A scale calibration record (or the certificate a lab gives you) has to show seven things:
- Which scale and where: asset ID, model, serial number, capacity (Max), display step d and, if marked, e and the accuracy class, and the location.
- The tolerance and its source: your process requirement, the manufacturer’s specification, or the legal tolerance for the class, which steps up with the load (NIST Handbook 44, Table 6; OIML R 76).
- The weights and the method: test weights with a valid certificate, their class, your procedure, and the conditions.
- The three tests (EURAMET cg-18, §5): errors at five or more loads across the range, repeatability with one load at least five times, and eccentricity with about a third of Max at the center and four corners.
- As found and as left: results before any adjustment and, after it, a new set; plus whether the scale was adjusted with its internal weight or an external one.
- The uncertainty of each error and the decision rule for pass or fail.
- The decision: in or out of tolerance, who did it and reviewed it, and the next due date.
The free template below has a field for each of these and fits on one Letter page. Print it from this page: no sign-up needed. Below it, the same form filled in with a worked example, then what to look for on a lab certificate.
Scale calibration record template
Scale and balance calibration record
One sheet per scale. The three EURAMET cg-18 tests with traceable test weights, as found and as left.
Use one sheet per scale. For a scale calibrated by an outside lab or service company, copy the key results from the certificate, record your own decision on the last block and attach the certificate. For in-house calibration, the sheet is the record. The loads, readings and positions follow EURAMET cg-18 as described on our scale calibration procedure page; it works for balances, bench scales and platform scales alike.
Example: a filled-in scale calibration record
The record below is for the same fictitious bench scale as the second worked example on our procedure page: same loads and indications, and every error, tolerance and pass/fail is calculated from them. The repeatability readings, the as-left eccentricity, the company, people and uncertainties are invented. Click a number to jump to its note.
Scale and balance calibration recordExample
One sheet per scale. The three EURAMET cg-18 tests with traceable test weights, as found and as left.
Which scale, and where
Max, the display step d and, for a scale with an accuracy class, the verification interval e: the tolerance is counted in e. SC-07 is a Class III bench scale, 15 kg × 5 g, so n = Max ÷ e = 3,000. Write down the location too: a certificate shouldn't be accepted for a scale moved after calibration, unless it has been shown to be immune to the move (EURAMET cg-18, 4.1.2).
The tolerance steps up with the load
Handbook 44 maintenance tolerance for Class III: 1 e (5 g) up to 500 e, 2 e (10 g) up to 2,000 e and 3 e (15 g) above, up to 4,000 e (Handbook 44, 2.20, Table 6). The acceptance tolerance, half as large, applies to a commercial scale's first official test within 30 days of corrective service (G-T.1, T.N.3.2). Write the steps on the record, not just the class.
Weights, warm-up and conditions
Class M1 weights suit a Class III scale (NIST HB 105-1, Table 1). Their error plus uncertainty must stay below one third of the tolerance at each load (Handbook 44, App. A 3.2): 5 g at 15 kg here. Air temperature recorded at least once (cg-18, 5.4); the scale warmed up and loaded to Max before the readings, which are not recorded (cg-18, 4.1.3).
Errors of indication: fails above 10 kg
Five loads from the minimum (20 e) to Max, one reading each, original indications written down (cg-18, 4.4.1). As found, the scale read high, increasingly with load: +20 g at 11 kg and +25 g at 15 kg, both outside ±15 g. After a span adjustment, the largest error is +5 g.
Repeatability
The same 7.5 kg load five times, zero checked after each removal (cg-18, 5.1). The standard deviation, 2.2 g, is under half a division. Compare it with the repeatability in the scale's specification: the 5 g limit here is an invented manufacturer value.
Eccentricity
A 5 kg load (Max/3) in the center and at the four corners (cg-18, 5.3). As found, the front-right reading was one division (+5 g) above the center, inside the 10 g tolerance at that load.
Adjustment, and with what
A span adjustment with an external 15 kg weight. The record should say whether the scale was adjusted with its internal weight or an external one, and which (cg-18, 8.2). After the adjustment all three tests are repeated as a separate as-left set (VIM 3.11, Note 3).
Decision rule, uncertainty and sign-off
Simple acceptance: an error passes if it is within the tolerance, and U is reported, so every as-left error passes. The as-left U of 6.1 g comes mostly from the 5 g display step and the repeatability; the rounding at zero and at the load alone give about 4.1 g (cg-18, 7.1.1). A stricter rule, |E| + U ≤ Tol (cg-18, 7.5.3, which uses the uncertainty of a weighing in normal use, larger still), could not show this scale to conform at two loads even with the calibration U: at 0.1 kg, where U alone exceeds the 5 g tolerance, and at 7.5 kg, where 5 + 6.1 > 10 g. The same results pass under one rule and not under the other: that is why the rule is agreed and goes on the record (Global ACI-TECH-1-002). The as-found failure opens an out-of-tolerance report.
