An out-of-tolerance (OOT) calibration result means that, when the instrument arrived for calibration, its error at one or more test points was outside the specified limits. The adjustment that usually follows fixes the instrument, not the measurements you made with it. So you assess and record whether the results since the last good calibration were affected, and act on the instrument and on any affected product according to risk (ISO 13485 clause 7.6, as audited under MDSAP; ISO/IEC 17025 clause 7.10 for labs). This guide walks through that assessment in seven steps, with a worked example, an annotated report and a printable template.
What “out of tolerance” means (and what it doesn’t)
A calibration compares the instrument with reference standards and reports the error at each test point. If an as-found error (before any adjustment) is beyond the limits that apply, the instrument was out of tolerance. In the terms of the decision-rule guidance, the item “failed”: at least one test point gave a value beyond the tolerance (Global ACI-TECH-1-002, formerly ILAC G8, Appendix B, Example 1). The limits can be the manufacturer’s specification, a legal maximum permissible error (VIM 4.26, quoted in ILAC G24, §3.13), or a tolerance you set for your own use.
The phrase is used for several different things, and search results mix them up:
| Term | What it describes | Example |
|---|---|---|
| Out of tolerance (calibration) | A calibration result: an as-found error beyond the limits | A pressure gauge reads +0.90 psi high against a ±0.75 psi tolerance |
| Out of calibration | An equipment status: calibration overdue, or status unknown | A caliper whose label is past its due date |
| OOS / OOT (pharmaceutical labs) | Out-of-specification or out-of-trend analytical results on a sample or batch | An assay result above the specification limit |
| Part out of tolerance | A product characteristic outside its drawing tolerance | A shaft diameter 0.02 mm over size |
This guide covers the first one: a measuring instrument that failed calibration, and what that means for everything it measured.
The out-of-tolerance process at a glance
Every framework that addresses OOT results asks for the same core: evaluate the effect on previous results, record it, and act on the equipment and on affected product (MDSAP Audit Approach, Tasks 13–14; ISO/IEC 17025 clause 7.10). The flow below turns that into seven steps. Two of them can end the process early, but only with a written rationale.
Calibration reports an as-found result out of tolerance
Confirm it is a real out-of-tolerance
Which tolerance, which decision rule, how large the uncertainty, and whether the failed points are in the range you use.
Out of tolerance where you use it?
No Record why, then return to service.
Take the instrument out of service and label it. Hold product in house that it measured.
From the last good calibration, or the last passing intermediate check, to the last use.
Products, lots, tests and other instruments calibrated with it inside the window.
Correct past results with the as-found error and compare them with the product limits, allowing for uncertainty.
Could any accepted result be wrong?
No No impact: record the evidence.
Re-test, rework, scrap or release with justification. Notify customers when product has shipped.
Root cause, interval review, corrective action, and a signed record.
Record approved and closed
Step 1: confirm it is a real out-of-tolerance
Before you open a long investigation, spend ten minutes on the certificate. Three questions decide whether this is an out-of-tolerance event for your use:
- Which tolerance was used? Labs usually check the manufacturer’s specification. If your own tolerance for the instrument is wider or narrower, compare the as-found errors with yours.
- What decision rule, and how large is the uncertainty? Whether a result is in or out depends on the decision rule (TECH-1-002, §5.2.3). Under a non-binary rule, a result just beyond the limit but within the guard band is a conditional fail, and close to the limit the chance that the instrument was actually in tolerance can approach 50 % (TECH-1-002, §5.2.3 and Appendix B, Example 2). A result beyond the limit by more than its expanded uncertainty is a clear fail.
- Did it fail where you use it? An error at 300 psi matters for tests done near 300 psi. If you only use the gauge from 0 to 100 psi and every point there passed, record that, with the evidence of how the gauge is used.
measured result ± expanded uncertainty U guard band (AL to TL) beyond the tolerance limit
A conditional fail is not a free pass. It means the evidence is uncertain, and the risk of the decision falls on whoever receives it (TECH-1-002, §1.2). When a lab reports an item as failed, it is the customer who has to investigate the impact on its products, which can lead to costly recalls (TECH-1-002, §6.4). That is a good reason to agree the decision rule with the lab before the calibration, as explained in our guide to reading a calibration certificate.
