How to write a calibration procedure (with an example and template)

What a calibration procedure must contain, section by section: scope, reference standards, test points, acceptance criteria, as-found and as-left, environment, uncertainty and ISO/IEC 17025 method validation. With an annotated example and a printable checklist.

Updated 20 primary sourcesHow we researchCite

On this page
  1. Calibration procedure vs program, work instruction and record form
  2. Do you need a written calibration procedure?
  3. What a calibration procedure should contain
  4. Example: an annotated calibration procedure for a digital pressure gauge
  5. How to choose test points
  6. Acceptance criteria: tolerance, TUR and decision rule
  7. Reference standards and equipment
  8. As found, adjustment and as left
  9. Environmental conditions
  10. Measurement uncertainty in the procedure
  11. Method verification and validation (ISO/IEC 17025 clause 7.2)
  12. Review, approval and document control
  13. Calibration procedure checklist (printable template)
  14. Where to find existing calibration procedures
  15. Common mistakes and myths
  16. Related guides
  17. FAQ
  18. Sources

A calibration procedure is the written, step-by-step method for calibrating one type of instrument: which parameters and ranges to check, with which reference standards, under what conditions, at which test points, how to decide pass or fail, and what to record. It has to be detailed enough for a qualified technician to carry out the calibration properly and in the same way every time (VIM 2.6; NASA-HDBK-8739.19-2, §10.2.1). It is not your calibration program, which covers the whole inventory, intervals and out-of-tolerance handling, and it is not the form you fill in. This guide covers what goes in each section, how to choose test points and acceptance criteria, as-found and as-left data, environmental conditions, method verification and validation under ISO/IEC 17025, and an annotated example with a printable checklist.

Calibration procedure vs program, work instruction and record form

These four documents get mixed up, and many “calibration SOP” templates online are actually program procedures or record forms. No standard defines all four terms. These are the working definitions we use across our guides:

Document What it answers Scope Example
Calibration program (program procedure, policy) Which items are calibrated, by whom, how often, and what happens when one fails All measuring equipment QP-07-01 Control of measuring equipment
Calibration procedure (method) How to calibrate this type of instrument, and how to judge the result One equipment type or family CP-PRS-002 Digital pressure gauges, 0–300 psi
Work instruction How to do one short task, often at the point of use One task WI-PRS-002A Monthly check at 300 psi
Record form Where the results of one calibration are written One calibration event F-PRS-002 Calibration record

Regulators expect both layers. FDA’s drug CGMP rule requires a written calibration program with specific directions, schedules, limits for accuracy and precision, and remedial action (21 CFR 211.160(b)(4)). MDSAP auditors accept a general program procedure plus more specific procedures for calibrating individual equipment (MDSAP Task 14, p. 108). In ISO/IEC 17025, a “method” can be read as a measurement procedure in the VIM sense (note to clause 7.2.1.1, as quoted by PJLA). For the program layer, see our guide to setting up a calibration program.

Do you need a written calibration procedure?

In practice, yes, if you calibrate anything in-house. The requirement is worded differently in each framework. For paid standards we describe requirements in our own words, based on the free sources cited.

Framework What it says about calibration procedures
ISO 9001 (clause 7.1.5) No explicit requirement for a written calibration procedure. It requires suitable resources, documented evidence of fitness for purpose and, where measurement traceability is required, calibration or verification against traceable standards at specified intervals or before use (ISO/TC 176 N1286; ISO 9001 APG). A procedure is the usual way to show that.
ISO/IEC 17025 (clause 7.2) Labs must use appropriate methods and procedures for all their activities, including uncertainty evaluation where relevant (7.2.1.1), keep them current and available to staff (7.2.1.2), and verify or validate them (7.2.1.5, 7.2.2). A standard that is clear enough for staff doesn’t need rewriting as an internal procedure (note to 7.2.1.3). Source: PJLA’s clause-by-clause slides.
ISO 13485 (clause 7.6) and FDA QMSR Documented procedures to make sure monitoring and measurement are carried out consistently with requirements; procedures must take environmental conditions into account. A general policy plus equipment-specific procedures is acceptable (MDSAP Tasks 13–14).
21 CFR 211.160(b)(4) (drug CGMP) A written program with specific directions, schedules, limits for accuracy and precision, and provisions for remedial action.
Former 21 CFR 820.72(b) (devices, until February 1, 2026) Calibration procedures had to include specific directions and limits for accuracy and precision. Replaced by the QMSR, which incorporates ISO 13485.
ANAB (accredited labs, AR 2251) A lab that calibrates its own equipment in-house needs, among other things, a documented procedure for each type of calibration (§5.1.5).
NVLAP Annex A (optional Z540-1 requirements) Procedures must state the range and tolerance or uncertainty of each parameter, describe the required standards generically, and be consistent with the accuracy required (NIST HB 150-2, A.3.1–A.3.3). Only applies to labs that ask for Z540-1 recognition.

Our ISO 9001 clause 7.1.5 guide and ISO/IEC 17025 guide cover the rest of each standard.

