Torque wrench calibration certificate and record template (free, printable)

A free, printable torque wrench calibration record that follows ISO 6789-2: test points, five readings per point, as found and as left, tolerance, uncertainty and decision, with a filled-in example and what a lab certificate should include.

By David Bosch16 primary sourcesHow we researchCite

On this page
  1. Torque wrench calibration record template
  2. Example: a filled-in torque wrench calibration record
  3. What a torque wrench calibration certificate should include
  4. How to fill in the template
  5. Where to keep it and for how long
  6. Get the PDF pack
  7. FAQ
  8. Sources

A torque wrench calibration record (or the certificate a lab gives you) has to show seven things:

  1. Which wrench: asset ID, serial number, type and class under ISO 6789-1, range and the direction tested.
  2. The tolerance and its source: typically ±4 % for click and dial wrenches with a maximum above 10 N·m, and ±6 % for those with a maximum of 10 N·m or less (ISO 6789-1, clause 5.1.5). The limit follows the wrench’s capacity, not the setting.
  3. The method and the reference: ISO 6789-2:2017 or your procedure, the calibration system with its certificate, and any adapters.
  4. The conditions: temperature and humidity.
  5. The readings: at the lowest value of the range (usually 20 % of maximum), 60 % and 100 %, five at each point, as found and, after any adjustment, as left.
  6. The uncertainty at each point and the decision rule for pass or fail.
  7. The decision: in or out of tolerance, what was done, who did it and reviewed it, and the next due date.

The free template below has a field for each of these. Print it from this page: no sign-up needed. Below it, the same form filled in with a worked example, then what to look for on a lab certificate.

Torque wrench calibration record template

Torque wrench calibration record

One sheet per wrench and direction. Test points and readings follow ISO 6789-2:2017; tolerance from ISO 6789-1 or the manufacturer.

Wrench
Asset ID
Manufacturer / model
Serial no.
Location / owner
Type (ISO 6789-1)Type I, indicatingType II, setting (click)
Class (A to G)
Range (lowest marked value to max.)
Graduation / resolution
UnitsN·mlbf·ftlbf·in
Direction on this sheetClockwiseCounterclockwise
Tolerance ± ____ %, fromISO 6789-1 (type, class, max.)ManufacturerASME B107.300Other:
Method, equipment and conditions
MethodISO 6789-2:2017In-house procedure no. / rev.:
Calibrated byIn-house (name):External lab (name, accreditation no.):
Calibration date
Certificate or record no.
Calibration system ID
System cert. no. / due date
System uncertainty interval, % (max. ¼ of tol.)
Adapters / accessories used
Temperature, °C (18 to 28, stable ±1)
Humidity, % RH (below 90)
Exercised at maximum torque before the readings (how many times)
As found (before any adjustment)
Test pointSet or ref. valueR1R2R3R4R5Mean dev. %Largest dev. %U, % (k = 2)Pass / fail
Lowest value (usually 20 %)
60 %
100 %
As left (after adjustment; if not adjusted, as found = as left)
Test pointSet or ref. valueR1R2R3R4R5Mean dev. %Largest dev. %U, % (k = 2)Pass / fail
Lowest value (usually 20 %)
60 %
100 %

R1 to R5: for a setting (click) wrench, the reference reading at release; for an indicating wrench, what the wrench shows at the applied reference torque. Deviation % = (R − set or reference value) ÷ set or reference value × 100 (ISO 6789-1 form). Every reading must be within the tolerance.

AdjustedNoYes (what):
Decision ruleSimple acceptanceGuard band:Other:
Decision and sign-off
As foundIn toleranceOut of tolerance: OOT report no.:
StatusReturned to serviceLimited use (labeled)Removed from service
Next due (date or cycles)
Status label appliedYesNo
Performed by / date
Reviewed by / date

Use one sheet per wrench and direction. For a wrench calibrated by an outside lab, copy the key results from the lab’s certificate, record your own decision on the last block, and attach the certificate to the record. For in-house calibration, the sheet is the record. The test points and five readings follow the ISO 6789-2 pattern for indicating tools and adjustable, graduated setting tools (Type I, and Type II Classes A, D and G); the full method is on our torque wrench calibration procedure page.

Example: a filled-in torque wrench calibration record

The record below is for the same fictitious click wrench as the worked example on our procedure page: same readings, and every deviation is calculated from them. The company, people, numbers and uncertainties are invented. Click a number to jump to its note.