Step 2: contain the instrument and suspect product
Stop the instrument from being used, and stop suspect product from moving:
- Pull and label the instrument. Labs accredited to ISO/IEC 17025 must take equipment that is defective or outside specified requirements out of service, and isolate or clearly label it until it is shown to work correctly (clause 6.4.9, summarized by PJLA). Drug manufacturers must not use instruments that fail to meet their specifications (21 CFR 211.160(b)(4)), and FDA’s API guidance says the same (ICH Q7, §5.34).
- Hold product still in house that was measured or accepted with the instrument since its last good calibration. You can release it later if the assessment shows no impact.
- Check what else depends on it. If the instrument was used to calibrate or check other instruments, those results are suspect too.
Step 3: set the look-back window
This is the question most people ask first: how far back do I have to look? The answer is to the last time you know the instrument was good. FDA’s guidance for active pharmaceutical ingredients says that when a critical instrument deviates from its calibration standards, you should investigate whether the deviation could have affected product made since the last successful calibration (ICH Q7, §5.35). That usually means the date of the previous calibration with in-tolerance as-left results.
The end of the window is the last use before the instrument was taken out of service.
Without other data, you cannot tell when the instrument drifted out: it might have happened right after the previous calibration or just before this one (PJLA, 17025 §7.10 webinar). Intermediate checks change that. A documented check that passed, at the points you use, is evidence that the instrument was good on that date, so the window can usually start there instead (ISO/IEC 17025 clause 6.4.10; ILAC G24, §4.9–4.10).
No intermediate checks
Review window: all 12 months
Monthly intermediate checks (passing until month 10)
Review window: 2 months
in tolerance (calibration) intermediate check passed found out of tolerance results to review
Step 4: trace where the instrument was used
The window tells you when. Now you need where: every product, lot, test, inspection and calibration in which the instrument was used during the window. MDSAP auditors check that the effect of an out-of-tolerance condition was evaluated for devices in process, finished and already released, based on risk (MDSAP Audit Approach, Task 13).
This is where most investigations get stuck. Practitioners on r/Metrology describe users with decades in the business who have no idea what a failed instrument touched. Typical sources to rebuild the list:
- Inspection and test records that name the instrument’s asset ID.
- Check-out logs for shared gauges, and the instrument’s location.
- Calibration records of other instruments where it served as the reference.
- Batch, lot or traveler records for the processes it supported.
If you can’t link the instrument to specific results, the conservative assumption is that every result in the window could be affected. That is a strong argument for recording the asset ID on every inspection record. See our guide to calibration records.
Step 5: assess the impact (a worked example)
For each use, ask one question: knowing the as-found error, could any result you accepted have been outside its limits? The simplest method corrects past readings for the error and compares them again with the product limits:
- Take the as-found error at the test point closest to where the instrument was used, with its sign. Error = reading − reference, so a gauge that reads high gives readings that are higher than the true value.
- Correct each recorded result: true value ≈ recorded value − error.
- Compare the corrected results with the product limits, keeping a margin for the calibration uncertainty.
- Flag every result that could have crossed a limit.
This assumes the full as-found error existed throughout the window, which is conservative if the instrument drifted gradually. It is our worked illustration, not a method prescribed by a standard.
Example. Let’s continue with the digital pressure gauge from our example calibration certificate. Its as-found error was +0.90 psi at 300 psi (tolerance ±0.75 psi, U = 0.12 psi) and +0.18 psi at 75 psi. In the 12-month window, it was used for two tests:
| Use | Test and limits | Error at that point | Recorded results | Corrected results | Impact |
|---|---|---|---|---|---|
| A | Relief valve set pressure, 297.0 to 303.0 psi | +0.90 psi | 297.4 to 302.6 psi | 296.5 to 301.7 psi | Yes: valves recorded below 298.02 psi are suspect |
| B | Leak test hold pressure, 70 to 80 psi | +0.18 psi | 73.1 to 76.8 psi | 72.92 to 76.62 psi | No: far from both limits |
In use A, a valve recorded at 297.8 psi passed at the time, but its corrected value is 296.9 psi, below the lower limit. Allowing for the 0.12 psi uncertainty, any valve recorded below 297.0 + 0.12 + 0.90 = 298.02 psi is suspect. In this fictitious example, that is 31 valves.