What a calibration procedure should contain

No official template exists. The structure below is our synthesis of four free sources: NASA’s list of what a laboratory calibration procedure typically includes (NASA-HDBK-8739.19-2, §10.2.1), EPA’s guidance on writing technical SOPs (EPA QA/G-6, §3–4), the optional Z540-1 requirements in NVLAP’s handbook (NIST HB 150-2, Annex A.3), and the layout of NIST’s own published procedures (NISTIR 6969, SOP 4: introduction and prerequisites, methodology, calculations, measurement assurance, uncertainty, certificate). NCSL International also publishes a recommended practice on calibration procedures (RP-3), which NASA cites for guidance; it is a paid document and we have not used it here.

Section What to write Basis
Header and document control Title, number, revision, effective date, page X of Y, approvals EPA §2.5, §3.1; ISO/IEC 17025 7.2.1.2
Purpose and scope Equipment type, models, parameters and ranges covered; limits of use; what is not covered EPA §4.1; NASA §10.2
References Manufacturer’s manual, published method, program procedure, decision rule, uncertainty budget, record form EPA §4.1
Safety and precautions Hazards (pressure, voltage, chemicals) and actions that could damage the instrument or invalidate results EPA §4.1
Personnel Training or qualification needed to perform and to review EPA §4.1; ANAB AR 2251 §5.1.2
Equipment and reference standards Required range and accuracy or uncertainty; ancillary equipment; allowed equivalents NASA §10.2.1; NIST HB 150-2 A.3.2
Environmental conditions Limits, stabilization and warm-up, what to record, what to do outside limits ISO/IEC 17025 6.3; NASA §10.2.1
Preliminary checks Inspection and functional tests; stop rules NASA §10.2.1; DKD-R 6-1 §5
Test points and sequence Points, direction, preloads, waiting times, mounting NASA §10.2.1; DKD-R 6-1 §7
Step-by-step instructions As found, adjustment criteria and method, as left NASA §10.2.1; EPA §1.4
Calculations Error, corrections, averaging, unit conversions EPA §4.1; NIST SOP 4 §3
Acceptance criteria and decision rule Tolerance per parameter and range; how uncertainty is applied NASA §10.2; NIST HB 150-2 A.3.1; Global ACI-TECH-1-002
Measurement uncertainty Budget to use, target uncertainty, minimum TUR ISO/IEC 17025 7.2.1.1; NIST GLP 9
Records, labels and out-of-tolerance Form, fields, where it is stored, labeling, what to do on a failure NASA §10.2.1; EPA §4.1
Revision history and review What changed, when, and the next review date EPA §2.3, §2.5–2.6

How much detail? EPA’s test is useful: someone with basic knowledge of the work but little experience should be able to follow it without supervision (EPA §2.1). The right level depends on how critical the measurement is, how often the procedure is used, how many people use it and how much training they get (EPA §3.0). NASA sets the same bar for in-situ procedures as for laboratory ones (§10.2.2).

Example: an annotated calibration procedure for a digital pressure gauge

The example below is an in-house procedure for digital pressure gauges of the same type as the gauge in our example calibration certificate and calibration record: 0–300 psi, 0.01 psi resolution, tolerance ±0.75 psi (±0.25 % of full scale). It is condensed onto one page; a real procedure would run to three or four. Everything in it is fictitious. Select a number to jump to its note.

1
Sample Manufacturing Co. · Calibration procedureCalibration of digital pressure gauges, 0–300 psi
Doc. no.
CP-PRS-002
Revision
C
Effective
2026-10-15
Page
Page 1 of 4
2

Purpose and scope

Digital gauge-pressure gauges, 0–300 psi, resolution 0.01 psi or finer, used to accept product or to set process pressure (for example Example Instruments DPG-300). In-house calibration in the metrology room.

Not covered: Absolute and differential gauges; transmitters with electrical output (CP-PRS-004); monthly intermediate checks (WI-PRS-002A).

Responsibilities: Performed by a technician qualified on this procedure (training record). Results reviewed by the metrology coordinator.

3

References

  • Example Instruments DPG-300 user manual, section 7 (zero and span adjustment)
  • DKD-R 6-1, Calibration of Pressure Gauges (method basis)
  • QP-07-01 Control of measuring equipment (program procedure)
  • DR-01 Decision rule · UB-PRS-002 Uncertainty budget
  • F-PRS-002 Calibration record form
4

Equipment and reference standards

Reference pressure standard
0–300 psi or wider; U ≤ 0.05 psi (k = 2); calibrated, traceable to the SI. Example: RS-003
Pressure source
Pneumatic hand pump or comparator, fine adjustment ≤ 0.01 psi
Ancillary
1/4 in NPT adapters, leak-free hoses, thermometer/hygrometer EM-01 (calibrated)

Other equipment may be used if it meets these specifications and has a current calibration.