Torque wrench calibration recordExample

One sheet per wrench and direction. Test points and readings follow ISO 6789-2:2017; tolerance from ISO 6789-1 or the manufacturer.

1Wrench
Asset IDTW-031
Manufacturer / modelExample Tools Co. / CW-100
Serial no.EX-58213
Location / ownerAssembly line 2 / M. Ortiz
Type (ISO 6789-1)Type I, indicatingType II, setting (click) (ticked)
Class (A to G)A (adjustable, graduated)
Range (lowest marked value to max.)20 to 100 N·m
Graduation / resolution0.5 N·m
UnitsN·m (ticked)lbf·ftlbf·in
Direction on this sheetClockwise (ticked)Counterclockwise
2Tolerance ± ____ %, fromISO 6789-1 (type, class, max.) (ticked)ManufacturerASME B107.300Other:±4 %
3Method, equipment and conditions
MethodISO 6789-2:2017 (ticked)In-house procedure no. / rev.:
Calibrated byIn-house (name): (ticked)External lab (name, accreditation no.):J. Lee
Calibration date2026-09-14
Certificate or record no.CR-2026-0412
Calibration system IDTCS-01 (transducer + loader)
System cert. no. / due dateEX-24-0917 / 2027-03-02
System uncertainty interval, % (max. ¼ of tol.)0.5 % (limit 1 %)
Adapters / accessories usedAD-03, 3/8-in, fixed position
Temperature, °C (18 to 28, stable ±1)21.4 (±0.3)
Humidity, % RH (below 90)46
Exercised at maximum torque before the readings (how many times)3 times at 100 N·m, clockwise (not recorded)
4As found (before any adjustment)
Test pointSet or ref. valueR1R2R3R4R5Mean dev. %Largest dev. %U, % (k = 2)Pass / fail
20 % (lowest)20.020.95 (outside tolerance)21.02 (outside tolerance)20.88 (outside tolerance)21.10 (outside tolerance)20.97 (outside tolerance)+4.92+5.500.8Fail (outside tolerance)
60 %60.061.661.961.461.861.7+2.80+3.170.6Pass
100 %100.0101.9102.4101.6102.2102.0+2.02+2.400.5Pass
5As left (after adjustment; if not adjusted, as found = as left)
Test pointSet or ref. valueR1R2R3R4R5Mean dev. %Largest dev. %U, % (k = 2)Pass / fail
20 % (lowest)20.020.4220.5120.3720.5520.45+2.30+2.750.8Pass
60 %60.060.961.260.761.061.1+1.63+2.000.6Pass
100 %100.0100.8101.3100.6101.1100.9+0.94+1.300.5Pass

R1 to R5: for a setting (click) wrench, the reference reading at release; for an indicating wrench, what the wrench shows at the applied reference torque. Deviation % = (R − set or reference value) ÷ set or reference value × 100 (ISO 6789-1 form). Every reading must be within the tolerance.

AdjustedNoYes (what): (ticked)release spring, per maker’s manual
6Decision ruleSimple acceptance (ticked)Guard band:Other:every reading within ±4 %; U reported
7Decision and sign-off
As foundIn toleranceOut of tolerance: OOT report no.: (ticked)OOT-2026-07
StatusReturned to service (ticked)Limited use (labeled)Removed from service
Next due (date or cycles)2027-09-14 or 5,000 cycles
Status label appliedYes (ticked)No
Performed by / dateJ. Lee, 2026-09-14
Reviewed by / dateR. Patel, 2026-09-15
  1. Which wrench, and what kind

    The asset ID ties the record, the status label and any certificate together. The type and class (ISO 6789-1, clause 4) and the maximum torque decide the tolerance, so they belong on the record, not only in the catalog. TW-031 is a Type II, Class A click wrench, 20 to 100 N·m.

  2. The tolerance and where it comes from

    ±4 % because the maximum torque is above 10 N·m (ISO 6789-1, clause 5.1.5, Table 4). The limit follows the capacity of the wrench, not the setting, so it is ±4 % at 20 N·m too. If the manufacturer claims a tighter limit, that limit applies. Write the source next to the number.

  3. Method, system and conditions

    The calibration system's relative uncertainty interval should be no more than a quarter of the wrench tolerance (ISO 6789-2, 4.3, as described by Norbar and Torque Lab): 0.5 % against a 1 % limit here. Temperature 18 to 28 °C, stable within ±1 °C, humidity below 90 % RH (Norbar; Torque Lab). One identified adapter, always mounted the same way (DKD-L 10-1, 3.1 item 4). The three exercises at maximum are not recorded as readings (AWS; Sushma Industries).