range of recorded results suspect after correction margin of U = 0.12 psi outside product limits (psi)
A quantitative method for higher stakes. NASA’s measurement decision risk handbook includes a “calibration feedback analysis”. For each item tested with the out-of-tolerance instrument, it estimates the probability that the item was accepted while actually out of tolerance (the false accept risk), using the instrument’s estimated bias, the uncertainty of that bias projected back to the test date, the value measured on the item, the test process uncertainty and the item’s limits. The result supports a choice between recalling the item, notifying its user, or taking no action (NASA-HDBK-8739.19-4, Appendix G). In NASA’s worked example, a multimeter found 2.81 mV below nominal, against a ±2.5 mV tolerance, gives a 10.1 % false accept risk for a DVD player attribute tested with it 370 days earlier (Table G-2).
Step 6: decide, act and notify
With the impact analysis done, each group of suspect results needs a disposition:
| Finding | Typical action |
|---|---|
| No accepted result could have crossed a limit | No product action. Record the evidence and the rationale |
| Suspect product still in house | Keep on hold; re-test with a calibrated instrument; then release, rework or scrap |
| Suspect product already shipped | Handle it through your nonconformance process: risk assessment, customer notification, and correction or recall where needed |
| Other instruments calibrated with it | Repeat the assessment for each of them and for what they measured |
A few rules apply whatever your sector:
- A recall is a possible outcome, not the default. ISO/IEC 17025 asks labs to notify customers and recall work “where necessary” (clause 7.10.1, as summarized by PJLA and IAEA), and NASA’s method explicitly includes “no action” (Appendix G).
- Shipped product is in scope. Drug manufacturers must investigate unexplained discrepancies even when the batch has already been distributed, and extend the investigation to other batches that may be affected (21 CFR 211.192). For medical devices, the MDSAP audit looks at released devices too (Task 13). If affected devices have shipped, the decision feeds your nonconformance, correction and removal process.
- Write down who decided and why. In many companies, quality provides the data and engineering and management make the call on shipped product. When you lack data, give the customer the size and direction of the error and the affected serial numbers so they can assess their own risk.
Step 7: close out (root cause, interval and records)
Containing the event isn’t the end. Three things prevent the next one:
- Root cause. Drift, damage, overload, a dropped gauge, a method problem. MDSAP auditors review adjustment, calibration and maintenance records to see whether inadequate procedures, or not following them, contributed to the problem (Task 13). ISO/IEC 17025 asks labs to take corrective action when a nonconformity could recur (clause 7.10.3).
- Interval review. An OOT result is evidence that the interval may be too long. In ILAC’s “staircase” method, the interval is shortened whenever a calibration finds the deviation outside a set fraction of the maximum permissible error. ILAC warns that a very long interval risks having to withdraw many results already reported (ILAC G24, §6.2). Adjustments to intervals should be based on documented technical and statistical analysis, and NIST lists lack of funding or loss of competent staff among the inadequate reasons for adjusting one (NIST GMP 11). See our guide to calibration intervals.
- Intermediate checks. A periodic check at the critical points shortens the next look-back window from months to weeks (Step 3).
Finally, update the equipment record with the as-found values. For A2LA-accredited calibration labs, equipment records must include the measured value of every parameter found out of tolerance (A2LA I112).
When you are the calibration lab
The same logic applies inside an accredited calibration lab, with one difference: the affected “product” is your customers’ calibrations. When a reference standard is found out of tolerance, ISO/IEC 17025 treats it as nonconforming work (clause 6.4.9 leads into 7.10). The lab’s procedure must:
- define who is responsible and who can stop work or hold reports, based on the lab’s risk levels;
- evaluate how significant the nonconformity is, with an analysis of its impact on earlier results;
- decide whether the work is acceptable;
- notify customers and recall work where necessary;
- define who authorizes resuming work;
and the lab must keep records of these steps (clauses 7.10.1–7.10.2, as summarized by PJLA and IAEA).