5

Environmental conditions

Temperature
18–28 °C (64–82 °F), stable within ±1 °C during the run
Stabilization
Gauge in the room ≥ 2 h; powered on ≥ 15 min before starting
Record
Temperature and humidity at start and end on F-PRS-002
Outside limits
Do not calibrate
6

Preliminary checks

  • Visual: display, housing, thread and sealing face undamaged; label legible
  • Battery and self-test OK; zero function works
  • Leak check: hold 300 psi for 1 min, no visible drop

Fails a check: tag "out of service", do not adjust, record as found "not calibratable".

7

Test points and sequence

% FS0 %25 %50 %75 %100 %
Point (psi)075150225300
Tol. (psi)±0.75±0.75±0.75±0.75±0.75

Ascending 0 → 300 psi, then descending 300 → 0 psi. One preload to 300 psi (≥ 30 s) before the first series. Gauge mounted vertically, as in use.

8

Procedure

  1. Connect the gauge and the reference to the pressure source at the same height. Vent to atmosphere.
  2. Preload to 300 psi, hold ≥ 30 s, vent. Record the zero reading, then zero the gauge if the model allows (as users do before each use).
  3. As found: apply each point, wait ≥ 30 s, record reference and gauge readings. Do not adjust anything.
  4. Compute error = gauge reading − reference value at each point.
  5. If any |error| > 0.50 psi, adjust zero and span per the user manual, section 7. Otherwise, do not adjust.
  6. As left (only if adjusted): repeat the full sequence and record it as a separate set.
9

Acceptance criteria

Tolerance
±0.75 psi at every point (±0.25 % FS; process requirement)
Decision rule
DR-01: simple acceptance, valid only while TUR ≥ 4:1. If U > 0.18 psi (TUR below 4:1), use guarded acceptance (w = U).
Adjust if
Adjustment threshold: 0.50 psi (our choice, two thirds of the tolerance).
10

Measurement uncertainty

UB-PRS-002: reference standard, gauge resolution, repeatability, hysteresis (up vs down), temperature.

Target: U ≤ 0.18 psi (k = 2), so TUR = 1.50 / (2 × 0.18) ≥ 4:1.

11

Records, labels and out-of-tolerance

  • F-PRS-002: as-found and as-left readings, U, reference ID, conditions, technician, date
  • Status label; update the equipment record and next due date
  • As found out of tolerance: open an OOT report (QP-07-03) and notify the owner the same day
12

Approval and revision history

PreparedL. Gómez, metrology technician2026-09-29
Reviewed (trial run)D. Chen, metrology technician2026-10-06
ApprovedR. Patel, metrology coordinator2026-10-08
Rev. CDescending series and leak check added; DR-01 referenced (after OOT-2026-007)2026-10-15
Rev. BReference standard RS-003 replaces RS-0012024-05-02

Next review: 2028-10 (or sooner if the method, equipment or a standard changes) · Uncontrolled when printed · Fictitious example by acribi.com. Company, people, document numbers and values are invented.

  1. Document control

    A unique number, a revision letter, an effective date and "page X of Y" on every page, so nobody works from an old or incomplete copy. EPA's guidance puts the title, ID number, revision date and approval signatures on the cover page (EPA QA/G-6, §3.1). ISO/IEC 17025 expects methods to be kept up to date and available to staff, under document control (clause 7.2.1.2).

  2. Purpose, scope and what is not covered

    Which equipment type and range the procedure applies to, and where it stops. The "not covered" line prevents the most common misuse: applying a gauge procedure to a transmitter, or treating a quick check as a calibration. EPA lists scope, applicability and limits of use as the first topic of a technical procedure (§4.1).

  3. References

    The documents the procedure relies on: the manufacturer's manual (for adjustment), the published method it is based on, the program procedure, the decision rule, the uncertainty budget and the record form. Quote numbers and revisions, so a change in any of them triggers a review of this procedure.

  4. Equipment and reference standards

    Describe the reference standard by what it must achieve (range and uncertainty) rather than by one serial number, and allow equivalent equipment that meets the specification. The optional Z540-1 requirements in NVLAP's handbook ask for that generic description (NIST HB 150-2, Annex A.3.2), and US Army calibration procedures add the permission to use equivalents (TB 9-6625-2372-24, §4). Here, U ≤ 0.05 psi keeps the reference a small part of the 0.18 psi target.

  5. Environmental conditions

    Limits, stabilization time, what to record and what to do when conditions are outside limits. ISO/IEC 17025 requires documented environmental requirements and monitoring where they influence results (clauses 6.3.1–6.3.3). The 18–28 °C range and ±1 °C stability come from DKD-R 6-1, §6. The stabilization times are example values: set yours from the manufacturer's warm-up time and your own data.

  6. Preliminary checks

    Inspection and functional tests before any measurement: damage, cleanliness, leaks, battery, self-test (NASA-HDBK-8739.19-2, §10.2.1; DKD-R 6-1, §5). The stop rule matters: a gauge that fails here is not adjusted first and then calibrated, or the as-found information is lost.