  4. As found: the 20 % point fails

    Five readings at each point. At 20 N·m the limit is 20.8 N·m, and all five readings are above it: mean deviation +4.92 %, largest +5.50 %. Even after subtracting the 0.8 % uncertainty, the mean is outside. The 60 % and 100 % points pass, so a one-point check at mid-range would have passed this wrench. The as-found results are written down before anyone touches the adjustment.

  5. As left: a separate set of readings

    After the adjustment, the whole sequence is repeated and recorded as a new set (VIM 3.11 Note 3: an adjusted instrument usually has to be calibrated again). Every reading is now inside ±4 %, the largest +2.75 % at 20 N·m. If nothing is adjusted, tick "No" and write "as found = as left" instead of copying the numbers.

  6. Decision rule

    Simple acceptance: a reading passes if it is inside ±4 %, and the uncertainty is reported next to it. That matches ISO 6789-1, which judges conformance to its limits without taking uncertainty into account (clause 5.1.5). Results close to the limit can be judged differently under a guard band; whichever rule you use, name it (Global ACI-TECH-1-002, formerly ILAC G8). See our guard banding guide.

  7. Decision, next due and sign-off

    An as-found failure opens an out-of-tolerance report: what was tightened with this wrench near 20 N·m since its last good calibration? See out-of-tolerance results. Next due uses the default associated with ISO 6789, 12 months or 5,000 cycles, whichever comes first (Norbar in Fastener + Fixing; UT Metrology). A second person reviews the record before the wrench goes back into use.

The deviation on this form is the ISO 6789-1 form: (reading − set value) ÷ set value. ISO 6789-2 expresses the error of the wrench relative to the reference reading, so the same result gives a slightly different number. Norbar’s example: 104 N·m read at a 100 N·m setting is +4 % one way and −3.85 % the other (Norbar, Calibration Laboratories). Pick one, and say which on your procedure.

What a torque wrench calibration certificate should include

ISO/IEC 17025 sets the minimum content of any calibration certificate from an accredited lab. Clause 7.8.2.1 covers the general items (a unique identification, the customer, the method, the item calibrated, the dates, the results and who authorized the report), and clause 7.8.4.1 adds the calibration items: measurement uncertainty, the conditions that affect the results, a traceability statement, results before and after any adjustment or repair when available, and a statement of conformity when one is given (NIST’s crosswalk of clause 7.8). ISO 6789-2 adds what is specific to torque tools (Norbar, Calibration Laboratories; Torque Lab):

On a lab’s ISO 6789-2 certificate Where it goes on the template
Wrench identification, type, class, range Wrench block
Method: ISO 6789-2:2017, plus “in conjunction with DKD-L 10-1” if the lab uses it (DKD-L 10-1, §3.2) Method
The calibration system and its traceability Calibration system ID, system cert. no.
Adapters and interchangeable parts used Adapters / accessories used
Temperature and humidity Conditions
Results at each test point, before and after adjustment As found and as left tables
Relative expanded uncertainty at each point, and the relative uncertainty interval U column (the interval combines the mean error, the expanded uncertainty and the error of the calibration system)
Pass or fail, the specification used and the decision rule Tolerance, decision rule, decision

Norbar notes that it is not unusual for a tool claimed at 4 % to end up with a relative uncertainty interval of 6 %. That is not a defect of the certificate: it is the honest size of the measurement, and the reason to record the uncertainty next to each result.

Red flags on a torque wrench certificate

  • A “certificate” that came in the box. Under ISO 6789:2017 the manufacturer supplies a declaration of conformance, not a calibration certificate; the 2017 edition dropped factory certificates because they lacked the information needed to be traceable (ISO 6789-1, foreword, items g and h). Some manufacturers still ship one, Harbor Freight’s ICON page among them. It is useful, but it is not a traceable calibration with uncertainty.
  • Only as-left results. If the lab adjusted the wrench and reports only the results after adjustment, you have lost the as-found data you need to judge past work. ISO/IEC 17025 asks for results before and after adjustment when they are available (clause 7.8.4.1). Ask for as-found data when you order the calibration.
  • One test point, or one reading per point. ISO 6789-2 tests the lowest value, 60 % and 100 %, with five readings each for most wrenches (Torque Lab; AWS). A wrench can pass at one point and fail at another, as the example shows.
  • No uncertainty. A certificate to ISO 6789-2 states the uncertainty at each point (Norbar; Torque Lab).
  • “Pass” with no tolerance or decision rule. A statement of conformity must say which specification was used and how uncertainty was taken into account (ISO/IEC 17025, clause 7.8.6, as described in Global ACI-TECH-1-002).
  • The wrong tolerance. Check that the tolerance matches the type, class and maximum torque of your wrench, or the manufacturer’s tighter claim. Many US click wrenches are rated to ASME B107.300; if that is your basis, write it on the record.
  • “NIST traceable” and nothing else. NIST’s policy puts the burden of supporting a traceability claim on whoever provides the result, and says a NIST test report number should not be used as the sole proof of traceability (NIST traceability FAQ, 5.1.6 and 5.2.4). Look for the calibration system and its own certificate.

