025