  7. Test points and sequence

    Five points at 0, 25, 50, 75 and 100 % of full scale, up and down, is the good practice NASA describes for laboratory calibrations (§10.2.1). The descending series shows hysteresis. Mount the gauge as it is used (DKD-R 6-1, §7). For a gauge specified at ±0.25 % of span, DKD-R 6-1 would ask for more: its sequence B, with 9 points, 2 preloads and 3 series (Table 1).

  8. Step-by-step instructions

    Short, numbered, imperative steps that a qualified technician can follow without guessing (NASA §10.2.1; EPA §1.4 and §2.1). Note how the steps fix the things that otherwise vary between technicians: preload, zeroing, waiting time, the sign of the error, and the order of as-found, adjustment and as-left.

  9. Acceptance criteria and decision rule

    The tolerance at each point and how measurement uncertainty is handled at the limits. Procedures should state the performance criteria used to decide conformance (NASA §10.2), and Z540-1 requires the tolerance or uncertainty for each parameter (NIST HB 150-2, Annex A.3.1). The decision rule follows Global ACI-TECH-1-002 (formerly ILAC G8). The 0.50 psi adjustment threshold is our example choice, not a rule from any standard.

  10. Measurement uncertainty

    The procedure names the uncertainty budget it uses and a target uncertainty: an upper limit set by the intended use of the results (VIM 2.6 Note 2; 2.34). Here the target keeps the TUR at 4:1 or better (1.50 / (2 × 0.18) ≈ 4.17), so the simple-acceptance rule in note 9 stays valid. The TUR uses the expanded uncertainty of the whole calibration, not that of the reference standard alone: 0.05 psi is one input to the budget, 0.18 psi is the ceiling for the result. The accredited lab that calibrates PG-014 itself reports U = 0.12 psi, a TUR of 6.2:1 (see the example certificate); the in-house target is a looser ceiling, so its minimum TUR is lower. Check yours with the uncertainty calculator and the TUR calculator.

  11. Records, labels and out-of-tolerance

    What gets recorded, on which form, and what happens next. NASA asks procedures to include instructions for recording, storing and analyzing data, and for labeling and handling the calibrated item (§10.2.1). The out-of-tolerance line links this procedure to your out-of-tolerance process; the form is described in our calibration records guide.

  12. Approval, revision history and review date

    Who wrote it, who tried it out, who approved it, what changed in each revision and when it is due for review. EPA recommends that someone other than the author tests a procedure before approval and that procedures are reviewed periodically (§2.2–2.6). The two-year review date here is an example, not a requirement.

Example calibration procedure for a fictitious company. Document numbers, people and values are invented; the values are chosen to be internally consistent. Source: structure based on NASA-HDBK-8739.19-2 §10.2.1, EPA QA/G-6 and NIST HB 150-2 Annex A; method elements from DKD-R 6-1; example values by Acribi.

Three choices make this procedure hold up. The reference standard is specified by performance, so replacing it doesn’t require a new revision. The steps force as-found data first and only allow an adjustment past a written threshold. And the target uncertainty is tied to the decision rule: if the calibration can’t reach U ≤ 0.18 psi, a simple pass/fail against ±0.75 psi is no longer valid, and the procedure says what to do instead.

How to choose test points

Test points decide what the calibration can prove. A gauge checked only at zero and full scale tells you nothing about the middle of the range, where it may be used every day. The sources agree on the principle and differ in effort:

Source Test points and sequence Applies to
NASA-HDBK-8739.19-2 §10.2.1 Enough points, including zero, to show conformance over the range used; good practice is one full ascending and descending cycle with 4 or 5 points each way (0, 25, 50, 75, 100 % of full scale) Laboratory calibration, any instrument
NASA-HDBK-8739.19-2 §10.2.2 Where many points aren’t practical, at least one up-and-down cycle with 3 points each way (0, 50, 100 %) In-situ calibration
DKD-R 6-1, §7 and Table 1 Points evenly spread over the range. Sequence A (uncertainty below 0.1 % of span): 9 points, 3 preloads, 2 up and 2 down series. B (0.1–0.6 %): 9 points, 2 preloads, 2 up and 1 down. C (above 0.6 %): 5 points, 1 preload, 1 up and 1 down. At least 30 s between steps Pressure gauges, transducers and transmitters
EURAMET cg-15, §3.4.1 A minimum set that reflects the working principle, every point the manufacturer requests and the customer’s needs; tables of minimum points per function and range Digital multimeters

Three more rules from the same sources:

  • Calibrate it the way it is used. If the instrument works with an accessory or fixture, calibrate it in that configuration (EURAMET cg-15, §3.3.7), and respect the mounting position (DKD-R 6-1, §7).
  • Calibrate at use conditions when they matter. Equipment used at process temperature may need calibration at that temperature (NASA §10.2.1).
  • Cover what you actually measure. If a caliper is used mostly at 150 mm, a test point near 150 mm is worth more than one at 10 mm. A caliper that reads zero with the jaws closed has been checked at one point, not calibrated over its range (NASA §10.2.1).