PJLA, an accreditation body, puts it simply: the as-found condition is the condition in which the standard was used. The lab traces which customer calibrations and statements of conformity relied on the standard during the window (“reverse traceability”) and keeps a record of the investigation (PJLA, 17025 §7.10 webinar). One standard can sit under hundreds of customer certificates, so the cascade can be large.
Out-of-tolerance report: annotated example and template
An OOT report (also called an out-of-tolerance investigation or impact assessment record) is the evidence auditors ask for. The example below records the pressure gauge case from Step 5. It is fictitious: the company, lots and results are invented. Select a number to jump to its note.
Instrument
- Asset ID
- PG-014
- Description
- Digital pressure gauge, 0 to 300 psi
- Serial no.
- 21-0456
- Location
- Final test bench 3 (valve line)
As-found error (psi)
| Nominal | 0 | 75 | 150 | 225 | 300 |
|---|---|---|---|---|---|
| Error | +0.05 | +0.18 | +0.37 | +0.57 | +0.90* |
* +0.90 psi at 300 psi: beyond ± 0.75 psi even after allowing for U (0.78 > 0.75). Confirmed out of tolerance. Gauge reads high.
Containment
Gauge tagged "Out of tolerance: do not use" on return from the lab (2026-09-18) and kept off bench 3. Valves in stock tested since the last good calibration placed on hold.
Review window
- Last good calibration
- 2025-09-15 · ECL-2025-03920 (as left in tolerance)
- Intermediate checks
- None recorded
- Window reviewed
- 2025-09-15 to 2026-09-11 (last use)
Uses traced and impact analysis
Correction applied: true value ≈ recorded value − as-found error at the nearest test point. A result counts as acceptable only if the corrected value is at least U (0.12 psi) inside the product limits.
| Use | Product test | Volume in window | Limits (psi) | Error (psi) | Recorded (psi) | Corrected (psi) | Impact |
|---|---|---|---|---|---|---|---|
| A | Relief valve RV-300, set-pressure test | 1,240 valves, 18 lots | 297.0 to 303.0 | +0.90 | 297.4 to 302.6 | 296.50 to 301.70 | Yes* |
| B | Manifold MF-75, leak test hold pressure | 3,600 manifolds, 41 lots | 70.0 to 80.0 | +0.18 | 73.1 to 76.8 | 72.92 to 76.62 | No |
* Use A: 31 valves recorded below 298.02 psi are suspect (9 in stock, 22 shipped). Use B: corrected values stay well inside the limits; no impact.
Disposition and notification
In stock (9): retested with calibrated gauge PG-022: 7 pass, 2 fail and scrapped.
Shipped (22): both customers notified through NC-2026-041 with serial numbers, size and direction of the error; replacement offered.
Use B: no action, rationale recorded above.
Root cause and prevention
Span drift at the top of the range; no damage or misuse recorded.
- Gauge adjusted by the lab (as left in tolerance); returned to service 2026-09-22
- Interval shortened from 12 to 6 months
- Monthly check at 300 psi against PG-022 added (intermediate check)
- CAPA-2026-019 opened to review all top-of-range pressure gauges
A unique, traceable record
Give each out-of-tolerance event its own number, with open and close dates. Auditors look for a record of the assessment, not just a corrected certificate.
The instrument and where it lives
Asset ID, serial number and location. The location is the first clue to where it was used.
The calibration event and the as-found data
Link the certificate and copy the as-found errors. Here, +0.90 psi at 300 psi stays beyond the ±0.75 psi tolerance even after allowing for the 0.12 psi uncertainty, so it is a confirmed out-of-tolerance (step 1).
Containment
When the instrument was tagged and pulled, and which product was put on hold (step 2).
The review window
From the last good calibration to the last use. With no intermediate checks, the whole interval is in scope (step 3).
Uses traced and impact analysis
One row per use, with the error at the point where the gauge was used. Use A is affected because many valves were close to the lower limit. Use B is not, even though the same gauge measured it (steps 4 and 5).
Disposition and notification
What happened to suspect product in stock and to product already shipped, and who was told what (step 6).
Root cause and prevention
Why it drifted, the new interval, the intermediate check that will shorten the next window, and the corrective action (step 7).
Approval
Who decided and when. Quality brings the data; the decision on shipped product usually involves engineering and management.
