Pipette calibration worksheet and certificate template (free, printable)

A free, printable pipette calibration worksheet for the ISO 8655-6 gravimetric method: three volumes, ten weighings each, Z factor, mean volume, systematic error and CV, as found and as left, with a filled-in example and what a pipette calibration certificate should include.

By David Bosch11 primary sourcesHow we researchCite

On this page
  1. Pipette calibration worksheet
  2. Example: a filled-in pipette calibration worksheet
  3. What a pipette calibration certificate should include
  4. How to fill in the worksheet
  5. Where to keep it and for how long
  6. Get the PDF pack
  7. FAQ
  8. Sources

A pipette calibration worksheet (and the certificate a lab gives you) has to show seven things:

  1. Which pipette and which tip: ID, model and serial number, volume range, single or multichannel, and the tip type, because the result is valid only for that combination.
  2. The limits and their source: ISO 8655-2, the manufacturer’s or your own, for both the systematic error and the CV.
  3. The method and the conditions: ISO 8655-6 (or a routine test under Part 7), the balance, water and air temperature, pressure and humidity.
  4. The readings: at three volumes (nominal, 50 % and 10 % or the lowest usable), ten balance readings each.
  5. The calculation: each reading times the Z factor, the mean volume, the systematic error and the CV, as found and, after any adjustment or service, as left.
  6. The uncertainty of each mean volume and the decision rule.
  7. The decision: within limits or not, what was done, who did it and reviewed it, and the next due date.

The free worksheet below has a field for each of these and fits on one Letter page. Print it from this page: no sign-up needed. Below it, the same worksheet filled in with a worked example, then what to look for on a lab certificate.

Pipette calibration worksheet

Pipette calibration worksheet

One sheet per pipette and channel. ISO 8655-6:2022 gravimetric method: three volumes, ten deliveries each.

Pipette
Asset ID
Manufacturer / model / serial no.
TypeFixedVariableSingle-channelMultichannel, channel:
Volume range / nominal, µL
Tip type (maker, part no.)
Test volumes Vs, µL, and limits (systematic ± % / CV %)
Limits fromISO 8655-2:2022ManufacturerOwn (equal or stricter)
Method, equipment and conditions
MethodISO 8655-6:2022Routine test (ISO 8655-7)Procedure no.:
Calibrated byIn-house (name):External lab (name, accreditation no.):
Date / record no.
Balance ID / resolution (ISO 8655-6 Table 1)
Water temp. start / end, °C
Air temp., °C / pressure, hPa / RH, %
Z factor, µL/mg / evaporation corr., mg
Equilibrated ≥ 2 h / pre-wet 5× / tips changed
Balance readings, mg (10 per volume; do not change the setting during a series)
Delivery10 % (or lowest): as found10 %: as left50 %: as found50 %: as leftNominal: as foundNominal: as left
1
2
3
4
5
6
7
8
9
10
Results
10 % (or lowest): as found10 %: as left50 %: as found50 %: as leftNominal: as foundNominal: as left
Mean, mg
V̄, µL
Syst. error, %
CV, %
Pass / fail

V̄ = (mean reading + evaporation corr.) × Z. Syst. error = (V̄ − Vs) ÷ Vs × 100. CV = 100 × s ÷ V̄ (s of the volumes, n − 1). Both must be within limits.

Adjusted or servicedNoAdjusted (tool), markedServiced (what):
Expanded uncertainty of each mean volume, µL (k = 2)
Decision ruleSimple acceptanceGuard band:Other:
Decision and sign-off
As foundIn limitsOut of limits: OOT report no.:
StatusReturned to serviceLimited use (labeled)Removed from service
Next due
Status label appliedYesNo
Performed by / date
Reviewed by / date

Use one sheet per pipette (one per channel for multichannel pipettes). Write the balance readings as they come, then fill in the results block: add any evaporation correction, multiply by Z, and calculate the mean, the systematic error and the CV, or let our pipette calculator do it. For a pipette calibrated by an outside service, copy the key results from the certificate, record your own decision and attach the certificate. The full method is on our pipette calibration procedure page.