Acceptance criteria: tolerance, TUR and decision rule

A procedure without acceptance criteria produces numbers, not decisions. It should state the range and the tolerance or uncertainty for each parameter calibrated (a requirement under Z540-1: NIST HB 150-2, Annex A.3.1) and the performance criteria for deciding conformance (NASA §10.2). FDA’s drug rule asks for limits for accuracy and precision in the written program (21 CFR 211.160(b)(4)).

Where the tolerance comes from. Pick one source and write it down next to the number:

  • the manufacturer’s specification (the default for most instruments);
  • a product standard for that instrument type, such as ASME B89.1.14 for calipers, ISO 6789 for torque tools or ISO 8655 for piston pipettes;
  • a process requirement: a tighter or wider limit derived from what the instrument measures. If you set a wider limit than the manufacturer’s, restrict the instrument’s use to work where that limit is good enough, and label it.

How uncertainty is handled at the limits. If the procedure leads to a pass or fail statement, ISO/IEC 17025 expects a decision rule that takes measurement uncertainty into account, documented and, for customers, agreed (Global ACI-TECH-1-002, which replaced ILAC G8 in 2026 with the same content). The choice is between simple acceptance, guarded acceptance with a guard band, and non-binary rules. Our guard banding guide explains each one.

The TUR. Many procedures require a minimum test uncertainty ratio, often 4:1, and allow simple acceptance only above it. That is a sound way to write it, as long as you know what it is: A2LA calls a TUR of 4:1 or greater the generally accepted industry standard (A2LA G136). Use the TUR calculator to check yours.

Reference standards and equipment

Describe what the reference standard must achieve, not just which one you own: the parameter, range, and tolerance or uncertainty, optionally with representative models (NIST HB 150-2, Annex A.3.2). US Army calibration procedures show the format well: a table of minimum equipment specifications chosen to give a 4:1 ratio, with the actual accuracy stated where 4:1 can’t be reached, and permission to use alternatives that meet the minimum and are in calibration (TB 9-6625-2372-24, §4). That is Army practice, not a general requirement, but it keeps the procedure valid when equipment changes.

NIST’s mass procedures add the prerequisites that belong at the start of any procedure: valid calibration certificates with values and uncertainties for every standard, traceability to the SI, standard uncertainties small enough for the level of calibration, and measuring equipment in good operating condition (NISTIR 6969, SOP 4, §1). ISO 9001 states the traceability part for any organization where measurement traceability is required: calibration or verification against standards traceable to international or national measurement standards, and a record of the basis used where none exist (clause 7.1.5.2).

As found, adjustment and as left

Calibration and adjustment are different operations. Calibration compares the instrument’s indications with reference values; adjustment changes the instrument so it reads closer to them, and after an adjustment the instrument normally has to be calibrated again (VIM 2.39 Note 2; 3.11 Note 3). When a procedure includes adjustments, the organization should keep records of the as-found and as-left conditions, and it is good practice to keep the measured values with their uncertainty (NASA §10.2). Calibration certificates must report results before and after adjustment or repair, where available (ISO/IEC 17025 clause 7.8.4.1), and ANAB expects before-and-after data even from non-accredited suppliers in the traceability chain (ANAB AR 2251, §2.2.3).

EURAMET’s multimeter guide describes three sequences, and a procedure should say which one applies:

Sequence Steps When
a) Standard Calibrate (as found) → adjust → calibrate (as left) Instruments that can be adjusted. The first calibration shows how the instrument behaved since the last one
b) Adjust first Adjust → calibrate Only when the arrival state doesn’t matter: a first calibration, or right after a repair, and only at the customer’s explicit request or after a repair by the lab
c) No adjustment Calibrate only Instruments that can’t be adjusted, or whose results are well inside limits

Source: EURAMET cg-15, §3.3.1–3.3.4. For sequence c), cg-15 gives an example of limits for skipping the adjustment on multimeters: 70 % of the annual specification at all points for less accurate instruments, 50 % for more accurate ones (§3.3.4). It is an example for one instrument class, not a general rule; set your own threshold and write it in the procedure. The adjustment itself follows the manufacturer’s method (§3.3.5). If repair or adjustment is needed just to make the instrument calibratable, agree it with the owner and document the parameters before and after (DKD-R 6-1, §5).

The sequence in the example procedure, as a decision flow:

  1. Preliminary checks

    Inspection and functional tests.

    Fails: tag out, record as found "not calibratable"

  2. As-found calibration

    Full sequence of test points, no adjustment of any kind. Record every reading.

  3. Compare with the tolerance

    Apply the decision rule at every point.

    Any point out of tolerance: open an out-of-tolerance report

  4. Decide whether to adjust

    Adjust only if a result passes the written threshold (0.50 psi in the example), using the manufacturer's method.

  5. As-left calibration

    If adjusted, repeat the full sequence and record it as a separate set. If not, the as-found results are also the as-left results.

  6. Record and label

    Results, uncertainty, standard, conditions, technician and date; status label; next due date.