Our calibration certificate guide covers accreditation symbols, scopes and the full review checklist.

How to fill in the template

Field What to write
Type and class From the wrench’s declaration or specification sheet (ISO 6789-1, clause 4). Not sure? Ask the manufacturer before you set a tolerance
Range From the lowest marked value to the maximum. If an older wrench has no defined lower limit, DKD allows starting at 20 %, provided the range is marked and stated (DKD-L 10-1, §3.3 item 8)
Units The units on the wrench scale. Don’t mix lbf·in and lbf·ft: they differ by a factor of 12
Direction One sheet per direction. Calibrate counterclockwise too if the wrench is made to work both ways (ISO 6789-1, clause 5.1.5)
Tolerance The ISO 6789-1 value for the type, class and maximum, or the manufacturer’s tighter claim
System uncertainty interval From the calibration system’s certificate; it should be no more than ¼ of the tolerance (about 1 % for a ±4 % wrench). Check it with the TUR calculator
Exercised How many times the wrench was loaded to its maximum before the readings: three per direction for indicating and adjustable, graduated wrenches (Type I and Type II Classes A, D and G), according to AWS and Sushma Industries
R1 to R5 Every reading, as taken. Don’t discard any (DKD-L 10-1, §3.3 items 10 and 14)
Mean and largest deviation Deviation of each reading in %; the largest one decides pass or fail, because the tolerance applies to every reading
U From the lab certificate, or from your own budget (try the uncertainty calculator)
Next due Your interval, or 12 months or 5,000 cycles from first use if you have none yet (Norbar in Fastener + Fixing; UT Metrology). Recalibrate sooner after an overload of 25 % or more (Norbar USA)

Where to keep it and for how long

File each sheet under the wrench’s asset ID, with the lab certificate attached, and add a line to the wrench’s equipment record: date, result, decision and next due date. That history is what lets you lengthen or shorten the interval later (see calibration intervals). How long to keep it depends on the standards and regulations you work under; our calibration records guide sets out the rules by standard. Print a matching status sticker with the calibration label generator.

Get the PDF pack

The template on this page is free to print, as it is. If you’d rather have it as a PDF, subscribe to Traceable, our newsletter, with the form on this page: subscribers get the printable pack with our calibration program starter kit and the instrument record templates, this one included.

Related:

FAQ

What should a torque wrench calibration certificate include?

The wrench identification (ID, serial number, type and class, range, direction), the method (ISO 6789-2:2017 or the lab's procedure), the calibration system used and its traceability, any adapters, temperature and humidity, the results at each test point (usually the lowest value of the range, 60 % and 100 %, with five readings each), results before and after any adjustment, the measurement uncertainty at each point, and, if the certificate says pass or fail, the tolerance and the decision rule used. It also needs a unique number, the dates and the name of the person who authorized it.

How often should a torque wrench be calibrated?

If you have no interval of your own, the default commonly cited from ISO 6789 is 12 months or 5,000 cycles, whichever comes first, counted from when the wrench goes into use. Recalibrate sooner after an overload of 25 % or more above its maximum, a drop, a repair or other misuse.

My new torque wrench came with a calibration certificate. Is that enough?

Check what it is. Under the 2017 edition of ISO 6789, the manufacturer supplies a declaration of conformance, not a calibration certificate, because factory certificates did not contain enough information to be traceable. Some manufacturers still ship a factory certificate. If your quality system, customer or auditor needs a traceable calibration with uncertainty, ask for one to ISO 6789-2 when you buy the wrench or before first use.

The certificate in the box is dated months before I bought the wrench. Is it still valid?

The default interval associated with ISO 6789 is counted from when the wrench is put into use, not from the date printed on its declaration or factory certificate. Record the date the wrench goes into service on your record, and plan the first calibration from there, unless your own procedure sets a different rule.