Example: a filled-in pipette calibration worksheet

The worksheet below is for the same fictitious pipette as the worked example on our procedure page: same balance readings and conditions, and Z, every mean volume, systematic error, CV and pass/fail is calculated from them. The company, people, limits and uncertainties are invented. Click a number to jump to its note.

Pipette calibration worksheetExample

One sheet per pipette and channel. ISO 8655-6:2022 gravimetric method: three volumes, ten deliveries each.

1Pipette
Asset IDPIP-014
Manufacturer / model / serial no.Example Lab Co. / VP-1000 / EX-61522
TypeFixedVariable (ticked)Single-channel (ticked)Multichannel, channel:
Volume range / nominal, µL100 to 1000 / 1000
Tip type (maker, part no.)Example Lab Co. T-1000
Test volumes Vs, µL, and limits (systematic ± % / CV %)100: ±1.5 / 0.6; 500: ±0.8 / 0.3; 1000: ±0.6 / 0.2
2Limits fromISO 8655-2:2022Manufacturer (ticked)Own (equal or stricter)
3Method, equipment and conditions
MethodISO 8655-6:2022 (ticked)Routine test (ISO 8655-7)Procedure no.:
Calibrated byIn-house (name): (ticked)External lab (name, accreditation no.):P. Novak
Date / record no.2026-10-12 / PC-2026-0377
Balance ID / resolution (ISO 8655-6 Table 1)BAL-04 / 0.1 mg
Water temp. start / end, °C21.4 / 21.4
Air temp., °C / pressure, hPa / RH, %21.6 / 1009 / 51
4Z factor, µL/mg / evaporation corr., mg1.00314 / 0
Equilibrated ≥ 2 h / pre-wet 5× / tips changed3 h / yes / after delivery 5
5Balance readings, mg (10 per volume; do not change the setting during a series)
Delivery10 % (or lowest): as found10 %: as left50 %: as found50 %: as leftNominal: as foundNominal: as left
198.1100.6495.6498.9987.9997.6
299.199.6493.0498.9988.1998.3
397.2100.1493.9500.2986.2996.8
498.8100.4493.7498.1985.0998.1
597.7100.9494.3499.6984.5999.1
697.7100.4492.3498.2986.1998.2
799.699.7493.2500.5984.61000.1
898.1100.5492.9499.9984.0996.2
997.999.9492.6500.7986.3997.7
1098.6100.8492.8499.4986.3996.4
6Results
10 % (or lowest): as found10 %: as left50 %: as found50 %: as leftNominal: as foundNominal: as left
Mean, mg98.28100.29493.43499.44985.90997.85
V̄, µL98.59100.61494.98501.01989.001000.99
Syst. error, %−1.41+0.61−1.00 (outside tolerance)+0.20−1.10 (outside tolerance)+0.10
CV, %0.75 (outside tolerance)0.450.200.180.140.12
Pass / failFail (outside tolerance)PassFail (outside tolerance)PassFail (outside tolerance)Pass

V̄ = (mean reading + evaporation corr.) × Z. Syst. error = (V̄ − Vs) ÷ Vs × 100. CV = 100 × s ÷ V̄ (s of the volumes, n − 1). Both must be within limits.

7Adjusted or servicedNoAdjusted (tool), marked (ticked)Serviced (what): (ticked)seals cleaned and replaced
Expanded uncertainty of each mean volume, µL (k = 2)0.8 at 100, 1.1 at 500, 1.5 at 1000
Decision ruleSimple acceptance (ticked)Guard band:Other:
8Decision and sign-off
As foundIn limitsOut of limits: OOT report no.: (ticked)OOT-2026-19
StatusReturned to service (ticked)Limited use (labeled)Removed from service
Next due2027-04-12; routine check monthly
Status label appliedYes (ticked)No
Performed by / dateP. Novak, 2026-10-12
Reviewed by / dateH. Weber, 2026-10-13
  1. The pipette and its tip

    Under ISO 8655:2022 the pipette and its tip are one system, and the result is valid only for the tip type used (ISO 8655-2:2022, foreword), so the tip goes on the record. PIP-014 is a variable, single-channel air-displacement pipette, 100 to 1,000 µL. A multichannel pipette gets one sheet per channel.