Environmental conditions

ISO/IEC 17025 requires facilities and environmental conditions that don’t invalidate results, documented environmental requirements, and monitoring, control and recording where the method requires it or conditions influence the results, including on-site work (clauses 6.3.1–6.3.5, as quoted by PJLA). ISO 13485 asks for the same consideration in equipment procedures (clause 7.6). In a procedure, that means four things: the limits, the stabilization or warm-up time, what to record, and a rule that you don’t calibrate outside the limits. PJLA’s slides on clause 6.3 show exactly this kind of wording as an example.

Typical values from the sources, all as examples rather than universal requirements:

Source Temperature Other conditions
PJLA example (clause 6.3) 68 °F ± 2 °F Minimum humidity; item stabilized in the lab at least 12 h; no calibration outside limits
DKD-R 6-1, §6 (pressure gauges) 18–28 °C, variation during calibration recommended within ±1 K Thermal equilibrium; warm-up time per manufacturer or experience; temperature recorded
NISTIR 6969, SOP 4, Table 1 (mass, by echelon) e.g. 18–23 °C, max. change ±2 °C in 12 h and ±1.5 °C per hour Relative humidity 40–60 %; stable for 24 h before calibration

How tight to go depends on where the work sits in the traceability chain: the higher the standard, the tighter the control (NASA §10.2.1).

Measurement uncertainty in the procedure

ISO/IEC 17025 expects appropriate methods for evaluating measurement uncertainty where relevant (clause 7.2.1.1). NIST goes further: a calibration is not complete until the expanded uncertainty has been determined and reported, and every NIST SOP includes how to calculate it (NISTIR 6969, GLP 9). Published procedures from NIST, EURAMET and DKD all carry their own uncertainty section.

In practice, the procedure should name the uncertainty budget it uses (or include it), list the contributions to evaluate, and state a target uncertainty: the upper limit set by how the results will be used (VIM 2.6 Note 2; 2.34). Tie it to the decision rule, as in the example. Our measurement uncertainty guide explains budgets, and the uncertainty calculator builds one.

Method verification and validation (ISO/IEC 17025 clause 7.2)

Clause 7.2 is the part of ISO/IEC 17025 that most calibration procedure guides skip. It separates three situations:

Situation What you must do Records
Adopt a published standard method that is clear enough for your staff Use it as written, at its latest valid edition; add details only where needed for consistent application (7.2.1.3) The method itself, under document control
Introduce any method, including standard and manufacturer methods Verify that you can perform it properly and achieve the required performance before use, and again if the issuer revises it (7.2.1.5) Records of the verification
Use a method you developed, a non-standard method, or a standard method outside its scope or modified Validate it as extensively as the application needs (7.2.2.1); revalidate if a change affects it (7.2.2.2) Validation procedure, requirements, performance characteristics, results and a statement of validity (7.2.2.4)

Source: ISO/IEC 17025:2017 clauses 7.2.1.3–7.2.2.4, as quoted in PJLA’s slides on clause 7.2.

Techniques for validation include calibration or evaluation of bias and precision with reference standards or materials, comparison with other validated methods, interlaboratory comparisons, systematic assessment of influence factors, robustness tests and uncertainty evaluation (note 2 to clause 7.2.2.1). PJLA suggests the same kinds of techniques for verification. The performance characteristics to check can include range, accuracy, uncertainty, linearity, repeatability, reproducibility, robustness and bias, as far as they matter to the customer (7.2.2.3). If a method doesn’t specify environmental conditions that influence the results, document them in your procedure anyway: that follows from clause 6.3.

Three more points from clause 7.2:

  • Manufacturer’s methods are acceptable. The standard lists methods specified by the equipment manufacturer among the recommended ones, next to those in international, regional or national standards (7.2.1.4). They still need verification.
  • Deviations from a method are allowed only if documented, technically justified, authorized and accepted by the customer (7.2.1.7).
  • Third-party procedures need checking. NASA notes that ASTM and ISA procedures are peer-reviewed by standards committees, while procedures posted on the NIST and GIDEP websites don’t go through an approval process. Its advice: treat every third-party procedure as guidance that needs verification and validation before use (NASA §10.2.1, p. 81).

Review, approval and document control

Procedures fail audits less often for what they say than for how they are controlled. EPA’s guidance on SOPs covers the essentials (EPA QA/G-6, §1.4 and §2–3):

  • Written by people who do the work, in active voice and present tense, step by step and without ambiguity. Flowcharts help.
  • Tried out by someone other than the author before approval, then reviewed and approved by qualified people.
  • Controlled: document number, revision and date on every page, signatures (electronic ones are accepted), obsolete versions withdrawn and archived.
  • Reviewed periodically, and updated whenever the process changes. EPA gives one to two years as an example of a review period, not a requirement. ISO 9001 and ISO/IEC 17025 set no frequency.
  • Checklists support the procedure; they don’t replace it.

Under ISO/IEC 17025, methods and the documents that support them must be kept current and available to staff (clause 7.2.1.2), and a revised method needs its verification repeated as far as necessary (7.2.1.5). A good trigger for review is any change of reference standard, equipment model, referenced standard edition or tolerance, and every out-of-tolerance investigation that points at the method.