Can I use this template for a torque screwdriver?

Yes. The structure is the same: identification, method, test points, readings, tolerance, uncertainty and decision. Change the tolerance: under ISO 6789-1, most torque screwdriver classes are held to ±6 %, and electronic indicating screwdrivers (Type I, Class E) to ±4 % if their maximum is above 10 N·m. Fixed and non-graduated setting screwdrivers follow a different reading scheme in ISO 6789-2, so check the standard before you use the five-reading table.

Do I need to record every reading, or only the mean?

Record every reading. The ISO 6789-1 tolerance applies to each reading, not to the mean, so one reading outside the limit is a fail. DKD's guidance on ISO 6789-2 also says not to discard readings or add preloads within a calibration step once the initial exercises are done.

Can I calibrate torque wrenches in-house and use this record instead of a lab certificate?

Yes, if your calibration system has a valid calibration certificate from an ISO/IEC 17025 lab or a national metrology institute, its relative uncertainty interval is no more than a quarter of the tolerance of the wrenches you calibrate, you follow a written procedure, and you evaluate the uncertainty. Your customer or auditor may still ask for an accredited certificate for some tools.

Sources

  1. ISO (sample distributed by iTeh Standards). ISO 6789-1:2017, Assembly tools for screws and nuts — Hand torque tools — Part 1 (official preview: foreword, clauses 1–5.1.8, Tables 3 and 4). 2017 — Paid standard; foreword items g and h, clauses 4, 5.1.3 and 5.1.5 read in the official preview. Under revision (ISO/DIS).
  2. ISO. ISO 6789-2:2017, Hand torque tools — Part 2: Requirements for calibration and determination of measurement uncertainty. 2017 — Paid standard, not read. Content described from DKD-L 10-1, AWS, Sushma Industries, Norbar and Torque Lab (two or more sources per statement).
  3. PTB / DKD. DKD-L 10-1, Determination of individual measurement uncertainty components according to DIN EN ISO 6789-2:2017. 2020 — §3.1 item 4 (adapters), §3.2 (method statement), §3.3 items 8, 10 and 14
  4. Advanced Witness Systems Ltd. Summary of new ISO standards: ISO 6789-1:2017 and ISO 6789-2:2017 (white paper v4). 2021 — Exercises and readings per type and class (3 exercises, 15 readings per direction for Type I and Type II A, D, G)
  5. Sushma Industries (R. Sadananda Murthy). Understanding the new torque wrench calibration standard: comparison between ISO 6789:2003 and ISO 6789:2017 (white paper v3.1). 2019
  6. Norbar Torque Tools. Understanding ISO 6789: Calibration Laboratories. 2017 — Calibration system and its certificate, ¼ rule, certificate content per point, Part 1 vs Part 2 deviation
  7. Fastener + Fixing Magazine. It's time to torque ISO 6789-2:2017 (N. Brodey, Norbar). 2017 — Declaration vs certificate; 12 months or 5,000 cycles from first use
  8. Torque Lab. What are the standard features you should see on an ISO 6789 compliant calibration certificate?. 2026
  9. UT Metrology. Torque Wrench Calibration: ASME vs ISO 6789 Explained
  10. Norbar USA. Torque wrench calibration (blog). 2014 — Recalibration after an overload of 25 % or more
  11. ASME. ASME B107.300-2021, Hand Torque Tools and Torque Testers (cover and table of contents). 2021
  12. Harbor Freight Tools. ICON 3/8 in. drive 20–100 ft-lb professional click torque wrench (product page) — Mentions a factory certificate of calibration; checked 2026-10-15
  13. NIST Office of Weights and Measures. ISO/IEC 17025 Crosswalk – Reporting the Results (clause 7.8). 2018 — Free summary of ISO/IEC 17025:2017 clause 7.8 (7.8.2.1, 7.8.4.1, 7.8.6)
  14. Global Accreditation Cooperation (Global ACI). Global ACI-TECH-1-002 (G), Guidelines on Decision Rules and Statements of Conformity (formerly ILAC G8:09/2019). 2026
  15. NIST. Metrological Traceability: Frequently Asked Questions and NIST Policy — Questions 5.1.6 and 5.2.4; checked 2026-10-29
  16. JCGM / BIPM. International Vocabulary of Metrology (VIM), JCGM 200:2012. 2012 — 3.11 Note 3 (calibration after adjustment)