The uncertainty column uses EURAMET’s formulas: the display step at zero and at the load, the repeatability and the eccentricity make up the uncertainty of the indication, and the weights add theirs (cg-18, §7.1). On a 0.1 mg analytical balance the reference mass, above all air buoyancy, can dominate instead (cg-18, App. H1); our procedure page walks through EURAMET’s own example.
What a scale calibration certificate should include
ISO/IEC 17025 sets the minimum content of any calibration certificate from an accredited lab. Clause 7.8.2.1 covers the general items (a unique identification, the customer, the method, the item calibrated, the dates, the results and who authorized the report), and clause 7.8.4.1 adds the calibration items: measurement uncertainty, the conditions that affect the results, a traceability statement, results before and after any adjustment or repair when available, and a statement of conformity when one is given (NIST’s crosswalk of clause 7.8). EURAMET cg-18 lists what is specific to weighing instruments (cg-18, §8):
| On a lab’s scale certificate | Where it goes on the template |
|---|---|
| Scale identification: Max, d, e and class, location | Scale block |
| Date, place, conditions, method, any limited range agreed | Method, weights and conditions |
| Whether the scale was adjusted, internally or externally, and with which weight | Adjusted |
| Test weights and how traceability is established | Test weights, weights cert. no. |
| Indications or errors at each load, before and after adjustment | Errors of indication table |
| Standard deviation, and whether it refers to a single reading | Repeatability table |
| Eccentricity results | Eccentricity table |
| Expanded uncertainty of each error, coverage factor and probability | U column (note k on the record) |
| Pass or fail, the tolerance and the decision rule | Tolerance, decision rule, decision |
Red flags on a scale certificate
- One weight and “CAL OK”. That is an adjustment. A calibration has at least five loads, repeatability and eccentricity, unless a smaller range was agreed and stated (cg-18, §5).
- Only as-left results. If the scale was adjusted and you only see the results after adjustment, you have lost the as-found data you need to judge past weighings (ISO/IEC 17025, clause 7.8.4.1). Ask for both.
- No word on the adjustment. The certificate should say whether the scale was adjusted with its internal weight or an external one (cg-18, §8.2).
- Weights with no class or certificate. You can’t check the one-third rule or the traceability (cg-18, §4.3.1; Handbook 44, App. A 3.2).
- Calibrated in the workshop, used on the shop floor. A scale moved after calibration needs at least a recheck (cg-18, §4.1.2).
- “Certified” read as legal for trade. A calibration certificate is not the official verification a commercial scale needs under Handbook 44, nor its NTEP certificate (NIST OWM, Weighing FAQs).
- “Pass” with no tolerance or decision rule. A statement of conformity must say which specification was used and how uncertainty was taken into account (clause 7.8.6, as described in Global ACI-TECH-1-002).
Our calibration certificate guide covers accreditation symbols, scopes and the full review checklist.
How to fill in the template
| Field | What to write |
|---|---|
| Max, d and e | From the data plate or the specification. Handbook 44 and OIML R 76 tolerances are counted in e |
| Tolerance by load | The steps for your class and source, e.g. ±1 e, ±2 e and ±3 e with their load limits. Use your process requirement if it is tighter |
| Test weights | ID, class and certificate. Their error (plus uncertainty, if used at nominal value) below ⅓ of the tolerance at each load; check with the TUR calculator. Let them reach room temperature first |
| Warm-up and exercise | The manufacturer’s warm-up time, then load to about Max at least once (cg-18, §4.1.3). Don’t record those readings |
| Site | Stable bench, no drafts or vibration, humidity 40–60 % if you can (NIST GMP 10, §1) |
| Errors of indication | At least five loads, e.g. the minimum, 25, 50, 75 and 100 % of Max, zero before each. Write the original indications; the errors come after (cg-18, §4.4.1) |
| Repeatability | One load, ideally one body, between half of Max and Max, five times or more (cg-18, §5.1); pharmaceutical procedures often use ten, as USP <41> asked until January 2026. Write every reading |
| Eccentricity | About Max/3, center first, then the four corners in any order (cg-18, §5.3) |
| U | From the lab certificate, or your own budget (try the uncertainty calculator). Write k if it isn’t 2 |
| Next due | Your interval. No scale standard sets one outside legal metrology; base it on risk and history |
Where to keep it and for how long
File each sheet under the scale’s asset ID, with the lab certificate and the weights’ certificates attached or referenced, and add a line to the scale’s equipment record: date, result, decision and next due date. That history is what lets you lengthen or shorten the interval later (see calibration intervals). How long to keep it depends on the standards and regulations you work under; our calibration records guide sets out the rules by standard. Print a matching status sticker with the calibration label generator.
Get the PDF pack
The template on this page is free to print, as it is. If you’d rather have it as a PDF, subscribe to Traceable, our newsletter, with the form on this page: subscribers get the printable pack with our calibration program starter kit and the instrument record templates, this one included.
Related:
- How to calibrate a scale: the method this record follows, with EURAMET’s worked example and USP <41>.
- Scale and balance calibration hub.
- How to read a calibration certificate.
- Out-of-tolerance results: what to do when the as-found result fails.
- TUR calculator and uncertainty calculator.
FAQ
What should a scale calibration certificate include?