Calibration procedure checklist (printable template)

Use this list to draft a new procedure or to review an existing one before an audit. Each line maps to a section of the structure above. Print it, or copy it into your own template. It is free, with no sign-up.

Calibration procedure checklist

Document control and scope
Equipment and conditions
Method
Decision and uncertainty
Records and control

Where to find existing calibration procedures

You rarely need to start from a blank page. These sources are free unless noted, and all of them are starting points that you verify and adapt, not ready-to-use procedures for your lab:

Source What you get Notes
NIST Office of Weights and Measures SOPs, good measurement practices and good laboratory practices for mass, volume and length US public domain. Published without a formal approval process (NASA §10.2.1)
EURAMET calibration guides (cg-xx) Method guidance for multimeters, weighing instruments, pressure balances, thermocouples and more Free; copyrighted, cite and don’t redistribute
DKD guidelines (DKD-R) Detailed methods, for example DKD-R 6-1 for pressure gauges Free; licensed for non-commercial use
US Army calibration procedures (TB 9-6625 series) Complete procedures for specific test equipment, in a three-section format (identification, equipment, calibration process) Only those marked Distribution Statement A are public. Check that you have the current revision
Manufacturer manuals and service notes Performance tests and adjustment instructions Many manufacturers don’t publish full calibration procedures (NASA §10.2.1)
NCSL International RP-3 Recommended practice on calibration procedures Paid

For instrument-specific procedures, see our instrument guides as they are published, starting with pressure gauges, scales and thermometers.

Common mistakes and myths

Myth or mistake What the sources say
“ISO/IEC 17025 requires a 4:1 (or 3:1) TUR.” It requires a decision rule that accounts for uncertainty. 4:1 is industry practice (A2LA G136; Global ACI-TECH-1-002).
“ISO 9001 requires a written calibration procedure.” It requires evidence of fitness for purpose and traceable calibration or verification (clause 7.1.5). A procedure is the usual means, not a stated requirement.
“Standard methods don’t need checking.” Every method must be verified before use (ISO/IEC 17025 7.2.1.5).
“We must rewrite every standard as an internal SOP.” Not if it is clear enough for staff (note to 7.2.1.3), though EPA notes a citation alone may not say how you do it.
“Procedures must be reviewed every year.” No standard we checked sets a frequency. EPA’s one to two years is an example.
“Don’t adjust if results are within 70 % of spec.” An example in EURAMET cg-15 for multimeters, not a general rule.
“If it reads zero, it’s calibrated.” Zero is one point. Conformance needs points across the range used (NASA §10.2.1).
“21 CFR 820.72 requires calibration procedures.” 820.72 was replaced on February 2, 2026. Device calibration now falls under ISO 13485 clause 7.6 through the QMSR.
“A NIST-published procedure is approved and ready to use.” NIST and GIDEP procedures don’t go through an approval process; verify and validate them (NASA §10.2.1).

FAQ

What should a calibration procedure include?

At minimum: the equipment it covers and its specification limits, the reference standards and other equipment needed (with their accuracy or uncertainty), environmental conditions, preliminary checks, the test points and step-by-step instructions, the acceptance criteria and decision rule, how to handle adjustment and as-found and as-left data, how to evaluate uncertainty, and what to record and how to label the instrument. Add document control: number, revision, approval and review date.

Can I use the manufacturer's calibration procedure?

Yes. ISO/IEC 17025 lists methods specified by the equipment manufacturer among the acceptable ones (clause 7.2.1.4). You still have to verify that you can perform it correctly before you use it (clause 7.2.1.5), and add what it leaves out, such as your tolerance, as-found recording and records. Many manufacturers publish adjustment instructions but no full calibration procedure: a procedure that adjusts before measuring loses the as-found data.

Do I need to validate a calibration procedure?

Under ISO/IEC 17025, you validate methods you developed yourself, non-standard methods, and standard methods used outside their scope or modified (clause 7.2.2.1). A published standard method used as intended only needs verification: evidence that your lab can perform it and get the required performance (clause 7.2.1.5). Keep records of either.

What are as-found and as-left data?

As-found data are the results measured when the instrument arrives, before anyone touches it. As-left data are the results after adjustment or repair. You need both when an instrument is adjusted: as-found tells you whether past measurements were right, as-left shows the state the instrument goes back into service in. If nothing is adjusted, one set of results serves as both.

How many test points should a calibration procedure use?

Enough to cover the range you use, including zero. NASA calls one ascending and one descending cycle with four or five points each way (for example 0, 25, 50, 75 and 100 % of full scale) good practice for laboratory calibrations. Instrument-specific guides go further: DKD-R 6-1 for pressure gauges asks for 5 or 9 points and two to four measurement series (up and down), depending on the uncertainty you need.

What tolerance should I put in a calibration procedure?