Use the template below for your own investigations. Print it, or copy the fields into your quality system.
Out-of-tolerance impact assessment template
What each standard requires
Requirements differ by sector, but they converge on the same idea. For paid standards we describe requirements in our own words, based on the official guidance cited.
| Framework | What it expects after an out-of-tolerance result |
|---|---|
| ISO 9001 (clause 7.1.5.2) | When measuring equipment turns out to be unfit for its intended purpose, check whether earlier measurement results may be compromised, and act as needed |
| ISO 13485 (clause 7.6) and FDA QMSR | Evaluate and document whether earlier measurement results are still valid, and act on the equipment and on any product affected. MDSAP auditors check the effect on in-process, finished and released devices, based on risk (Tasks 13–14). FDA inspects medical device calibration under this clause since the QMSR took effect on February 2, 2026 (CP 7382.850); it replaced the former 21 CFR 820.72 |
| IATF 16949 (clause 7.1.5.2.1) | Adds record-keeping expectations, including the out-of-specification readings found at calibration and an assessment of the risk to the product’s intended use. Check the standard and your customer-specific requirements for details such as customer notification |
| ISO/IEC 17025 (clauses 6.4.9, 7.10) | Remove the equipment from service, examine the effect, and run the nonconforming work procedure: impact analysis on previous results, customer notification and recall of work where necessary, records |
| FDA drug CGMP (21 CFR 211.160(b)(4), 211.192) and ICH Q7 (§5.33–5.35) | Written remedial action when instruments fail; no use of failing instruments; thorough investigation of discrepancies even after distribution. ICH Q7 (guidance) recommends reviewing product made since the last successful calibration |
For the measuring-resources clause in depth, see our page on ISO 9001 clause 7.1.5.
Common myths
| Myth | What the sources say |
|---|---|
| “An OOT result means you must recall product.” | No source requires an automatic recall. You assess and document the impact by risk, and the outcome can be no action (MDSAP Tasks 13–14; ISO/IEC 17025 clause 7.10.1; NASA-HDBK-8739.19-4, Appendix G). |
| “Once it’s adjusted back into tolerance, the problem is closed.” | The adjustment fixes the instrument. The assessment is about the results made while it was in use. |
| “Look back one year” or “to the last audit”. | Look back to the last successful calibration (ICH Q7, §5.35). FDA rejected a shorter window in the 2013 warning letter above. |
| “The calibration lab assesses the impact on our product.” | The customer investigates the impact on its products (TECH-1-002, §6.4). The lab handles the impact of its own work (ISO/IEC 17025 clause 7.10). |
| “Errors under 10 % of tolerance can be ignored.” | No standard sets such a threshold. The impact depends on the product results and limits (NASA-HDBK-8739.19-4, Appendix G). |
| “A ‘Fail’ on the certificate always means it was out of tolerance.” | It depends on the decision rule. Near the limit, a conditional fail may have been in tolerance (TECH-1-002, §5.2.3 and Appendix B, Example 2). |
| “A certificate that says PASS or FAIL is enough.” | You need the as-found values and their uncertainty to assess impact (ISO/IEC 17025 clause 7.8.4.1; NASA-HDBK-8739.19-2, §10.2.7). |
| “21 CFR 820.72 requires…” | Since February 2, 2026, the QMSR incorporates ISO 13485 and section 820.72 no longer exists (21 CFR Part 820). |
Next steps: check how much margin your calibrations leave with the TUR calculator, and see how out-of-tolerance handling fits into a complete calibration program.
FAQ
Do I have to recall product after an out-of-tolerance calibration?
Not automatically. You have to assess and record whether the results made with the instrument since its last good calibration were affected, and act according to risk. The outcome can be no action, a re-test, a customer notification or a recall. What auditors look for is the assessment and the documented rationale, including for product already released.
How far back do I need to review?
To the last time you know the instrument was good: normally the last calibration with in-tolerance as-left results, or the last passing intermediate check that covers the affected range. FDA guidance for APIs says to look at everything made since the last successful calibration, and in a 2013 warning letter FDA rejected a review that covered a shorter period.
Does the calibration lab assess the impact on my product?