The scale identification (ID, model, serial number, capacity, display step d, and e and accuracy class if marked), the place and date, the method, the environmental conditions, whether and how the scale was adjusted (internal or external weight), the test weights and their traceability, the results of the three tests (errors at several loads, repeatability and eccentricity), results before and after any adjustment, the expanded uncertainty of each error with its coverage factor, and, if it says pass or fail, the tolerance and the decision rule. It also needs a unique number and the name of the person who authorized it.
Is a scale calibration certificate the same as legal-for-trade certification?
No. A calibration certificate reports the errors of the scale and their uncertainty. Legal verification is the official check that a scale used to buy, sell or charge by weight meets the legal tolerances. In the US the rules are in NIST Handbook 44, and a commercial scale also needs an active NTEP Certificate of Conformance and approval by the state where it is used. A calibration certificate does not make a scale legal for trade.
Who can calibrate a scale?
A calibration lab or scale service company, or your own staff with test weights that have a valid calibration certificate and a written procedure. For audits, customers and regulated work, a certificate from a lab accredited to ISO/IEC 17025 for that calibration is the easiest evidence to defend. Pressing the CAL button is an adjustment, not a calibration.
How can I check if my scale is calibrated?
Look for a calibration label and a certificate or record for that scale, with results at several loads and a date that is still within your interval. To check it yourself, weigh a certified weight that you did not use to adjust it. Repeatability and off-center checks don't even need a calibrated weight: any stable object weighed several times, and at the center and corners of the pan, should give the same reading.
Which calibration weights do I need, and do they need a certificate?
Yes, for a calibration the weights need a valid calibration certificate traceable to the SI. Choose the class from the scale class: NIST Handbook 105-1 suggests ASTM class 1 or OIML F1 for Class I balances and ASTM 3 to 6 or OIML M1 to M2-3 for Class III bench and platform scales. Then check the numbers: the weight's error (plus its uncertainty, if used at nominal value) should be below one third of the scale tolerance at that load.
What is different about an analytical balance certificate?
Mostly the size of the uncertainty and what goes into it. On a balance with a 0.1 mg display step, air buoyancy and the weights' certified values matter, so the certificate should say whether weights were used at nominal or certified values and how buoyancy was handled. Balances used in pharmaceutical work also fall under USP <41>, which links calibration with routine checks and the minimum weight; its revision became official on February 1, 2026.
Is the certificate still valid if I move the scale?
Not automatically. A different place can change the readings through local gravity, the environment or the transport itself, and EURAMET's guide says a certificate shouldn't be accepted as evidence of traceability for a scale moved after calibration unless it has been shown to be immune to those effects. Calibrate scales where they are used, and at least recheck them after a move.
Sources
- EURAMET. EURAMET Calibration Guide No. 18 (cg-18), Guidelines on the Calibration of Non-Automatic Weighing Instruments, version 4.0. 2015 — §4.1–4.4, §5, §6, §7.1, §7.5.3, §8, App. H1. Cited, not reproduced
- NIST Office of Weights and Measures. NIST Handbook 44, Specifications, Tolerances, and Other Technical Requirements for Weighing and Measuring Devices, 2026 edition. 2026 — G-T.1; 2.20 Scales: T.N.3.2, Table 6; App. A 3.2
- OIML. OIML R 76-1:2006, Non-automatic weighing instruments, Part 1: Metrological and technical requirements. 2006 — 3.5.1–3.5.2 (Table 6), 3.6.2, 3.7.1
- NIST Office of Weights and Measures. NIST Handbook 105-1, Specifications and Tolerances for Field Standard Weights. 2019 — Table 1 (weight classes for each scale class)
- NIST Office of Weights and Measures. NIST GMP 10, Good Measurement Practice for Understanding Factors Affecting Weighing Operations (in NISTIR 6969-2019). 2019 — §1 environment
- NIST Office of Weights and Measures. Weighing FAQs — Handbook 44 adoption, NTEP Certificate of Conformance, state approval; checked 2026-10-16
- United States Pharmacopeia. USP General Chapter <41> Balances, proposed revision (Pharmacopeial Forum, 2024): preview. 2024
- ECA Academy (gmp-compliance.org). USP: Chapters Balances <41> and Weighing on an Analytical Balance <1251> final. 2025 — Both chapters official February 2026
- United States Pharmacopeia. USP General Chapter <41> Balances (USP 38–NF 33 text, official 2013 to January 2026) — Superseded February 1, 2026; ten repeatability readings
- NIST Office of Weights and Measures. ISO/IEC 17025 Crosswalk – Reporting the Results (clause 7.8). 2018 — Free summary of ISO/IEC 17025:2017 clause 7.8 (7.8.2.1, 7.8.4.1, 7.8.6)
- Global Accreditation Cooperation (Global ACI). Global ACI-TECH-1-002 (G), Guidelines on Decision Rules and Statements of Conformity (formerly ILAC G8:09/2019). 2026
- JCGM / BIPM. International Vocabulary of Metrology (VIM), JCGM 200:2012. 2012 — 2.39 calibration, 3.11 adjustment