The tolerance the instrument must meet for its use: the manufacturer's specification, a product standard for that type of instrument, or a tighter limit derived from your process. Write it down per parameter and range, together with the decision rule that says how measurement uncertainty is handled at the limits.

Who should write calibration procedures?

Someone who knows the measurement and does the work, with review by a second qualified person. EPA's guidance for writing procedures recommends that experienced people write them and that someone other than the author tries them out before approval.

How often should calibration procedures be reviewed?

Whenever the method, equipment or a referenced standard changes, and periodically. ISO 9001 and ISO/IEC 17025 set no fixed frequency. EPA gives every one to two years as an example of a periodic review, not as a requirement.

Does ISO 9001 require a written calibration procedure?

Not in so many words. ISO 9001 requires suitable measuring resources, documented evidence that they are fit for purpose and, where measurement traceability is required, calibration or verification against traceable standards at specified intervals or before use (clause 7.1.5). A written procedure is the usual way to make that repeatable and to show an auditor how it is done.

Where can I find calibration procedures for my equipment?

Start with the manufacturer's manual and service documentation, then free public sources: NIST Office of Weights and Measures SOPs for mass, volume and length, EURAMET calibration guides, DKD guidelines, and US Army calibration procedures released for public distribution. Treat any third-party procedure as a starting point that you verify and adapt.

Sources

  1. JCGM / BIPM. International Vocabulary of Metrology (VIM), JCGM 200:2012. 2012 — 2.6 measurement procedure, 2.34 target measurement uncertainty, 2.39 calibration, 3.11 adjustment
  2. NASA. NASA-HDBK-8739.19-2, Measuring and Test Equipment Specifications. 2010 — §10.2 (pp. 78–79), §10.2.1 (pp. 79–81), §10.2.2 (pp. 81–82), §10.2.7
  3. US EPA. EPA QA/G-6, Guidance for Preparing Standard Operating Procedures (EPA/600/B-07/001). 2007 — §1.4, §2.1–2.6, §3.0–3.1, §4.1
  4. Perry Johnson Laboratory Accreditation (PJLA). Requirements in Section 7.2, Selection, Verification and Validation of Methods (webinar slides). 2023 — ISO/IEC 17025:2017 clauses 7.2.1.1–7.2.2.4
  5. Perry Johnson Laboratory Accreditation (PJLA). A look at ISO/IEC 17025 Sections 6.3 & 6.4 (webinar slides). 2022 — ISO/IEC 17025:2017 clauses 6.3.1–6.3.5 (with an example of environmental requirements in a procedure), 6.4.3
  6. NIST NVLAP. NIST HB 150-2e2024, NVLAP Calibration Laboratories. 2024 — §7.2 and Annex A (optional ANSI/NCSL Z540-1 requirements), A.3.1–A.3.3
  7. ANAB. AR 2251, Accreditation Requirements: ISO/IEC 17025 Calibration Laboratories. 2026 — §2.2.3, §5.1.1–5.1.7 (in-house calibrations)
  8. NIST Office of Weights and Measures. NISTIR 6969-2019, Selected Laboratory and Measurement Practices and Procedures to Support Basic Mass Calibrations. 2019 — SOP 4 (weighing by double substitution), GLP 9
  9. NIST Office of Weights and Measures. Calibration Procedures (SOPs, GMPs and GLPs)
  10. PTB / DKD. DKD-R 6-1, Calibration of Pressure Gauges (03/2014, revision 3). 2014 — §5 (note), §6, §7 and Table 1
  11. EURAMET. EURAMET Calibration Guide No. 15, Guidelines on the Calibration of Digital Multimeters (v3.0). 2015 — §3.3–3.4
  12. EURAMET. EURAMET Calibration Guide No. 18, Guidelines on the Calibration of Non-Automatic Weighing Instruments (v4.0). 2015
  13. Department of the Army. TB 9-6625-2372-24, Calibration Procedure for Nav/Comm Test Set, Aeroflex Model IFR 4000. 2012 — Distribution Statement A; cited as an example of format only (Sections I–III, equipment table)
  14. Global Accreditation Cooperation (Global ACI). Global ACI-TECH-1-002 (G), Guidelines on Decision Rules and Statements of Conformity (formerly ILAC G8:09/2019). 2026
  15. A2LA. G136, Guidance on Decision Rules in Calibration. 2022
  16. MDSAP (hosted by FDA). MDSAP AU P0002.009, Audit Approach (Tasks 13–14, monitoring and measuring equipment). 2024
  17. eCFR. 21 CFR Part 211 (§211.160(b)(4)). 2026
  18. eCFR (historical version). 21 CFR Part 820 as in force on January 15, 2026 (former §820.72). 2026 — Superseded on February 2, 2026; cited as history
  19. ISO/TC 176/SC 2 (copy hosted by IAQG). ISO/TC 176/SC 2/N1286, Guidance on the requirements for Documented Information of ISO 9001:2015. 2016
  20. ISO/TC 176 and IAF. ISO 9001 Auditing Practices Group, Guidance on Monitoring and measuring resources. 2016