No. The lab reports the as-found results. Only you know where and how the instrument was used. Accreditation guidance on decision rules (Global ACI-TECH-1-002, formerly ILAC G8) notes that when a lab returns an item as failed, it is the customer who has to investigate the impact on its products. The lab is responsible for the impact of its own work, for example when one of its reference standards is found out of tolerance.
Is out of tolerance the same as out of calibration?
No. Out of tolerance describes a calibration result: the as-found error was beyond the limits. Out of calibration usually describes a status: the calibration is overdue or unknown. An instrument can be out of calibration and still within tolerance, or freshly calibrated and found out of tolerance.
Is there a percentage of tolerance below which an out-of-tolerance error can be ignored?
No standard sets one. Rules such as under 10 % of tolerance is negligible are internal conventions at best. The impact depends on how close your product results were to the product limits and on the risk of a wrong accept, which is why you correct the results and look at them.
Do I need to investigate when the instrument passed calibration?
Usually not as an out-of-tolerance event. But a pass near the limit, with a large calibration uncertainty or a low test uncertainty ratio, still carries a risk that the instrument was out of tolerance. Use the TUR calculator to see how large that risk is, and consider a guard band.
What label goes on an instrument that failed calibration?
The standards cited here do not prescribe a label text. ISO/IEC 17025 asks labs to take such equipment out of service and isolate or clearly label it until it is shown to work correctly. A common practice is to remove the calibration label, mark the instrument as failed or out of service, and give it a new due date only after it passes again.
Does a lab have to report its out-of-tolerance results to the accreditation body?
We found no such requirement. ANAB, for example, asks accredited labs to notify it promptly of unsatisfactory proficiency testing results, which is a different thing. Out-of-tolerance standards are handled through the lab's own nonconforming work process.
Sources
- MDSAP (hosted by FDA). MDSAP AU P0002.009, Audit Approach (Tasks 13–14: monitoring and measuring equipment; impact analysis of equipment found out of specification). 2024 — pp. 107–108
- FDA. Compliance Program 7382.850, Inspection of Medical Device Manufacturers (Attachment A). 2026
- eCFR. 21 CFR Part 820, Quality Management System Regulation. 2026
- eCFR. 21 CFR Part 211 (§211.160(b)(4), §211.192). 2026
- FDA / ICH. Q7 Good Manufacturing Practice Guidance for Active Pharmaceutical Ingredients, Rev. 1 (§5.3 Calibration). 2016
- FDA. Warning Letter OOWL-13-01, Allergy Laboratories, Inc. (observation 1.b). 2013 — Archived copy; the original fda.gov URL no longer resolves
- Perry Johnson Laboratory Accreditation (PJLA). ISO/IEC 17025:2017 Section 7.10 Nonconforming Work & Section 8.6 Improvement (webinar slides). 2021
- Perry Johnson Laboratory Accreditation (PJLA). Requirements in Section 6.4 Equipment (webinar slides, clause 6.4.9). 2023
- IAEA / SCK CEN. Nonconformity (training slides on ISO/IEC 17025 clause 7.10). 2021
- Global Accreditation Cooperation Inc. (Global ACI). Global ACI-TECH-1-002 (G), Guidelines on Decision Rules and Statements of Conformity (formerly ILAC G8:09/2019). 2026 — Same content as ILAC G8; sections renumbered +1
- ILAC / OIML. ILAC G24:2022 / OIML D 10:2022, Guidelines for the determination of recalibration intervals. 2022
- NASA. NASA-HDBK-8739.19-4, Estimation and Evaluation of Measurement Decision Risk (Appendix G, Calibration Feedback Analysis). 2010
- NASA. NASA-HDBK-8739.19-2, Measuring and Test Equipment Specifications (§10.2.7 Calibration Records). 2010
- NIST Office of Weights and Measures. GMP 11, Assignment and Adjustment of Calibration Intervals for Laboratory Standards. 2019
- A2LA. I112, Calibration Program Handbook
- ANAB. AR 2251, Accreditation Requirements: ISO/IEC 17025 Calibration Laboratories. 2026
- ISO/TC 176 and IAF. ISO 9001 Auditing Practices Group, Guidance on Measurement Traceability. 2016