  2. Which limits

    The example uses the manufacturer's specification (fictitious) as its limits: ±0.6 % and CV 0.2 % at 1,000 µL. To state compliance with ISO 8655, your limits must be equal to or stricter than ISO 8655-2's; at nominal volume for this size, those are 0.8 % and 0.3 % (Eppendorf, 2023).

  3. Balance, room and conditions

    A 0.1 mg balance is enough for a 1,000 µL pipette: ISO 8655-6 Table 1 is keyed to the nominal volume, not to the volume tested (5.2). Water temperature at the start and end, plus air temperature, pressure and humidity (8.1.4). Everything equilibrated in the room for at least 2 hours (7.2), and the tip pre-wet five times (8.3.2).

  4. Z turns milligrams into microliters

    Z = 1.00314 µL/mg for water at 21.4 °C, air at 21.6 °C, 1009 hPa and 51 % RH (EURAMET cg-19, equations 1 to 4). Skipping it would make every volume about 0.3 % low. Evaporation matters most below 50 µL (ISO 8655-6, 7.3); here the lab determined it as negligible against the 0.1 mg resolution at these volumes, so the correction is zero (8.1.3).

  5. Ten readings per volume

    At least ten deliveries per volume (ISO 8655-6, 8.1.2), smallest volume first as DKD-R 8-1 suggests, without changing the setting during a series, and the tip changed at least once per volume (Sartorius). Write every balance reading as read: the scatter is part of the result.

  6. Results: two ways to fail

    As found, the mean was 1.1 % low at 1,000 µL and 1.0 % low at 500 µL, outside the systematic limits. At 100 µL the mean was inside (−1.41 % against ±1.5 %), but the CV of 0.75 % was not: the deliveries scattered too much. As left, all three volumes pass both limits, with systematic errors from +0.10 % to +0.61 %.

  7. Service, adjustment and marking

    A large CV is not adjusted out: an imprecise pipette usually needs repair or replacement (Troemner). Here the seals were cleaned and replaced, then the volume was adjusted with the manufacturer's tool and the pipette marked so the change is visible (ISO 8655-2, 5.3). The whole test was repeated as the as-left set (VIM 3.11, Note 3).

  8. Uncertainty, decision and sign-off

    A calibration states the uncertainty of each mean volume (Sartorius). The values here are invented, scaled from DKD's sample budget for a variable-volume pipette (0.15 % of nominal volume), with 75 % and 50 % of that at the middle and lowest volumes, as DKD-R 8-1 does (Appendix B and the note before the appendices). Simple acceptance; if results sit close to a limit, agree a decision rule (Global ACI-TECH-1-002). The as-found failure opens an out-of-tolerance report: everything pipetted since the last good calibration was about 1 % short at the upper volumes.

The systematic error says whether the pipette delivers the right amount on average (trueness), and the CV how much single deliveries scatter (precision). They fail for different reasons: a systematic error can usually be adjusted out, while a large CV points to seals, leaks or the tip fit (DKD-R 8-1, §5.3; Troemner).

What a pipette calibration certificate should include

ISO/IEC 17025 sets the minimum content of any calibration certificate from an accredited lab. Clause 7.8.2.1 covers the general items (a unique identification, the customer, the method, the item calibrated, the dates, the results and who authorized the report), and clause 7.8.4.1 adds the calibration items: measurement uncertainty, the conditions that affect the results, a traceability statement, results before and after any adjustment or repair when available, and a statement of conformity when one is given (NIST’s crosswalk of clause 7.8). ISO 8655-6 lists what a pipette report must contain (clause 10, as summarized by Sartorius); as found and as left come from ISO/IEC 17025 7.8.4.1, and Sartorius notes that the 2022 edition calls for an as-found test before maintenance or repair:

On a pipette calibration certificate Where it goes on the worksheet
Pipette identification, tips or other exchangeable parts, delivery (Ex) or content (In) Pipette block, tip type
Test conditions: temperatures, pressure, humidity Water temp., air temp. / pressure / RH
Volume of each delivery and number of measurements per volume Balance readings table (times Z)
Systematic and random errors (as found and as left, if adjusted or repaired) Results table
Limits used, if any Test volumes and limits, limits from
Expanded uncertainty of each mean volume, per volume and channel, when required Expanded uncertainty field
Method: ISO 8655-6:2022, any deviation, and the mass-to-volume formula Method, Z factor
Date and operator Date, performed by

Red flags on a pipette certificate

  • Fewer volumes or readings, called a calibration. Three volumes × ten measurements per channel is the Part 6 calibration of a variable pipette (ISO 8655-6, 8.1). Fewer is a routine test under Part 7: fine if that is what you ordered and the certificate says so.
  • No as-found results. Sartorius notes that the 2022 edition calls for an as-found calibration or test before maintenance or repair. Without it you can’t judge past results.
  • No tip type. The result is valid only for the tip used (ISO 8655-2:2022, foreword).
  • No Z factor or conditions. Without them, you can’t tell whether readings in mg were converted to µL properly.
  • No uncertainty, or no limits. A calibration states the uncertainty of each mean volume, and a pass needs the limits it was judged against (Sartorius; ISO/IEC 17025, clause 7.8.6, as described in Global ACI-TECH-1-002).
  • “Complies with ISO 8655” against looser limits. Own or manufacturer limits must be equal to or stricter than ISO 8655-2’s to support that claim (Eppendorf, 2023; SelectScience).
  • Numbers that look too clean. Ten readings should show some scatter, and the time spent should fit 30 weighings per single-channel pipette.

Our calibration certificate guide covers accreditation symbols, scopes and the full review checklist.

How to fill in the worksheet

Field What to write
Volume range / nominal The range on the pipette and the nominal (largest) volume. It decides the balance and the test volumes
Tip type Maker and part number of the tips used. Approved tips, at least two per volume
Balance ID and resolution from ISO 8655-6 Table 1 for the nominal volume: 0.001 mg below 20 µL, 0.01 mg from 20 to below 200 µL, 0.1 mg from 200 µL to 10 mL
Conditions Water temperature at the start and end; air temperature, pressure and humidity. The room at (20 ± 3) °C (ISO 8655-6, 7.2); water within 0.5 °C of the air (ISO 8655-6, clause 6)
Z factor From the water temperature, air temperature, pressure and humidity (EURAMET cg-19), or from a published Z table
Evaporation Determine it for small volumes, especially below 50 µL, and add it to each reading (ISO 8655-6, 7.3 and 8.1.3)
Balance readings One column per volume and set. Don’t change the volume setting during a series (ISO 8655-6, 8.2)
Results Mean reading × Z = mean volume; systematic error and CV against the limits. Both must pass
Uncertainty From your budget or the lab certificate. The balance is rarely the main contribution: handling and temperature differences usually are (DKD-R 8-1, Appendix A). Try the uncertainty calculator

Where to keep it and for how long

File each worksheet under the pipette’s asset ID, with the service certificate attached, and add a line to the pipette’s equipment record: date, result, decision and next due date. Keep routine checks with it: they run at shorter intervals than calibrations (Sartorius) and show drift early, which is what lets you lengthen or shorten the interval later (see calibration intervals). How long to keep it depends on the standards and regulations you work under; our calibration records guide sets out the rules by standard. Print a matching status sticker with the calibration label generator.

Get the PDF pack

The worksheet on this page is free to print, as it is. If you’d rather have it as a PDF, subscribe to Traceable, our newsletter, with the form on this page: subscribers get the printable pack with our calibration program starter kit and the instrument record templates, this one included.

Related:

FAQ

What should a pipette calibration certificate include?

The pipette identification and the tip type used, the test conditions (water and air temperature, pressure, humidity), the volume of each delivery and the number of measurements per volume, the systematic and random errors at each volume, as found and as left if the pipette was adjusted or repaired, the limits used, the expanded uncertainty of each mean volume, the method (ISO 8655-6:2022 and any deviation from it) with the formula used to convert weights into volumes, the date and the operator.

How many weighings does a pipette calibration need?

Under ISO 8655-6:2022, a variable-volume pipette is tested at three volumes (nominal, 50 %, and 10 % or the lower limit of the usable range, whichever is greater), with at least ten measurements at each: 30 weighings per channel. A fixed-volume pipette is tested at its nominal volume. A routine test under Part 7 may use fewer volumes and at least four measurements per volume, but it is not a calibration to Part 6.

Why multiply the balance reading by a Z factor?

Because 1 µL of water does not weigh exactly 1 mg. Z corrects for the density of water at the test temperature and for air buoyancy; near room temperature it is about 1.003 µL/mg. Treating mg as µL builds an error of about 0.3 % into every result, more than a third of the 0.8 % permissible systematic error of a 1,000 µL pipette.

What are the ISO 8655 limits for a pipette?

ISO 8655-2:2022 sets maximum permissible systematic and random errors for the pipette and tip together, at 100 %, 50 % and 10 % of nominal volume. At the nominal volume of a single-channel air-displacement pipette between 100 and 5,000 µL, they are 0.8 % systematic and 0.3 % random error (CV). For the 50 % and 10 % limits, check Table 1 of the standard. Your own or the manufacturer's limits must be equal or stricter if you want to state ISO 8655 compliance.

I need a pipette calibration order form. Is this it?

No: an order form is what a calibration service asks you to send with your pipettes, and each service has its own. This page is the worksheet for doing or checking the calibration. When you order, ask for three volumes and ten measurements per channel to ISO 8655-6, as-found results before any service, the tip type used, the limits applied and the uncertainty, so the certificate you get back has what this worksheet records.

How do I fill it in for a multichannel pipette?

Use one sheet per channel. Each channel is tested on its own: on a single-channel balance, all channels aspirate together and only the channel being measured dispenses into the weighing vessel. A 12-channel pipette at three volumes means 360 weighings.

Can I do the calculations in a spreadsheet?

Yes, as long as the formulas are checked. Our free pipette calculator, on the procedure page, does the same math as this worksheet: it converts readings with Z and gives the mean volume, systematic error, standard deviation and CV, and it is tested against published values. It does not calculate the measurement uncertainty.

Sources

  1. ISO. ISO 8655-6:2022, Piston-operated volumetric apparatus, Part 6: Gravimetric reference measurement procedure for the determination of volume (official preview). 2022 — Paid standard. Read on the free official preview (clauses 4, 5.2, 7.2, 7.3, 8.1–8.3, Table 1); clauses cited, not reproduced
  2. ISO. ISO 8655-2:2022, Piston-operated volumetric apparatus, Part 2: Pipettes (official preview). 2022 — Foreword (pipette and tip as one system), 5.3 (marking an adjustment)
  3. EURAMET. EURAMET Calibration Guide No. 19 (cg-19), Guidelines on the Determination of Uncertainty in Gravimetric Volume Calibration, version 4.1. 2025 — §3 equations (1) to (4). Cited, not reproduced
  4. PTB / Deutscher Kalibrierdienst. DKD-R 8-1, Calibration of piston-operated pipettes with air cushions, edition 12/2011, revision 1 — §5.3, §7, §8.9, Appendices A and B
  5. Sartorius. Pipette Calibration and Compliance: A Practical Guide to ISO 8655 (includes the white paper Gravimetric Calibration and Testing of POVA according to Part 6 and 7 of the ISO 8655:2022). 2023 — Test types, tip change, multichannel, report contents (ISO 8655-6 clause 10)
  6. SelectScience. Pipette calibration ISO 8655:2022: everything you need to know. 2022
  7. Eppendorf. Change of measurement values according to ISO 8655:2022. 2023 — Maximum permissible errors at nominal volume; own limits equal or stricter
  8. Troemner. Pipette Standards Handbook — Imprecise pipettes need repair or replacement
  9. NIST Office of Weights and Measures. ISO/IEC 17025 Crosswalk – Reporting the Results (clause 7.8). 2018 — Free summary of ISO/IEC 17025:2017 clause 7.8 (7.8.2.1, 7.8.4.1, 7.8.6)
  10. Global Accreditation Cooperation (Global ACI). Global ACI-TECH-1-002 (G), Guidelines on Decision Rules and Statements of Conformity (formerly ILAC G8:09/2019). 2026
  11. JCGM / BIPM. International Vocabulary of Metrology (VIM), JCGM 200:2012. 2012 — 3.11 adjustment